FDA (Devices) recall action 92250

Permobil recalled 2 products on 2023-07-05

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-07-05
Critical
0
Units
25,101

Why these products were recalled

When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2023-07-05

    PushTracker E2 watch app may crash, disabling wheelchair motor controls

    Permobil's PushTracker E2 watch controls wheelchair power assist motors. The application may crash when running multiple processes, causing the motor to keep running and preventing users from stopping it using normal controls.

    Product
    PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-07-05

    PushTracker E3 software failure may prevent stopping SmartDrive motor

    A software defect in the PushTracker E3 watch can cause the SmartDrive motor to run uncontrollably, and users may not be able to stop the device using tap controls. Approximately 6,196 units worldwide are affected.