FDA (Devices) recall action 94508

Philips North America Llc recalled 30 products on 2024-05-29

The FDA (Devices) publishes a recall like this one as 30 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 30 pages.

Products
30
Announced
2024-05-29
Critical
0
Units
8,631

Why these products were recalled

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

26–30 of 30

  • HighFDA (Devices)··2024-05-29

    Philips Panorama HFO MR Systems Loose Terminal Connection May Create Fire Hazard

    Philips Panorama HFO MR diagnostic imaging systems (Model 781350) contain a loose terminal connection in the mains distribution unit that may create a hotspot and cause smoke or fire. Approximately 118 units are affected worldwide.

    Product
    Panorama HFO- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781350
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Recalls Ingenia 3.0T MR Systems for Loose Mains Terminal Fire Hazard

    Philips is recalling 1,517 Ingenia 3.0T magnetic resonance (MR) systems due to a potentially loose terminal connection in the mains distribution unit that may create a hotspot and cause fire or smoke.

    Product
    Ingenia 3.0T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782103 (2) 781377 (3) 781342
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips MRI System Recalled for Potential Fire in Power Distribution Unit

    Philips is recalling 2 Achieva 1.5T MRI systems worldwide due to a potential loose connection in the power distribution unit that could cause smoke or fire in hospital technical rooms.

    Product
    Achieva 1.5T Initial system Model Number (REF): 781178
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Achieva 1.5T MR Systems Electrical Connection Fire Hazard

    Philips is recalling 1,585 Achieva 1.5T MR systems worldwide due to a loose g-MDU L3 terminal connection that may create an electrical hotspot causing fire or smoke. No injuries have been reported.

    Product
    Achieva 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 781296 (2) 781343
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Achieva 3.0T MRI System Recall: Fire Hazard Risk

    Philips recalls 386 Achieva 3.0T MRI systems due to a loose terminal connection that may create a hotspot and cause smoke or fire in the equipment room.

    Product
    Achieva 3.0T Model Number (REF): (1) 781345; (2) 781344; (3) 781278; (4) 781277; (5) 781177
    Category
    Distribution
    0 states