The Recall Desk

Hazard

Steam Leak recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
1
Most recent
2025-07-17

Of the 1 steam leak recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1 (100%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all steam leak recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • SevereCPSC·25389·2025-07-17

    CasaClean handheld steamers can leak steam or spew hot boiling water

    CasaClean Handheld Steamers, model DF-958, are being recalled because they can leak steam or spew hot boiling water. Thirteen consumers have received burns. About 23,000 units are affected.

    Product
    CasaClean Handheld Steamers
    Category
    Consumer Product
    Distribution
    Distributed nationwide