Siemens Syngo RT Oncologist software update addresses safety issues
Siemens Medical Solutions is updating Syngo RT Oncologist software versions 4.2 and 4.3 to fix safety and performance issues.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, which typically indicates a temporary or medically reversible adverse health consequence. The source describes the action as providing a software update for safety and performance fixes without reporting specific injuries or hospitalizations, fitting the criteria for a moderate severity precautionary recall.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 14 units of the Syngo RT Oncologist software, an optional accessory to linear accelerator systems used for radiation therapy planning. The recall involves units currently running software version 4.2 or 4.3.
The recall is intended to provide a software update that contains important safety and performance fixes. These fixes address safety-related issues previously described in a Customer Safety Notice regarding UFSN-RTT/RTO 4.3 Adaptive Targeting [Auto-registration].
The affected units were distributed to facilities in Massachusetts, Missouri, Ohio, Pennsylvania, Utah, and Wisconsin. Siemens is providing the update to resolve the identified safety concerns.
The recalled product
- Product
- Syngo RT Oncologist, an optional accessory to the linear accelerator system and permits localization, contouring, segmentation, image review, and review and approval of treatment plan parameters. In addition, it includes tools and administrative functions to aid in the diagnosis,
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Hazard
- Affected units
- 14
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Model Number of device- 10568604
Distribution
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · software defect
See all →- HighMidmark IQecg Digital ECG Software Recall for Inaccurate Waveforms
FDA (Devices) · 2026-09-02
- HighMidmark Digital ECG Software Recall for Incorrect Waveform Display
FDA (Devices) · 2026-09-02
- ModerateEndoFLIP System Recall Due to Incorrect Catheter Expiration Flags
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighMitsubishi Recalls Outlander Models for Rearview Camera Software Defect
NHTSA · 2026-08-25