Aerogen Recalls FRIWO Power Adapters for Aeroneb Nebulizer Systems
Aerogen is recalling FRIWO AC/DC power adapters for Aeroneb nebulizers because the cover may break and dislodge before use.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device component with a physical defect (broken cover) that poses a risk of harm, but the source text does not report any specific injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Aerogen Ltd. is recalling 1,223 FRIWO AC/DC power adapters (Model GPP10) used in the Aeroneb Solo and Pro Nebulizer systems. The recall was initiated after the company received a small number of complaints indicating that the power adapter cover had broken and dislodged from the body of the adapter prior to use.
The affected adapters were distributed worldwide, including in the United States (nationwide, with specific mentions of CA, GA, IL, MS, NC, and TX) and numerous international markets such as Australia, Canada, China, and the United Kingdom. The adapters are components supplied with various Aeroneb Professional and Solo Nebulizer System models, including AG-AS3000 and AG-PX1000 series units.
Consumers should stop using the affected power adapters immediately. Users can identify the affected units by checking the model number (GPP10) and cross-referencing the lot numbers provided in the official recall documentation. Aerogen is the recalling firm responsible for this action.
The recalled product
- Product
- The Aeroneb Professional Nebulizer System is a portable medical device for multiple patient use that is intended to aerosolize physician-prescribed solutions for inhalation to patients on and off ventilation or other positive pressure breathing assistance. The Aeroneb…
- Manufacturer
- Aerogen Ltd.
- Affected units
- 1,223
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- FRIWO AC/DC Adapter Model Number: GPP10
- Lot Numbers: 9144140519057
- 9144140616062
- 9144140702155
- 9144140710034
- 2) Model Number: AG-AS3000-IN
- Lot Numbers: 9125140522116
- 9125140526101
- 9125140527077
- 9125140609002
- 9125140609007
- 9125140616064
- 9125140627057
- 9125140702166
- 9125140710031
- 9125140805010
- 9125140805134
- 3) Model Number: AG-AS3000-NE
- Lot Numbers: 9122140520089
- 9122140521089
Distribution
Part of a larger recall action
This product is one of 4 recalled by Aerogen Ltd. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 4 products in this recall →Related recalls
Same manufacturer · Aerogen Ltd.
See all →- ModerateAerogen Nebulizer Power Adapter Recall Due to Broken Cover
FDA (Devices) · 2014-10-29
- ModerateAerogen Recalls FRIWO Power Adapters for Aeroneb Nebulizer Systems
FDA (Devices) · 2014-10-29
- ModerateAerogen Nebulizer Power Adapter Recall Due to Broken Cover
FDA (Devices) · 2014-10-29
Same hazard · physical defect
See all →- HighThermodilution Catheter May Bend During Use, Causing Monitor Errors
FDA (Devices) · 2023-04-19
- ModerateBard Clean-Cath Intermittent Catheter Recalled for Packaging and Tip Defects
FDA (Devices) · 2021-04-28
- HighFerring Recalls ZOMA-Jet 5 Demonstration Kits Due to Defective Delivery System
FDA (Drugs) · 2021-04-07
- HighROi CPS Recalls Custom Procedure Packs with Flaking Gauze Sponges
FDA (Devices) · 2021-01-06
- HighGE Healthcare Recalls Flow Sensor Service Parts with Damaged Tubes
FDA (Devices) · 2020-11-11
Similar recalls
- SevereDraeger Oxylog 3000 Plus Ventilators Recalled Due to Potentiometer Fault
FDA (Devices) · 2014-04-16
- CriticalDraeger Perseus A500 Anesthesia Workstation Recalled for Power Switch Failure
FDA (Devices) · 2015-12-23
- HighAC Power Adapter for LIFEPAK 15 Defibrillator May Fail to Charge or Power On
FDA (Devices) · 2023-12-27
- SevereFDA Recalls 324 Puritan Bennett 980 Ventilators Due to Software Issue
FDA (Devices) · 2014-11-05
- SevereVentilation Device Recall: Set2Go Ventilation 12 Breathing Circuit Connections May Loosen
FDA (Devices) · 2023-05-24
- SevereVentlab AirFlow Resuscitator Recalled for Inlet Valve Leakage
FDA (Devices) · 2012-11-07