The Recall Desk

Hazard

Bridle Line Failure recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2022-07-28

Of the 1 bridle line failure recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all bridle line failure recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighCPSC·22764·2022-07-28

    Ceramic Pulleys for Cabrinha Switchblade and Drifter Kites Recalled for Injury Hazard

    Adventure Sports is recalling about 1,950 ceramic pulleys used in Cabrinha Switchblade and Drifter kiteboarding kites because the pulleys can cause bridle line breakage, reducing steering and control and risking kiteboarder injury.

    Product
    Ceramic Pulley for Switchblade and Drifter kite sailing kites
    Category
    Consumer Product
    Distribution
    Distributed nationwide