The Recall Desk

Manufacturer

Ventec Life Systems, Inc.

5 recalls in our database name Ventec Life Systems, Inc. as the manufacturing or recalling firm.

What the numbers show

Total
5
Critical
1
Severe
3
Most recent
2026-07-01

Of the 5 recalls from Ventec Life Systems, Inc. we have scored against our rubric, 1 (20%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 3 (60%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–5 of 5

  • SevereFDA (Devices)·Z-2525-2026·2026-07-01

    VOCSN+Pro Respiratory System Backup Battery Alarm Failure Recall

    Ventec Life Systems is recalling VOCSN+Pro respiratory systems due to a damaged backup battery that may prevent the backup alarm from activating during power loss, delaying caregiver awareness of a hazardous condition.

    Product
    VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1942-2026·2026-05-13

    VOCSN V+Pro Ventilator System Respiratory Device Safety Recall

    Ventec Life Systems is recalling VOCSN V+Pro ventilator units because they may not have been fully tested for high-pressure conditions, which could cause oxygen leaks and increase fire risk during use.

    Product
    VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
    Category
    Medical Device
    Distribution
    4 states
  • SevereFDA (Devices)·Z-1196-2025·2025-03-05

    Ventec VOCSN Ventilators Recalled Due to Incorrect Service Parts

    Ventec Life Systems recalled VOCSN Multi-Function Ventilators (30 units in CA, NY, MO) due to incorrect parts used during service. Defects may cause unexpected shutdown or impaired respiratory monitoring and support.

    Product
    VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002
    Category
    Medical Device
    Distribution
    3 states
  • CriticalFDA (Devices)·Z-1070-2024·2024-03-06

    VOCSN Pediatric breathing package spiral wrap detachment risk during ventilation

    Ventec Life Systems recalls VOCSN Patient Breathing Packages because the spiral wrap may detach during ventilation, compromising the breathing circuit and creating a risk of serious injury or death, especially for ventilator-dependent patients.

    Product
    VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ven
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-1902-2021·2021-06-30

    Ventec VOCSN Ventilators May Shut Down Due to Component Failure

    Ventec VOCSN ventilators may unexpectedly shut down or lose blower function during patient use due to component failure. This could prevent the device from providing ventilation therapy. About 1,934 devices are affected.

    Product
    Ventec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900-001; V+C+Pro/Japan/PRT-00900-100; V+C/English/PRT-01098-000; V+O+C+S+N/English/PRT-01100-00
    Category
    Medical Device
    Distribution
    Distributed nationwide

About this company

Community-sourced company facts from Wikidata, not from any recall agency. This is background on the corporate entity — it says nothing about the recall(s) listed on this page, which come from the agency source notice above.

Ventec Life Systems

American medical device company

Country
United States
Founded
2012

Official website