The Recall Desk

FDA (Devices) recall action 96743

Dexcom, Inc. recalled 4 products on 2025-06-25

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2025-06-25
Critical
3
Units
703,687

Why these products were recalled

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • CriticalFDA (Devices)··2025-06-25

    Dexcom One Continuous Glucose Monitoring System Receiver Speaker Defect

    Dexcom, Inc. is recalling Dexcom One Continuous Glucose Monitoring System receivers due to potential defective foam or assembly errors that may cause the speaker to lose contact with internal components, resulting in missed audible alerts for dangerous blood glucose levels.

    Product
    Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2)
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2025-06-25

    Dexcom G6 Glucose Receiver recalled for missed audible alerts

    Dexcom G6 Glucose Receiver units may have defective foam or assembly errors that disable audible alerts for dangerously low or high blood glucose, risking seizures, loss of consciousness, or death.

    Product
    Dexcom G6 Glucose Receiver, SKUs (Part Number): STK-FM-001 (MT27408-1) & STK-FR-001 (MT27408-1)
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-06-25

    Dexcom One+ Continuous Glucose Monitoring System Receiver Speaker Defect

    Dexcom, Inc. is recalling 26,291 units of the Dexcom One+ Continuous Glucose Monitoring System due to defective foam or assembly errors that may prevent audible alerts for low or high blood glucose levels, potentially leading to seizures, loss of consciousness, or death.

    Product
    Dexcom One+ Continuous Glucose Monitoring System, SKU(Part Number): STK-D7-013 (MT28285-1), STK-D7-014 (MT28285-1), STK-D7-103 (MT28285-2)
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-06-25

    Dexcom G7 Glucose Receiver recalled due to defective foam or assembly error

    Dexcom, Inc. is recalling Dexcom G7 Glucose Receiver units due to defective foam or assembly error that may cause the speaker to lose contact with the circuit board, preventing audible alerts for low or high blood glucose.

    Product
    Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013 (MT26403-1) , STK-GT-019 (MT26403-4) , STK-GT-100 (MT26403-2), STK-GT-109 (MT26403-2) , STK-GT-113…
    Category
    Distribution
    0 states