Philips Allura Xper FD20/15 OR Table System Startup Failure Due to BIOS Battery Depletion

Philips is recalling certain Allura Xper FD20/15 OR Tables because the BIOS battery may deplete faster than expected, preventing the system from starting without warning. No user warnings appear before the battery becomes depleted.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: This is an FDA Class II device recall involving a functional failure (BIOS battery depletion preventing system startup) with no reported hospitalizations, injuries, or illnesses. The hazard is operational disruption rather than direct patient harm.

Plain-English summary

Philips Medical Systems Nederland B.V. is recalling the Allura Xper FD20/15 OR Table, Model 722059, due to a BIOS battery defect. The BIOS battery may deplete faster than expected based on design specifications.

When the BIOS battery is fully depleted, the operating room table cannot start up and complete its initialization. The system provides no warning message to users before the battery reaches critically low levels or becomes completely depleted.

Affected devices include 8 units worldwide: 2 in the United States and 6 internationally distributed across numerous countries.

Philips Medical Systems Nederland B.V. has initiated a recall of the affected units. Healthcare facilities and users of this equipment should contact Philips for information about service or replacement options.

The recalled product

Product
Allura Xper FD20/15 OR Table; Model Numbers: 722059; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Hazard
  • battery-depletion
  • startup-failure
  • no-warning

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (10)

  • Model Numbers: 722059
  • UDI-DI: 00884838059122
  • Serial Numbers: 7
  • 33
  • 35
  • 40
  • 38
  • 41
  • 37
  • 2

Distribution

Distributed nationwide across the United States.