FDA (Devices) recall action 95515

AVID Medical, Inc. recalled 73 products on 2024-11-20

The FDA (Devices) publishes a recall like this one as 73 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 73 pages.

Products
73
Announced
2024-11-20
Critical
0
Units
7,259

Why these products were recalled

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

26–50 of 73

  • HighFDA (Devices)··2024-11-20

    Medical surgical kits recalled for loose metal flakes in instruments

    AVID Medical is recalling Halyard CYSTO PDS medical convenience kits due to loose metal flakes that may detach from sponge forceps and towel clamps. These flakes could enter a surgical site and cause foreign body reactions.

    Product
    Halyard CYSTO PDS - Medical convenience kits Model Number: GVVA002-05
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard SAMMC ANGIOGRAPHY Medical Kits Recalled for Loose Metal Flakes

    AVID Medical recalls Halyard SAMMC ANGIOGRAPHY medical convenience kits due to loose metal flakes on sponge forceps and towel clamps. Metal flakes may enter surgical sites, potentially causing local reactions or foreign body complications.

    Product
    Halyard SAMMC ANGIOGRAPHY - Medical convenience kits Model Number: SAMM066-15
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Medical Convenience Kits Recalled for Detachable Metal Flakes

    AVID Medical recalls Halyard TAH-BUNDLE medical kits with sponge forceps and towel clamps. Loose metal flakes may detach and enter surgical sites, potentially causing local or foreign body reactions.

    Product
    Halyard TAH-BUNDLE - Medical convenience kits Model Number: LANC039-29
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard ORTHO PACK surgical instrument kits recalled due to metal flakes

    AVID Medical is recalling Halyard ORTHO PACK kits because sponge forceps and towel clamps may shed loose metal flakes that could enter a patient's surgical site, causing local or foreign body reactions.

    Product
    Halyard ORTHO PACK -Medical convenience kits Model Number: MMOK006-04
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard IR Biopsy Pack recalled for loose metal flakes in surgical kit

    AVID Medical is recalling Halyard IR BIOPSY PACK due to loose metal flakes in forceps and towel clamps that could detach and enter the surgical site, potentially causing local or foreign body reactions.

    Product
    Halyard IR BIOPSY PACK - Medical convenience kits Model Number: AVMD019-01
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Surgical Sponge Forceps and Towel Clamps Recalled for Loose Metal Flakes

    AVID Medical is recalling Halyard CATH LAB PACK surgical kits because sponge forceps and towel clamp components may shed loose metal flakes that could enter a patient's surgical site, posing a risk of local or foreign body reactions.

    Product
    Halyard CATH LAB PACK - Medical convenience kits Model Number: VANW016-05
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Biopsy Pack surgical instruments recalled due to loose metal flakes

    AVID Medical is recalling Halyard BIOPSY PACK surgical kits due to loose metal flakes from sponge forceps and towel clamps. Metal flakes could enter a patient's surgical site, potentially causing local or foreign body reactions.

    Product
    Halyard BIOPSY PACK - Medical convenience kits Model Number: VANW019-04
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Angio Tray surgical kits recalled for loose metal flakes

    AVID Medical recalls Halyard ANGIO TRAY (Model VAPX021-05) surgical kits due to potential loose metal flakes on forceps and clamps that could enter surgical sites and cause local or foreign body reactions.

    Product
    Halyard ANGIO TRAY - Medical convenience kits Model Number: VAPX021-05
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Medical Convenience Kits Recalled for Detachable Metal Flakes in Surgical Instruments

    AVID Medical is recalling Halyard Shoulder Pack medical convenience kits because the sponge forceps and towel clamps contain loose metal flakes that could detach and enter a patient's surgical site during use, potentially causing foreign body reactions.

    Product
    Halyard SHOULDER PACK - Medical convenience kits Model Number: PANJ005
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Medical Drainage Pack Components Recalled for Potential Metal Fragments

    AVID Medical is recalling the Halyard IR Abcess Drainage Pack due to sponge forceps and towel clamps that may have loose metal flakes. Metal flakes could detach and enter a patient's surgical site, causing local or foreign body reactions.

    Product
    Halyard IR ABCESS DRAINAGE PACK - Medical convenience kits Model Number: AVMD018
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Surgical Instruments Recalled Due to Potential Metal Flake Contamination

    AVID Medical recalled Halyard POST PARTUM TUBAL LIGATION surgical kit components due to loose metal flakes that could detach from sponge forceps and towel clamps. Metal flakes entering a patient's surgical site could cause local or foreign body reactions.

    Product
    Halyard POST PARTUM TUBAL LIGATION - Medical convenience kits Model Number: LANC026-19
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Laparotomy GYN Packs recalled due to loose metal flakes

    AVID Medical recalls Halyard LAPAROTOMY GYN surgical packs due to loose metal flakes on sponge forceps and towel clamps. Metal fragments may enter the surgical site and cause local or foreign body reactions.

    Product
    Halyard LAPAROTOMY GYN PACK - Medical convenience kits Model Number: SAMM018-13
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Pacemaker Pack surgical kits recalled for loose metal flakes

    AVID Medical is recalling Halyard Pacemaker Pack surgical kits because sponge forceps and towel clamps contain loose metal flakes that could detach and enter patient surgical sites.

    Product
    Halyard PACEMAKER PACK - Medical convenience kits Model Number: MAMC223-08
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Knee Arthroscopy Packs recalled due to loose metal flakes

    AVID Medical recalls Halyard Knee Arthroscopy Packs due to loose metal flakes that could detach from forceps and clamps. Metal fragments may enter the surgical site, potentially causing local or foreign body reactions.

    Product
    Halyard KNEE ARTHROSCOPY PACK - Medical convenience kits Model Number: MAMC204-16
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard ANGIO PACK surgical kits recalled for loose metal fragments

    AVID Medical recalls Halyard ANGIO PACK surgical kits due to loose metal flakes that could detach and enter a patient's surgical site. The recall affects 284 units distributed nationwide.

    Product
    Halyard ANGIO PACK - Medical convenience kits Model Number: WRNM047-05
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard CPT Cardiac Cath Pack Recalled Due to Loose Metal Flakes in Forceps and Clamps

    AVID Medical is recalling Halyard CPT Cardiac Cath Pack convenience kits due to small metal flakes that may detach from sponge forceps and towel clamps. Metal flakes could enter a surgical site undetected and cause local or foreign body reactions.

    Product
    Halyard CPT CARDIAC CATH PACK - Medical convenience kits Model Number: SHAN430-02
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Surgical instruments recalled for loose metal flake contamination risk

    AVID Medical is recalling Halyard RAD PACK medical convenience kits containing sponge forceps and towel clamps. Small metal flakes may detach and enter surgical sites, potentially causing foreign body or local tissue reactions.

    Product
    Halyard RAD PACK - Medical convenience kits Model Number: VANW017-04
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Pacemaker Tray Surgical Kits Recalled for Loose Metal Flakes

    AVID Medical recalls Halyard PACEMAKER TRAY surgical kits nationwide due to potential loose metal flakes that could detach from sponge forceps and towel clamps and contaminate surgical sites.

    Product
    Halyard PACEMAKER TRAY - Medical convenience kits Model Number: VMPB0003-10
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Angiography Drape Pack Recalled for Potential Metal Flakes in Surgical Site

    AVID Medical recalls Halyard angiography drape packs due to potential metal flakes that could detach from sponge forceps and towel clamps. These flakes could enter a surgical site undetected and cause local or foreign body reactions.

    Product
    Halyard ANGIOGRAPHY DRAPE PACK - Medical convenience kits Model Number: VMCC021-06
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard cardiac surgical kit recalled for loose metal flakes risk

    AVID Medical is recalling Halyard cardiac surgical kits (Model ESJH017-03) due to sponge forceps and towel clamps that may shed small metal flakes during surgery. These metal flakes could enter the patient's surgical site, potentially causing local or foreign body reactions.

    Product
    Halyard KIT, CARDIAC ROBOTIC - Medical convenience kits Model Number: ESJH017-03
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Total Knee Surgical Kit Recalled for Loose Metal Flakes

    AVID Medical is recalling Halyard total knee surgical kits due to loose metal flakes on sponge forceps and towel clamps that could enter and contaminate the surgical site. The defect poses a risk of local reactions or foreign body complications.

    Product
    Halyard KIT, TOTAL KNEE - Medical convenience kits Model Number: ESJH012-07
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Medical surgical kits recalled due to loose metal flakes

    AVID Medical is recalling Halyard TOTAL KNEE surgical kits (Model EUOR002-05) because sponge forceps and towel clamps may shed small metal flakes that could contaminate surgical sites and cause reactions.

    Product
    Halyard KIT, TOTAL KNEE - Medical convenience kits Model Number: EUOR002-05
    Category
    Distribution
    0 states