Philips is recalling five Allura Xper FD20/20 OR Table systems because the BIOS battery may deplete faster than designed, halting start-up.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: Scored 2 as a component failure whose stated effect is that the system will not start; the notice describes no in-use hazard or patient harm.

Plain-English summary

Allura Xper FD20/20 OR Table interventional X-ray systems, model number 722039, are being recalled because the BIOS battery may deplete faster than was anticipated during the design phase.

When the battery is depleted, the start-up process of the system is halted and the system will not start. No user messages are shown before the battery is low on power or depleted.

The recall covers five units, all outside the United States, carrying UDI-DI 00884838054271 and serial numbers 2, 4, 5, 6 and 8. Distribution was worldwide. The FDA has classified the action as Class II.

Affected sites should follow Philips' instructions for battery replacement and plan for the possibility of a system failing to start without prior warning.

The recalled product

Product
Allura Xper FD20/20 OR Table; Model Numbers: 722039; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Hazard
  • battery-depletion
  • startup-failure
  • device-unavailability

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (7)

  • Model Numbers: 722039
  • UDI-DI: 00884838054271
  • Serial Numbers: 2
  • 4
  • 5
  • 6
  • 8

Distribution

Distributed nationwide across the United States.