The Recall Desk

Hazard

Lp Gas Hazard recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2023-10-02

Of the 1 lp gas hazard recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all lp gas hazard recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighNHTSA·23V665000·2023-10-02

    2024 Holiday Rambler Invicta Motorhome Griddle Fire Hazard Recall

    REV Recreation Group is recalling certain 2024 Holiday Rambler Invicta motorhomes. The exterior griddle may be stowed while still connected to the LP gas line, creating a fire risk.

    Product
    FLEETWOOD — FLEETWOOD, HOLIDAY RAMBLER FORTIS, INVICTA
    Category
    Vehicle
    Distribution
    Distributed nationwide