The Recall Desk

FDA (Devices) recall action 95515

AVID Medical, Inc. recalled 73 products on 2024-11-20

The FDA (Devices) publishes a recall like this one as 73 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 73 pages.

Products
73
Announced
2024-11-20
Critical
0
Units
7,259

Why these products were recalled

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–25 of 73

  • HighFDA (Devices)··2024-11-20

    Surgical procedure kits recalled due to loose metal flakes in instruments

    AVID Medical is recalling Halyard BASIC PROCEDURE PACK surgical instrument kits because sponge forceps and towel clamps may contain loose metal flakes that could enter a patient's surgical site and cause local or foreign body reactions.

    Product
    Halyard BASIC PROCEDURE PACK - Medical convenience kits Model Number: GRAN019-06
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Extremity Pack Surgical Instruments Recalled for Loose Metal Flakes

    AVID Medical recalls Halyard Extremity Pack surgical instruments (model MMOK009-03) because small metal flakes may detach from sponge forceps and towel clamps and enter patients' surgical sites. No injuries have been reported.

    Product
    Halyard EXTREMITY PACK - Medical convenience kits Model Number: MMOK009-03
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Total Knee Surgical Kit Recalled for Loose Metal Flakes

    AVID Medical is recalling Halyard total knee surgical kits due to loose metal flakes on sponge forceps and towel clamps that could enter and contaminate the surgical site. The defect poses a risk of local reactions or foreign body complications.

    Product
    Halyard KIT, TOTAL KNEE - Medical convenience kits Model Number: ESJH012-07
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard EYE PLASTIC TRAY Medical Kits Recalled for Loose Metal Flakes

    AVID Medical recalls Halyard EYE PLASTIC TRAY medical kits due to loose metal flakes in sponge forceps and towel clamps. Metal flakes may contaminate surgical sites, potentially causing foreign body reactions or local tissue damage.

    Product
    Halyard EYE PLASTIC TRAY - Medical convenience kits Model Number: SAMM054-19
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Surgical Kit Components May Shed Metal Flakes

    AVID Medical is recalling Halyard ANGIO ARTERIOGRAM PACK surgical kits due to metal flakes that may detach from forceps and clamps. Metal fragments could enter a patient's surgical site and cause local or foreign body reactions.

    Product
    Halyard ANGIO ARTERIOGRAM PACK - Medical convenience kits Model Number: MMCC004-01
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard cardiac surgical kit recalled for loose metal flakes risk

    AVID Medical is recalling Halyard cardiac surgical kits (Model ESJH017-03) due to sponge forceps and towel clamps that may shed small metal flakes during surgery. These metal flakes could enter the patient's surgical site, potentially causing local or foreign body reactions.

    Product
    Halyard KIT, CARDIAC ROBOTIC - Medical convenience kits Model Number: ESJH017-03
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Surgical kits recalled for potentially detachable metal flakes

    AVID Medical is recalling Halyard VA lumbar laminectomy surgical kits because sponge forceps and towel clamps may have loose metal flakes that could detach and enter a patient's surgical site.

    Product
    Halyard VA LUMBAR LAMINECTOMY - Medical convenience kits Model Number: VMED009-03
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Surgical instrument bundle recalled for potentially detachable metal flakes

    AVID Medical is recalling Halyard VAG hysterectomy instrument bundles due to loose metal flakes on surgical instruments that could contaminate surgical sites.

    Product
    Halyard VAG HYSTERECTOMY BUNDLE PK - Medical convenience kits Model Number: LANC022-29
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Surgical forceps and clamps recall due to potential loose metal flakes

    AVID Medical is recalling Halyard OR Endovascular Tray surgical kits due to loose metal flakes that could detach from forceps and clamps and potentially enter the surgical site.

    Product
    Halyard OR ENDOVASCULAR TRAY - Medical convenience kits Model Number: SAMM048-15
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Medical surgical kits recalled for loose metal flakes in instruments

    AVID Medical is recalling Halyard CYSTO PDS medical convenience kits due to loose metal flakes that may detach from sponge forceps and towel clamps. These flakes could enter a surgical site and cause foreign body reactions.

    Product
    Halyard CYSTO PDS - Medical convenience kits Model Number: GVVA002-05
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard SAMMC ANGIOGRAPHY Medical Kits Recalled for Loose Metal Flakes

    AVID Medical recalls Halyard SAMMC ANGIOGRAPHY medical convenience kits due to loose metal flakes on sponge forceps and towel clamps. Metal flakes may enter surgical sites, potentially causing local reactions or foreign body complications.

    Product
    Halyard SAMMC ANGIOGRAPHY - Medical convenience kits Model Number: SAMM066-15
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Medical Convenience Kits Recalled for Detachable Metal Flakes

    AVID Medical recalls Halyard TAH-BUNDLE medical kits with sponge forceps and towel clamps. Loose metal flakes may detach and enter surgical sites, potentially causing local or foreign body reactions.

    Product
    Halyard TAH-BUNDLE - Medical convenience kits Model Number: LANC039-29
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Surgical Kit Components Recalled for Potential Metal Fragment Detachment

    AVID Medical is recalling Halyard ORAL SURGERY HEAD medical convenience kits (Model SAMM029-15) due to potential loose metal flakes that could detach from forceps and clamps and enter the surgical site, posing risk of local or foreign body reactions.

    Product
    Halyard ORAL SURGERY HEAD - Medical convenience kits Model Number: SAMM029-15
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard MINOR PACK Medical Kits Recalled for Loose Metal Fragments in Instruments

    AVID Medical is recalling Halyard MINOR PACK medical kits because sponge forceps and towel clamps may shed loose metal fragments. These could enter surgical sites undetected, potentially causing local or foreign body reactions.

    Product
    Halyard MINOR PACK - Medical convenience kits Model Number: MMOK013-03
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard surgical kit recalled for loose metal flakes in forceps and clamps

    AVID Medical is recalling Halyard KIT, SPINE surgical convenience kits due to loose metal flakes on sponge forceps and towel clamps. The flakes could detach and enter a patient's surgical site, potentially causing local or foreign body reactions.

    Product
    Halyard KIT, SPINE - Medical convenience kits Model Number: EUOR004-03
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Interventional Tray Medical Kits Recalled for Loose Metal Flakes

    AVID Medical, Inc. is recalling the Halyard Interventional Tray medical kits because sponge forceps and towel clamps may shed small metal flakes. These loose flakes could enter a patient's surgical site undetected, causing local or foreign body reactions.

    Product
    Halyard INTERVENTIONAL TRAY - Medical convenience kits Model Number: PSSC004-03
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard ORTHO PACK surgical instrument kits recalled due to metal flakes

    AVID Medical is recalling Halyard ORTHO PACK kits because sponge forceps and towel clamps may shed loose metal flakes that could enter a patient's surgical site, causing local or foreign body reactions.

    Product
    Halyard ORTHO PACK -Medical convenience kits Model Number: MMOK006-04
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Angiography Pack Surgical Instruments Recalled Due to Loose Metal Flakes

    AVID Medical is recalling Halyard angiography pack surgical instruments due to loose metal flakes that could detach and enter a patient's surgical site, potentially causing foreign body reactions.

    Product
    Halyard ANGIOGRAPHY PACK - Medical convenience kits Model Number: MAMC218-18
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard IR Biopsy Pack recalled for loose metal flakes in surgical kit

    AVID Medical is recalling Halyard IR BIOPSY PACK due to loose metal flakes in forceps and towel clamps that could detach and enter the surgical site, potentially causing local or foreign body reactions.

    Product
    Halyard IR BIOPSY PACK - Medical convenience kits Model Number: AVMD019-01
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard ENDOVASCULAR surgical kits recalled for loose metal flakes

    Halyard ENDOVASCULAR Model VMED008-02 surgical convenience kits are recalled due to potential loose metal flakes in sponge forceps and towel clamps. These flakes could enter a surgical site and cause local or foreign body reactions.

    Product
    Halyard ENDOVASCULAR - Medical convenience kits Model Number: VMED008-02
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Pacemaker Drape Surgical Instruments Recalled for Loose Metal Flakes

    AVID Medical recalls Halyard Pacemaker Drape surgical instruments because metal flakes may detach from sponge forceps and towel clamps. Loose metal fragments could enter surgical sites undetected and cause local or foreign body reactions.

    Product
    Halyard PACEMAKER DRAPE - Medical convenience kits Model Number: VAST020-06
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Lower Extremity Pack surgical kits recalled for loose metal fragments

    AVID Medical is recalling Halyard Lower Extremity Pack surgical kits because sponge forceps and towel clamps may shed loose metal flakes that could enter a patient's surgical site undetected, potentially causing local or foreign body reactions.

    Product
    Halyard LOWER EXTREMITY PACK - Medical convenience kits Model Number: PANJ001-03
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Surgical Sponge Forceps and Towel Clamps Recalled for Loose Metal Flakes

    AVID Medical is recalling Halyard CATH LAB PACK surgical kits because sponge forceps and towel clamp components may shed loose metal flakes that could enter a patient's surgical site, posing a risk of local or foreign body reactions.

    Product
    Halyard CATH LAB PACK - Medical convenience kits Model Number: VANW016-05
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    Halyard Biopsy Pack surgical instruments recalled due to loose metal flakes

    AVID Medical is recalling Halyard BIOPSY PACK surgical kits due to loose metal flakes from sponge forceps and towel clamps. Metal flakes could enter a patient's surgical site, potentially causing local or foreign body reactions.

    Product
    Halyard BIOPSY PACK - Medical convenience kits Model Number: VANW019-04
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-11-20

    AVID Medical Recalls Halyard EP PACK Kits Due to Metal Flakes

    AVID Medical, Inc. is recalling Halyard EP PACK medical convenience kits because sponge forceps and towel clamps may release metal flakes that could enter a patient's surgical site.

    Product
    Halyard EP PACK - Medical convenience kits Model Number: VANW018-06
    Category
    Distribution
    0 states