The Recall Desk

FDA (Devices) recall action 96856

Baxter Healthcare Corporation recalled 2 products on 2025-07-16

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-07-16
Critical
0
Units
19,738

Why these products were recalled

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • SevereFDA (Devices)··2025-07-16

    Hillrom Q-Link 13 patient lift components may falsely latch and detach

    The Hillrom Q-Link 13 component used with mobile patient lifts is being recalled because it may allow false latching of the quick-release hook. 9,899 units are affected.

    Product
    Hillrom Q-Link 13 component, Product Code 3156509, used with the following Mobile Patient Lifts: 1) Uno 102 EE Mobile lift - Product Code 2010004; 2) Viking L Mobile lift - Product Code 2040044; 3) Viking XL Mobile lift - Product Code 2040043; 4) Viking M Mobile lift…
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2025-07-16

    Hillrom LikoScale patient lift components may falsely latch and detach

    Hillrom LikoScale components used with patient lift adapter kits are being recalled because the Q-link 13 may allow false latching of the quick-release hook. 9,839 units are affected.

    Product
    Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.
    Category
    Distribution
    0 states