Zyno Medical LLC recalled 2 products on 2019-07-31

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2019-07-31
Critical
0
Units
—

Why these products were recalled

Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2019-07-31

    Zyno Medical Administration Sets Recalled Due to Inverted Pinch Clamp Defect

    Zyno Medical LLC is recalling administration sets where an inverted pinch clamp may cause blood to be drawn from the patient instead of delivering fluid.

    Product
    Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code: B2-70072-F
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2019-07-31

    Zyno Medical Administration Sets Recalled Due to Inverted Pinch Clamp Defect

    Zyno Medical LLC is recalling administration sets with inverted pinch clamps that may cause blood to be drawn from patients instead of delivering fluid.

    Product
    Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072
    Category
    Distribution
    1 state