The Recall Desk

Hazard

Corrosive Injury recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2025-03-06

Of the 1 corrosive injury recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all corrosive injury recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighCPSC·25173·2025-03-06

    Taylor Water Technologies Recalls Phosphate Reagent Bottles Due to Chemical Burn Risk

    Taylor Water Technologies is recalling about 10,000 Phosphate Reagent #1 bottles sold with test kits. The sulfuric acid bottles lack required child-resistant packaging, posing a risk of severe chemical burns.

    Product
    Phosphate Reagent #1 Bottles sold within the Taylor Phosphate Test Kits
    Category
    Consumer Product
    Distribution
    Distributed nationwide