The Recall Desk

Manufacturer

NovaSignal Corp.

4 recalls in our database name NovaSignal Corp. as the manufacturing or recalling firm.

What the numbers show

Total
4
Critical
0
Severe
0
Most recent
2024-07-03

Of the 4 recalls from NovaSignal Corp. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–4 of 4

  • HighFDA (Devices)·Z-2217-2024·2024-07-03

    NovaGuide 2 Ultrasound System Recall Due to Grounding Cable Disconnection Risk

    NovaGuide 2 Traditional Ultrasound systems may have grounding cables that can be inadvertently disconnected during use, creating risk of electrical shock and equipment malfunction. Twenty-eight systems have been distributed nationwide.

    Product
    NovaGuide 2 Traditional Ultrasound , REF: NSC-TCDNG2
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2216-2024·2024-07-03

    NovaGuide 2 Ultrasound Systems Recalled for Electrical Shock Risk

    FDA is recalling 45 NovaGuide 2 Intelligent Ultrasound systems due to a defective grounding cable that could become disconnected and cause electrical shock, electrical interference, or equipment malfunction.

    Product
    NovaGuide 2 Intelligent Ultrasound, REFs: NSC-NVGSYS2 & NSC-NVGSYS2-CA
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1049-2023·2023-02-08

    Lucid TCD 2.0 Transcranial Ultrasound System Recalled for Hard Drive Failure

    The Lucid TCD 2.0 System has a faulty hard drive that causes the device to freeze and reboot. Eventually, the system becomes completely inoperable.

    Product
    Lucid TCD 2.0 System, Model: NA-LUCM1. Transcranial Ultrasound System
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1050-2023·2023-02-08

    Ultrasound system hard drive defect causes device freezing and failure

    FDA recall of NovaGuide Intelligent Ultrasound System due to faulty hard drive causing device freezing and eventual system failure.

    Product
    NovaGuide Intelligent Ultrasound System, Model: NA-RBTSYS1. Comprised of NovaBot Automated Headset (AKA: NeuralBot, Model: NA-RBT1) and Lucid TCD 2.0 Transcranial Doppler Ultrasound (Model: NA-LUCM1)