The Recall Desk

Hazard

Infusion recalls

106 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
106
Critical
2
Severe
0
Most recent
2026-01-21

Of the 106 infusion recalls we have scored against our rubric, 2 (2%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all infusion recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

26–50 of 106

  • HighFDA (Devices)·Z-0714-2026·2025-12-03

    ULTRASITE set 352049-C recalled over backflow and inability to prime

    B. Braun Medical is recalling 41,400 ULTRASITE sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    ULTRASITE set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 2 ULTRASITE VLV, 10
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0695-2026·2025-12-03

    IV set 490193 recalled over backflow and inability to prime

    B. Braun Medical is recalling 111,312 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15DR HF 4-WAY
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0702-2026·2025-12-03

    IV set 354208 recalled over medication backflow and priming failure

    B. Braun Medical is recalling 126,050 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15DR, CARESIT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0682-2026·2025-12-03

    IV set 490488 recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 187,464 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0698-2026·2025-12-03

    IV set 490274 recalled over backflow into the primary container

    B. Braun Medical is recalling 98,568 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 3 CARESITE LAD
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0711-2026·2025-12-03

    SafeDAY set 352635 recalled over piggyback backflow and inability to prime

    B. Braun Medical is recalling 200,917 SafeDAY sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    SafeDAY set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 60 DROP W/2 SAFEDAY; Cat
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0631-2026·2025-12-03

    Extension set 490629 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 13,128 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXTENSION SET, 25 IN.; Catalog Numb
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0707-2026·2025-12-03

    Outlook set 354213 recalled over backflow into the primary container

    B. Braun Medical is recalling 18,072 units outlook sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK IV SET 15DROP W/3 CARESITE; C
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0578-2026·2025-12-03

    IV set 490487 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 23,160 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    FoundationOne Liquid CDx (F1LCDx), TECH-0009 version 6.0
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0703-2026·2025-12-03

    IV set 490301 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 31,392 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV AD.SET W/15 DR 128 IN, E
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0632-2026·2025-12-03

    Short extension sets recalled over piggyback backflow and occlusion

    B. Braun Medical is recalling 13,600 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT SET W/ B/C VLV, ULTRAPORT, 57 I
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0664-2026·2025-12-03

    IV set 490315 recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 11,832 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15 DR/ML,2 CA
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0630-2026·2025-12-03

    IV extension sets recalled over backflow between piggyback and primary containers

    B. Braun Medical is recalling 43,452 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. EXTENSION SET, 18 IN; Catalog Numb
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0604-2026·2025-12-03

    ADDitIV sets recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 100 ADDitIV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. V1782F UNIV IV SET W/0.2 MICRON FILTER; C
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0665-2026·2025-12-03

    IV set 490489 recalled over piggyback backflow and inability to prime

    B. Braun Medical is recalling 3,120 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15DR, 92 IN W/
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0668-2026·2025-12-03

    IV set 490504 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 36,936 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0710-2026·2025-12-03

    Rate flow sets recalled over backflow and inability to prime

    B. Braun Medical is recalling 150,664 RATE FLOW sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    RATE FLOW set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. RATEFLOW ADMIN 15 DROP W/2 CARESITE
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0669-2026·2025-12-03

    High volume IV sets recalled over piggyback backflow and occlusion

    B. Braun Medical is recalling 190,625 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DROP W/ CAR
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0647-2026·2025-12-03

    Single lot IV sets recalled over backflow and priming occlusion

    B. Braun Medical is recalling 48 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN 15D, 2 B/C VLV, 3
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0626-2026·2025-12-03

    Caresite set 354207 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 381,850 Caresite primary gravity sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Primary Gravity Filtered Set with 2 CARE
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0603-2026·2025-12-03

    Primary gravity sets recalled over backflow from secondary containers

    B. Braun Medical is recalling 43,900 ADDitIV primary gravity sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. Primary Gravity IV Set, 2 Non-Needle-Fre
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0691-2026·2025-12-03

    IV set 490408 recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 24,072 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 2 CARESITES,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0646-2026·2025-12-03

    IV set 457506 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 1,019 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET W/2 CARESITE & 1 ULT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0694-2026·2025-12-03

    IV administration sets recalled over backflow from piggyback containers

    B. Braun Medical is recalling 2,544 IV administration sets over the potential for medication to flow back from secondary piggyback containers into primary containers, and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 124 IN.; Catal
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0645-2026·2025-12-03

    IV set 490067 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 407,800 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET 15DR 5CARESITE 1 ULR
    Category
    Medical Device
    Distribution
    0 states