The Recall Desk
Agency classClass IOur severity:SevereFDA (Devices)·Z-0128-2027·Announced 2026-10-28

Ventec VCSN+Pro Ventilator Recalled for Potential Connector Disconnection

Ventec Life Systems is recalling VCSN+Pro ventilators because an internal cable connector may disconnect, potentially interrupting ventilation therapy.

The agency’s classification

FDA (Devices): Class I

Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as Class I because the internal cable connector may disconnect, potentially interrupting ventilation therapy, which presents a reasonable probability of serious adverse health consequences. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

Ventec Life Systems, Inc. is recalling 4,368 units of the VCSN+Pro ventilator (Model Number: PRT-01184-000, Package PN: PRT-01172-000). The device is used for ventilatory support and was distributed nationwide in the United States, including Puerto Rico.

The recall is due to a defect where the internal flow transducer cable connector may disconnect from the internal board. This disconnection can potentially result in a system reboot and an interruption of ventilation therapy.

The source text does not report any illnesses, injuries, or deaths associated with this defect. Consumers and healthcare providers should contact Ventec Life Systems for instructions on how to address this issue.

The recalled product

Product
VCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support.
Affected units
4,368

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All Serial Numbers: UDI: 00855573007860

Distribution

Distributed nationwide across the United States.

The notice also names 1 state specifically:

Similar recalls