The Recall Desk

FDA (Devices) recall action 82038

Philips Medical Systems Nederlands recalled 64 products on 2019-03-06

The FDA (Devices) publishes a recall like this one as 64 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 64 pages.

Products
64
Announced
2019-03-06
Critical
0
Units
718,464

Why these products were recalled

There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

26–50 of 64

  • SevereFDA (Devices)··2019-03-06

    Philips Achieva XR MRI Systems Recalled for Ceiling Speaker Hazard

    Philips is recalling 11,226 Achieva XR MRI systems because ceiling speakers may fall and be pulled by the magnetic field.

    Product
    Achieva XR 1.5T/3.0T, Model Number 781153. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-03-06

    Philips MRI System Recall Due to Falling Ceiling Speaker Hazard

    Philips is recalling 11,226 MRI systems because ceiling speakers may fall and be pulled by the magnetic field.

    Product
    SmartPath to dStream for 1.5T, Model Number 781260. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-03-06

    Philips Intera 0.5T MRI System Recalled for Falling Speaker Hazard

    Philips is recalling 11,226 Intera 0.5T MRI systems because ceiling speakers may fall due to magnetic attraction.

    Product
    Intera 0.5T Standard, Model Number 781101. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2019-03-06

    Philips Intera 1.0T MRI System Recalled for Ceiling Speaker Hazard

    Philips is recalling Intera 1.0T MRI systems because ceiling speakers may fall and be pulled by the magnetic field.

    Product
    Intera 1.0T Power/Pulsar, Model Number 781103. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2019-03-06

    Philips MRI System Recall Due to Ceiling Speaker Magnetic Hazard

    Philips is recalling Intera Achieva 1.5T Nova MRI systems because ceiling speakers may be pulled by the magnetic field and fall.

    Product
    Intera Achieva 1.5T Nova, Model Number 781172. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-03-06

    Philips Achieva MRI Systems Recalled Due to Ceiling Speaker Hazard

    Philips is recalling 11,226 Achieva 1.5T MRI systems because ceiling speakers may fall and be pulled by the magnetic field.

    Product
    Achieva 1.5T, Model Number 781296. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-03-06

    Philips Multiva MRI System Recalled for Ceiling Speaker Hazard

    Philips is recalling Multiva 1.5T MRI systems because ceiling speakers may fall due to magnetic attraction.

    Product
    Multiva 1.5T 16 R5 (ZBO), Model Number 781073. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-03-06

    Philips MRI System Recall for Falling Ceiling Speaker Hazard

    Philips is recalling Intera Achieva 1.5T Pulsar MRI systems because ceiling speakers may fall due to magnetic attraction.

    Product
    Intera Achieva 1.5T Pulsar, Model Number 781171. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-03-06

    Philips MR-OR 3.0T MRI Systems Recalled for Falling Speaker Hazard

    Philips is recalling 11,226 MR-OR 3.0T MRI systems because ceiling speakers may fall and be pulled by the magnetic field.

    Product
    MR-OR 3.0T, Model Number 781438. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-03-06

    Philips MRI System Recall Due to Falling Ceiling Speaker Hazard

    Philips is recalling Intera 1.5T MRI systems because ceiling speakers may fall and be pulled by the magnetic field.

    Product
    Intera 1.5T Omni/Stellar, Model Number 781104. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-03-06

    Philips Achieva 3.0T MRI System Recalled for Falling Speaker Hazard

    Philips is recalling 11,226 Achieva 3.0T MRI systems because ceiling speakers may fall and be pulled by the magnetic field.

    Product
    Achieva 3.0T X-series, Model Number 781277. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-03-06

    Philips Ingenia 1.5T MRI System Recalled for Falling Speaker Hazard

    Philips is recalling 11,226 Ingenia 1.5T MRI systems because ceiling speakers may fall and be pulled by the magnetic field.

    Product
    NZM101 Ingenia 1.5T Linac, Model Number 781483. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2019-03-06

    Philips MRI System Recall Due to Ceiling Speaker Magnetic Hazard

    Philips is recalling Intera 1.5T MRI systems because ceiling speakers may fall and be pulled by the magnetic field.

    Product
    Intera 1.5T Power/Pulsar, Model Number 781105. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-03-06

    Philips MRI System Recall Due to Falling Ceiling Speaker Hazard

    Philips is recalling Intera Achieva 1.5T MRI systems because ceiling speakers may fall and be pulled by the magnetic field.

    Product
    Intera Achieva 1.5T IT Nova Dual, Model Number 781176. Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states