The Recall Desk

FDA (Devices) recall action 94063

In2Bones, SAS recalled 2 products on 2024-03-27

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-03-27
Critical
0
Units
122

Why these products were recalled

This Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.B.S Compression screws diameter 6.5mm length 80mm. The screws of the batches concerned by the recall may have different lengths from those issued on their labels.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2024-03-27

    Medical Device Recall: Orthopedic Compression Screws with Incorrect Dimensions

    In2Bones, SAS is recalling I.B.S Compression screws due to a batch mix-up. Affected screws may have different lengths than indicated on their labels, creating a risk of improper surgical implantation.

    Product
    I.B.S¿ 6.5-C Compression screw - diam 6.5mm lg 80mm,Part Number S65 ST180
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-03-27

    I.B.S. 6.5-C Compression Screw Batch Recalled for Dimensional Mix-Up

    A batch of I.B.S compression screws may have incorrect dimensions compared to their labels due to a batch mix-up. The 59 affected units were distributed nationwide in Tennessee.

    Product
    I.B.S¿ 6.5-C Compression screw - diam 6.5mm lg 75mm,Part Number S65 ST175
    Category
    Distribution
    1 state