Natus Medical Recalls Medix NatalCare LX Incubators Due to Wheel Defect
Natus Medical is recalling 104 Medix NatalCare LX Intensive Care Incubators because a defective base weld may cause wheels to bend out of alignment.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II and involves a structural defect in a medical device used for neonatal intensive care. While no injuries are reported, the risk of harm from a malfunctioning incubator is significant, fitting the 'High' severity criteria for risk-of-harm products.
Plain-English summary
Natus Medical Incorporated is recalling 104 units of the Medix NatalCare LX Intensive Care Incubator. The affected item numbers are DMNCLX1, KITNC-STLX0003, KITNC-STLX0006, and KITNC-STLX0007. The recall covers specific serial codes listed in the agency's structured data.
The recall is due to a defective base where the weld holding the wheel bushing to the tube is inadequate. This defect has resulted in several complaints of wheels bending out of alignment. The devices were distributed in the United States to the states of AL, CA, FL, IL, IN, KY, LA, MA, MS, NJ, NY, OK, PA, SC, and VA, as well as internationally to Pakistan.
Hospitals and healthcare facilities should stop using the affected incubators and contact Natus Medical Incorporated for instructions on repair or replacement. The FDA has classified this as a Class II recall.
The recalled product
- Product
- The Medix NatalCare LX Intensive Care Incubator. Neonatal incubator. Item numbers: DMNCLX1; KITNC-STLX0003; KITNC-STLX0006; and KITNC-STLX0007.
- Manufacturer
- Natus Medical Incorporated
- Affected units
- 104
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 5045
- 5046
- 5047
- 5048
- 5051
- 5053
- 5056
- 5057
- 5096
- 5097
- 5128
- 5129
- 5130
- 5131
- 5132
- 5133
- 5134
- 5135
- 5136
- 5138
Distribution
Related recalls
Same manufacturer · Natus Medical Incorporated
See all →- HighExternal Drainage System Collection Bags Recalled for Sterility Assurance Issues
FDA (Devices) · 2024-01-10
- HighEDS 3 CSF External Drainage System recalled for incomplete sterility testing
FDA (Devices) · 2024-01-10
- HighNatus Cranial Access Kit Recalled Due to Defective Sterile Surgical Drapes
FDA (Devices) · 2022-12-28
- ModerateNatus Medical Recalls Olympic Pasteurmatic 3000 and 3500 Disinfection Systems
FDA (Devices) · 2017-07-19
- HighNatus Medical Recalls neoBLUE 2 Phototherapy Lights for Intensity Defect
FDA (Devices) · 2017-05-10
Same hazard · structural defect
See all →- SevereSubaru Recalls 2026 Forester and Crosstrek for Ball Joint Defect
NHTSA · 2026-10-19
- SevereVolvo Recalls EX30 and EX30CC Vehicles for Head Restraint Defect
NHTSA · 2026-10-09
- Severe
- Severe
- SevereDecathlon Recalls Van Rysel GRVL AF Bicycles Due to Crash Hazard
CPSC · 2026-10-08
Similar recalls
- HighNatus Medical Recalls Medix NatalCare ST-LX Neonatal Incubators
FDA (Devices) · 2016-04-20
- HighICU Medical Recalls NanoClave Administration Sets Due to Seal Defect
FDA (Devices) · 2020-10-28
- SevereImactis NaviKit Stereotaxic Accessory Recalled for Mechanical Defect
FDA (Devices) · 2021-03-24
- HighICU Medical Recalls NanoClave Extension Sets Due to Seal Defect
FDA (Devices) · 2020-10-28
- HighNatus Cranial Access Kit Recalled Due to Defective Sterile Surgical Drapes
FDA (Devices) · 2022-12-28
- HighICU Medical Recalls Smallbore Extension Sets Due to Seal Defect
FDA (Devices) · 2020-10-28