Chest Tube Insertion Tray Recalled Due to Over-Sterilization of Suture Component
Medline recalls CENTURION SnagFree Chest Tube insertion trays because a silk suture component was sterilized twice instead of once, potentially compromising its integrity. The affected lot contains 18 kits distributed in California and Texas.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II medical device recall involving a surgical component (silk suture) that was not processed according to specifications. Although no illnesses or injuries have been reported, the affected chest tube insertion tray poses a potential risk of harm during use if the over-sterilized suture component fails or loses integrity.
Plain-English summary
Medline Industries is recalling CENTURION SnagFree Chest Tube insertion trays (model CHT545) because one lot of 18 kits was subjected to a processing error that does not meet product specifications.
The chest tube insertion tray contains a silk suture component that is restricted to one sterilization cycle. The affected lot was processed through two sterilization cycles, which may compromise the integrity of the suture component.
The affected kits were distributed in California and Texas. No reported illnesses or injuries have been associated with this recall.
The recalled product
- Product
- CENTURION SnagFree Chest Tube insertion Tray, Reorder CHT545
- Manufacturer
- MEDLINE INDUSTRIES, LP - Northfield
- Affected units
- 18
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- UDI/DI 20653160222979
- Lot Code 2022050690
Distribution
Related recalls
Same manufacturer · MEDLINE INDUSTRIES, LP - Northfield
See all →- CriticalIntubation trays recalled after laryngoscope handle lights failed to illuminate
FDA (Devices) · 2025-10-01
- SevereMedline Medical Convenience Kits with Defective Laryngoscope Handles
FDA (Devices) · 2025-10-01
- HighGood Neighbor alcohol prep pads recalled for alcohol below labeled concentration
FDA (Drugs) · 2025-09-03
- HighSurgical kits recalled because Foley tray inserts carry incorrect product information
FDA (Devices) · 2025-09-03
- HighSurgery trays recalled because they contain recalled tourniquet cuffs
FDA (Devices) · 2025-09-03
Same hazard · sterilization defect
See all →- SevereMedline Convenience Kits recalled for potential sterility assurance defects
FDA (Devices) · 2026-05-13
- ModerateMedline Burn Care Kits Recalled Due to Sterilization Calibration Issues
FDA (Devices) · 2026-05-13
- HighMedline PPE Kit recalled for sterilization calibration issues
FDA (Devices) · 2026-05-13
- SevereMedline Convenience Kits recalled for potential sterilization issues
FDA (Devices) · 2026-05-13
- HighMedline Convenience Kits recalled due to sterilization equipment calibration issues
FDA (Devices) · 2026-05-13
Similar recalls
- HighSurgical kits recalled due to compromised sterile wrapping
FDA (Devices) · 2023-02-22
- ModerateSurgical Circumcision Kits Recalled Due to Non-Sterile Outer Wrapping
FDA (Devices) · 2023-02-22
- SevereMedline recalls Centurion Laceration Trays for potential sterility defect
FDA (Devices) · 2024-03-06
- SevereMedline circumcision surgical kits recalled for sterility concerns
FDA (Devices) · 2024-03-06
- HighCenturion Medical Recalls Suture Trays Due to Sterile Packaging Concerns
FDA (Devices) · 2019-08-14
- SevereMedline Tracheostomy Care Trays Recalled for Sterility Issues
FDA (Devices) · 2024-03-06