The Recall Desk

FDA (Devices) recall action 93656

American Contract Systems, Inc. recalled 18 products on 2024-02-07

The FDA (Devices) publishes a recall like this one as 18 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 18 pages.

Products
18
Announced
2024-02-07
Critical
0
Units
42,797

Why these products were recalled

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–18 of 18

  • HighFDA (Devices)··2024-02-07

    Custom Procedural Convenience Kits Recalled for Sterilization Validation Failure

    American Contract Systems is recalling 244 custom procedural kits (Baby PICC, Pediatric PICC, and PORT PACK) distributed nationwide. Components subjected to sterilization without proper validation documentation may have reduced functionality or elevated sterilization residuals.

    Product
    Custom procedural convenience kits and trays, general hospital use, labeled as: a) Baby PICC Kit, kit number UIBP90J; b) Baby PICC Kit, kit number UIBP90K; c) PORT PACK (PS 058330), kit number UIPO82AB; d) Pediatric PICC Kit, kit number UIPP76K;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Gastroenterological and Urological Procedure Kits Recalled for Sterilization Process Failures

    American Contract Systems is recalling AHN BIOPSY STERILE procedure kits nationwide because several components were improperly sterilized, potentially compromising device functionality and drug efficacy.

    Product
    Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: AHN BIOPSY STERILE, kit number AHBP12A
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Surgical Procedure Kits Recalled Due to Improper Sterilization

    American Contract Systems is recalling 7,881 surgical procedure kits nationwide because components were sterilized without proper validation, risking loss of functionality or elevated chemical residue.

    Product
    Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; c) MINI MINOR AGH ASC PACK, kit number AGMM60M; d) OUTPATIENT PORT ACCESS KIT…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Surgical kits for childbirth recalled for sterilization and functionality risks

    American Contract Systems is recalling C-section surgical kits that were sterilized without proper assessment. Components may have lost function or drug efficacy, or have excess sterilization residuals.

    Product
    Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) C-SECTION PACK, kit number FDCS01AA; b) C-SECTION PACK, kit number FDCS01AB; c) C-SECTION STERILE PACK, kit number HGCS68J; d) C-SECTION PACK, kit number HICS50H; e)…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Custom Surgical Procedure Kits Recalled for Sterilization Assessment Failures

    American Contract Systems has recalled 2,717 custom surgical procedure kits used in neurosurgery nationwide. Components were added without proper assessment of their suitability for the sterilization process.

    Product
    Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Urological procedural kits recalled due to sterilization defects

    American Contract Systems recalled 1,031 urological procedure kits nationwide due to sterilization process defects. Components may have damaged functionality, reduced drug efficacy, or elevated sterilization residuals.

    Product
    Custom procedural convenience kits and trays, urological, labeled as: a) UROLOGY DAVINCI PACK, kit number HGUR87G; b) CYSTO PACK, kit number LVCY20; c) URO ROBOTIC PACK, kit number LVUR44;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Custom Surgical Convenience Kits May Lack Proper Sterilization Validation

    American Contract Systems, Inc. recalls surgical convenience kits whose added components may lack proper sterilization validation, potentially resulting in loss of functionality or retained sterilant residue.

    Product
    Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) Breast Plastics Pack, kit number BBPL20D; b) MINOR PLASTIC, kit number WPMP16H
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Ophthalmic procedural kits recalled due to inadequate sterilization validation

    American Contract Systems is recalling 940 ophthalmic surgical kits nationwide due to undocumented components that may lack functionality, drug efficacy, or contain elevated sterilization residuals.

    Product
    Custom procedural convenience kits and trays, Ophthalmic use, labeled as: a) EYE PACK, kit number LVEY27; b) Laminectomy, kit number UIOL34AG
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Orthopedic surgical kits recalled for unassessed external component sterilization

    American Contract Systems recalls 5,524 orthopedic surgical kits due to unassessed sterilization of external components that may lose functionality, drug efficacy, or have excess sterilization chemical residue.

    Product
    Custom procedural convenience kits and trays, Othopedic use, labeled as: a) HIP FRACTURE, kit number AHHF15C; b) HIP FRACTURE, kit number AHHF15D; c) Hand Pack, kit number BBHP10E; d) Lower Extremity, kit number BBLE40B; e) Custom Podiatry, kit number BOCP70J; f)…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Obstetrical and Gynecological Surgical Kits Recalled for Sterilization Assessment Defects

    American Contract Systems is recalling 1,020 obstetrical and gynecological surgical kits because components were sterilized without documented assessment of their suitability to the process, potentially causing loss of functionality or drug efficacy.

    Product
    Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L; b) OB/GYN DAVINCI PACK A RG BASIN, kit number HGOB98G; c) DAVINCI GYN PACK, kit number HHDV17; d) KIT, GYNE ROBOTIC, kit number MMGR44D…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Custom Surgical Procedure Kits Recalled for Improper Sterilization Assessment

    American Contract Systems is recalling 3,196 custom surgical procedure kits because sterilized components lacked documented suitability assessment. The components may have lost functionality or accumulated excessive sterilization residuals.

    Product
    Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: a) LAP KIDNEY MODULE, kit number AGKD50L; b) ROBOTIC PROSTATE PACK - 264852, kit number ANPR34S; c) LAP CHOLE - 213388, kit number BFLC02AD; d) GENERAL LAPAROSCOPY PACK, kit…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Cardiac central line insertion kits lack documented sterilization validation

    American Contract Systems is recalling 290 cardiac procedural kits nationwide because components were subjected to sterilization without documented assessment of temperature suitability, potentially causing loss of functionality or efficacy.

    Product
    Custom procedural convenience kits and trays, cardiac, labeled as: a) CENTRAL LINE INSERTION PACK, kit number LLCE52; b) Adult Central Line Insertion Supply Kit, kit number UICL44N; c) Pediatric Central Line Insertion, kit number UIPL77K
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Custom Ear, Nose and Throat Surgical Kits Recalled Due to Sterilization Validation Failure

    American Contract Systems is recalling 804 custom ear, nose and throat surgical kits nationwide because components were sterilized without validation that they could tolerate the process, potentially affecting functionality or drug efficacy.

    Product
    Custom procedural convenience kits and trays, ear, nose & throat, labeled as: a) T AND A PACK AGH ASC, kit number AGTA48J; b) OPSC EAR PACK, kit number LLEO66; c) T AND A PACK, kit number MMTA41F
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Custom Anesthesia Kits Recalled Over Sterilization Validation Failure

    American Contract Systems recalls 182 anesthesia procedure kits due to improper sterilization assessment of added components. The kits may have compromised functionality, reduced drug efficacy, or elevated sterilant residue.

    Product
    Custom procedural convenience kits and trays, anesthesia, labeled as: a) HEART ANESTHESIA BUNDLE NS, Kit Number LMHA01V; b) OR/Anesthesia Central Line Insertion Kit, Kit Number UIAN43L
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Custom OB/GYN Procedural Kits Recalled for Sterilization Process Failure

    American Contract Systems recalled 144 custom OB/GYN procedural kits after discovering components were sterilized without proper documentation, potentially losing functionality or efficacy. Affected units were distributed nationwide.

    Product
    Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LavH, kit number BBLV21C; b) LAP HYST 247540, kit number BFHY75U
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Custom Cardiac Procedural Kits Recalled for Inadequate Sterilization Assessment

    American Contract Systems is recalling 4394 custom cardiac procedural kits due to components added without proper assessment for sterilization exposure, which may result in loss of functionality or drug efficacy.

    Product
    Custom procedural convenience kits and trays, cardiac, labeled as: a) HEART CATHETERIZATION TRAY, kit number JRHC45G; b) TAVR PACK, kit number LLTV10; c) CATH PED LF SJH, kit number SJCA21F; d) CATH HEART ADL LF SJH, kit number SJCH22F; e) INT RAD PORTS-LF - 206060…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Cardiac surgical kits recalled due to improper sterilization exposure

    American Contract Systems is recalling 12,291 cardiac procedure kits nationwide. Components were exposed to sterilization without documented assessment of suitability, potentially causing loss of functionality or efficacy.

    Product
    Procedural convenience kits and trays, cardiac, labeled as: a) EP LAB BI V IMPLANT PACK, kit number AGBI64J; b) LAB DEVICE IMPLANT, kit number AGDI40J; c) LAB DEVICE IMPLANT, kit number AGDI40K; d) MINOR VASCULAR, kit number AGMV26O; e) OPEN HEART ACCESSORY, kit…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Obstetrical Labor and Delivery Kits Recalled for Potential Functionality Loss

    American Contract Systems has recalled Labor and Delivery Packs because certain components may have lost functionality or efficacy due to unvalidated sterilization exposure.

    Product
    Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LABOR & DELIVERY PACK, kit number HILD52F; b) LABOR & DELIVERY PACK, kit number HILD52G
    Category
    Distribution
    0 states