Sorin SCP Pump Control Panel Recalled for Circuit Board Malfunction
Sorin Group USA is recalling SCP Pump Control Panels due to a circuit board defect that may cause display malfunctions and incorrect flow or pressure readings.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device used in cardiopulmonary bypass where incorrect flow and pressure displays pose a risk of harm to patients, but the source text does not report any specific injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products without reported injury.
Plain-English summary
Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding specific serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel. The affected units are identified by Product Code 60-02-50 and specific serial numbers including 60S09131, 60S09133-60S09135, 60S09140-60S09142, and 60S08954. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuits for typical durations of six hours or less.
The recall is due to an out-of-specification circuit board component within the panel. This defect may lead to a malfunction characterized by a blank or flashing display. Additionally, the flow and pressure display information may be incorrect following such a malfunction.
A total of 8 units are affected, with 2 units distributed in the USA (including Michigan and New York) and 6 units distributed outside the USA in Belgium, Canada, Japan, India, Russia, and the United Arab Emirates. Healthcare facilities and users should check their inventory for these specific serial numbers and contact Sorin Group USA for further instructions.
The recalled product
- Product
- SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping thro
- Manufacturer
- Sorin Group USA, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- Product Code: 60-02-50
- Serial Numbers: 60S09131
- 60S09133-60S09135
- 60S09140-60S09142
- 60S08954.
Distribution
Part of a larger recall action
This product is one of 2 recalled by Sorin Group USA, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Sorin Group USA, Inc.
See all →- HighSorin FlexFlow Venous Cannula Recalled for Sharpness Defect
FDA (Devices) · 2017-11-15
- HighSorin Group Aortic Arch Cannula Recalled for Excess Plastic on Tip
FDA (Devices) · 2017-08-16
- HighLivaNova Perfusion Tube Sets Recalled Due to Potential Sterile Barrier Breach
FDA (Devices) · 2017-08-09
- HighLivaNova Perfusion Tube Sets Recalled for Potential Sterile Barrier Breach
FDA (Devices) · 2017-08-09
- HighLivaNova Perfusion Tube Sets Recalled for Potential Sterile Barrier Breach
FDA (Devices) · 2017-08-09
Same hazard · display malfunction
See all →- HighHyundai vehicles instrument panel display software error recall
NHTSA · 2026-01-26
- ModerateMIRA Surgical System software update recall for motion display issue
FDA (Devices) · 2025-06-11
- HighTerumo CDI OneView Monitor Probe May Fail to Display Oxygen Saturation Data
FDA (Devices) · 2025-04-09
- HighEssenz HLM Heart-Lung Machines May Fail Due to Software Defect
FDA (Devices) · 2024-12-04
- HighMEDLINE Digital Blood Pressure Monitor Recalled Due to Microchip Faults
FDA (Devices) · 2024-10-16