The Recall Desk

Hazard

Guard Detachment recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
1
Most recent
2022-08-04

Of the 1 guard detachment recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1 (100%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all guard detachment recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • SevereCPSC·22196·2022-08-04

    DeWALT 12-inch Sliding Compound Miter Saws Recalled for Injury Hazard

    DeWALT is recalling approximately 1.36 million 12-inch sliding compound miter saws because the rear safety guard can break or detach, creating a risk of projectiles and laceration injuries. Consumers should stop using the saws and contact DeWALT for a free repair.

    Product
    DeWALT 12-inch Sliding Compound Miter Saws
    Category
    Consumer Product
    Distribution
    Distributed nationwide