FDA (Devices) recall action 94508

Philips North America Llc recalled 30 products on 2024-05-29

The FDA (Devices) publishes a recall like this one as 30 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 30 pages.

Products
30
Announced
2024-05-29
Critical
0
Units
8,631

Why these products were recalled

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–25 of 30

  • HighFDA (Devices)··2024-05-29

    Philips Achieva 1.5T MR Systems: Loose Terminal Connection Fire Risk

    Philips is recalling Achieva 1.5T MR diagnostic systems worldwide due to a potential loose connection in the mains distribution unit that could create a hotspot and cause smoke or fire in hospital technical rooms.

    Product
    Achieva 1.5T Conversion- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 781283; (2) 781346
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Medical Device Recalled for Fire Hazard in Electrical Connection

    Philips is recalling the SmartPath to dStream for 3.0T medical device due to a loose electrical connection in the mains distribution unit that may cause smoke or fire in hospital technical rooms. Three units are affected.

    Product
    SmartPath to dStream for 3.0T Model Number (REF): 782145
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Ingenia Ambition X MR systems recalled for loose terminal connection fire hazard

    Philips is recalling 509 Ingenia Ambition X MR systems due to loose terminal connections in the g-MDU L3 unit that may create hotspots and cause smoke or fire in hospital technical rooms.

    Product
    Ingenia Ambition X- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 782138; (2) 782109; (3) 781356
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Recalls Ingenia 3.0T MR Systems for Loose Mains Terminal Fire Hazard

    Philips is recalling 1,517 Ingenia 3.0T magnetic resonance (MR) systems due to a potentially loose terminal connection in the mains distribution unit that may create a hotspot and cause fire or smoke.

    Product
    Ingenia 3.0T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782103 (2) 781377 (3) 781342
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Achieva 1.5T MR Systems Electrical Connection Fire Hazard

    Philips is recalling 1,585 Achieva 1.5T MR systems worldwide due to a loose g-MDU L3 terminal connection that may create an electrical hotspot causing fire or smoke. No injuries have been reported.

    Product
    Achieva 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 781296 (2) 781343
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Achieva 3.0T MRI System Recall: Fire Hazard Risk

    Philips recalls 386 Achieva 3.0T MRI systems due to a loose terminal connection that may create a hotspot and cause smoke or fire in the equipment room.

    Product
    Achieva 3.0T Model Number (REF): (1) 781345; (2) 781344; (3) 781278; (4) 781277; (5) 781177
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Magnetic Resonance System MR RT Upgrades Recalled for Fire Risk

    Philips is recalling MR RT Upgrades Magnetic Resonance diagnostic systems due to a loose terminal connection in the mains distribution unit that could cause smoke or fire. Two units with serial numbers 70980 and 71569 are affected worldwide.

    Product
    MR RT Upgrades- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781440
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    MRI System Mains Distribution Unit Connection Fire Risk

    Loose terminal connections in the mains distribution unit of Philips MRI systems may create electrical hotspots that could cause smoke or fire. The company is recalling 617 affected units worldwide.

    Product
    SmartPath to dStream for 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 782146; 782112; 781260
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Magnetic Resonance System Loose Terminal Connection May Cause Fire

    Philips MR RT imaging systems may have loose electrical connections that could cause smoke or fire in hospital technical rooms. Two units with serial numbers 45 and 47 are affected worldwide.

    Product
    MR RT - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781439
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    MR Imaging System Recall for Loose Electrical Terminal Connection Fire Risk

    Philips AchievaXR MR systems may have loose terminal connections in the power distribution unit that could create hotspots and cause smoke or fire. The FDA has recalled 35 affected units worldwide.

    Product
    AchievaXR -Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): 781253
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Sonalleve MR HIFU 3.0T diagnostic system fire risk recall

    Philips is recalling 4 Sonalleve MR HIFU 3.0T diagnostic systems worldwide due to loose electrical connections in the mains distribution unit that may create fire hazards in hospital settings.

    Product
    Sonalleve MR HIFU 3.0T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781361
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    MRI Scanner Terminal Connection Fire and Smoke Risk

    Philips Ingenia 3.0T CX MRI scanners may experience loose terminal connections in the global Mains Distribution Unit (g-MDU) that could create fire or smoke hazards. The recall affects 350 units distributed worldwide.

    Product
    Ingenia 3.0T CX Model Number (REF): (1) 782105 (2) 781271
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Evolution 1.5T MR Systems: Loose Terminal Connection Fire Risk

    Philips Evolution 1.5T MR systems may have loose g-MDU terminal connections that could create electrical hotspots and cause fires. The recall affects 56 units distributed worldwide.

    Product
    Evolution upgrade 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 782116; (2) 782148
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Intera 1.5T MRI System Mains Connection Defect Poses Fire Risk

    Philips is recalling 200 Intera 1.5T MRI systems worldwide due to loose mains distribution unit connections that may cause electrical hotspots, creating fire and smoke hazards in hospital technical rooms.

    Product
    Intera 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781295
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips MR System 1.5T Marlin MRI devices recalled due to fire hazard

    Philips is recalling 10 MR System 1.5T Marlin MRI devices worldwide due to a fire and smoke hazard. A loose electrical connection in the mains distribution unit may create a hotspot that causes smoke or fire in hospital technical rooms.

    Product
    MR system 1.5T Marlin- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781483
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Ingenia Ambition S MRI systems recalled for potential fire hazard

    Philips is recalling Ingenia Ambition S MRI systems due to a potential fire hazard. A loose terminal connection in the electrical distribution unit may create a hotspot and smoke or fire in hospital technical rooms.

    Product
    Ingenia Ambition S Model Number (REF): (1) 782139; (2) 782133; (3) 782108; (4) 781359
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Medical Device Mains Distribution Unit Recall Due to Fire Risk

    The Evolution upgrade 3.0T mains distribution unit is being recalled because its L3 terminal connection may become loose, creating a hotspot that could cause fire or smoke in hospital technical rooms.

    Product
    Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips MR 5300 magnetic resonance systems' mains distribution unit may ignite

    Philips has recalled 81 MR 5300 MRI systems worldwide due to a loose electrical connection that may overheat and cause fire or smoke in hospital technical rooms.

    Product
    MR 5300- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782152 (2) 782110
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips MRI System Recalled for Potential Fire in Power Distribution Unit

    Philips is recalling 2 Achieva 1.5T MRI systems worldwide due to a potential loose connection in the power distribution unit that could cause smoke or fire in hospital technical rooms.

    Product
    Achieva 1.5T Initial system Model Number (REF): 781178
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    MRI System Recalled Due to Loose Terminal Connection Fire Hazard

    Philips Ingenia Elition X- MRI systems have a loose terminal connection in the power distribution unit that may create a hotspot and cause smoke or fire in hospital settings. No injuries have been reported.

    Product
    Ingenia Elition X- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781358
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Panorama HFO MR Systems Loose Terminal Connection May Create Fire Hazard

    Philips Panorama HFO MR diagnostic imaging systems (Model 781350) contain a loose terminal connection in the mains distribution unit that may create a hotspot and cause smoke or fire. Approximately 118 units are affected worldwide.

    Product
    Panorama HFO- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781350
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Medical MR System Electrical Terminal Connection May Create Fire Hazard

    A Philips Achieva 3.0T for PET MR diagnostic system (Serial 43006) may have a loose mains distribution unit terminal that could overheat and cause smoke or fire.

    Product
    Achieva 3.0T for PET -Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): 781477
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Medical Imaging System Recalled Due to Fire Risk from Loose Connection

    Philips has recalled 81 Marlin 1.5T MR diagnostic systems due to a loose g-MDU terminal connection that may create a fire hazard in hospital technical rooms.

    Product
    Marlin 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781474
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips dStream MR System Terminal Connection May Become Loose and Cause Fire

    Philips is recalling its Upgrade dStream MR system due to a potential fire hazard. A loose terminal connection in the g-MDU may create a hotspot causing smoke or fire.

    Product
    Upgrade dStream - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): 782127
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-29

    Philips Ingenia 1.5T MRI machines recalled for loose power connection fire risk

    Philips is recalling 2,262 Ingenia 1.5T MRI systems worldwide because a loose terminal connection in the power distribution unit can overheat and cause fire or smoke.

    Product
    Ingenia 1.5T Model Number (REF): (1) 782140; (2) 782115; (3) 782101; (4) 781396; (5) 781341; (6) 781315
    Category
    Distribution
    0 states