Allura Xper FD20/15 OR Table; Model Numbers: 722059; Associated 510(k)s: K102005, K130638, K130842
Pending LLM rewrite. Source: FDA_DEVICE Z-0011-2026.
Severity not yet assessed
We have not scored this recall against our rubric yet, so we are not showing a severity rating or our usual guidance. The agency’s own notice below is the authoritative source — read it first.
Plain-English summary
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
The recalled product
- Product
- Allura Xper FD20/15 OR Table; Model Numbers: 722059; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Category
- Medical Device — Devices
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (10)
- Model Numbers: 722059
- UDI-DI: 00884838059122
- Serial Numbers: 7
- 33
- 35
- 40
- 38
- 41
- 37
- 2
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- SevereAbiomed Impella CP Smart Assist Set Heart Pump Introducer Sheath Leakage
FDA (Devices) · 2026-07-22
- HighMedtronic DLP One-Piece Pediatric Arterial Cannula Recalled for Foreign Material
FDA (Devices) · 2026-07-22
- HighVasoview Hemopro 2 Endoscopic Vessel Harvesting System Recall
FDA (Devices) · 2026-07-22
- SevereMedtronic Bio-Medicus Adult Cannula Kit Recalled for Foreign Material
FDA (Devices) · 2026-07-22
- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22