Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

24151–24175 of 40339

  • LowFDA (Devices)·Z-1520-2018·2018-05-02

    Quidel MicroVue Intact PTH EIA Recalled for Potential Calibration Issues

    Diagnostic Hybrids, Inc. is recalling 26 units of the Quidel MicroVue Intact PTH EIA due to a potential loss of separation between calibrators that may impact assay calibration.

    Product
    Quidel MicroVue Intact PTH EIA, Model 8044. An enzyme immunoassay for the quantification of Intact Parathyroid Hormone (PTH) in serum.
    Category
    Medical Device
    Distribution
    0 states
  • LowFDA (Devices)·Z-1518-2018·2018-05-02

    CooperVision Clariti 1 Day Contact Lenses Recalled for Incorrect Expiration Date

    CooperVision is recalling specific lots of Clariti 1 Day contact lenses because the secondary packaging displays an incorrect expiration date.

    Product
    56% 1 Day Silicone Hydrogel (-2.50), 10 pack, UV Blocking, daily disposable contact lenses The Clariti 1 Day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of refractiveametropia…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • LowFDA (Devices)·Z-1479-2018·2018-05-02

    Medtronic DxTerity Diagnostic Catheter Recalled for Incorrect Box Labeling

    Medtronic is recalling 855 units of DxTerity Diagnostic Catheters because the outer box displays an incorrect 6 French size indicator, though the catheters inside are correct.

    Product
    Medtronic DxTerity Diagnostic Catheter, JL 4.0, REF DXT5JL40, marked with 6F color code on box.
    Category
    Medical Device
    Distribution
    0 states
  • LowFDA (Devices)·Z-1464-2018·2018-05-02

    Ace Surgical Supply Recalls Mispackaged Double Blade Shoe Unigraft Knives

    Ace Surgical Supply Co., Inc. is recalling six units of Double Blade Shoe F/Unigraft Knives because a single blade shoe was mistakenly packaged instead of the intended double blade.

    Product
    Double Blade Shoe F/Unigraft Knife 1.5 Thick X 9MM SKU 6500009 Used to achieve epithelial graft in the oral cavity.
    Category
    Medical Device
    Distribution
    3 states
  • LowFDA (Devices)·Z-1517-2018·2018-05-02

    CooperVision Recalls Clariti 1 Day Contact Lenses for Incorrect Expiration Date

    CooperVision is recalling Clariti 1 Day contact lenses because the secondary packaging displays an incorrect expiration date.

    Product
    56% 1 Day Silicone Hydrogel (-4.50), 10 pack, UV Blocking, daily disposable contact lenses The Clariti 1 Day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of refractiveametropia…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • LowFDA (Devices)·Z-1515-2018·2018-05-02

    CooperVision Recalls Clariti 1 Day Contact Lenses for Incorrect Expiration Date

    CooperVision is recalling Clariti 1 Day contact lenses because the secondary packaging displays an incorrect expiration date.

    Product
    clariti 1 day/Fresh Day 90 pack (-1.00), UV Blocking, 90 silicon hydrogel, daily disposable contact lenses The Clariti 1 Day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • LowFDA (Devices)·Z-1516-2018·2018-05-02

    CooperVision Recalls Clariti 1 Day Contact Lenses for Incorrect Expiration Date

    CooperVision is recalling Clariti 1 Day contact lenses because the secondary packaging displays an incorrect expiration date.

    Product
    clariti 1 day (-1.25) 90 pack, UV Blocking, 90 silicone hydrogel, daily disposable contact lenses The Clariti 1 Day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1420-2018·2018-04-25

    Autochair Milford Smartbase Recalled Due to Leg Folding Hazard

    Autochair is recalling Milford Smartbase units because a weak spring and misaligned locking pin caused legs to fold, dropping a person.

    Product
    Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-1409-2018·2018-04-25

    Philips Brilliance iCT SP CT Systems Recalled for Emergency Stop Defect

    Philips is recalling 19 Brilliance iCT SP CT systems because closing the emergency stop may disable the horizontal motor, allowing the couch to move during patient loading.

    Product
    Brilliance iCT SP - Model no. 728311 Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1460-2018·2018-04-25

    Cook Vacuum Pump Recalled for Electrical Shock and Burn Risk

    Cook Inc. is recalling 132 Cook Vacuum Pumps (K-MAR-5200-US) worldwide due to defective crimp connectors that may cause device failure, electric shock, or burns.

    Product
    Cook Vacuum Pump (Cook Aspiration Unit) K-MAR-5200-US, GPN G51607. Intended for the aspiration of body fluids and cells; in particular oocyte aspiration.
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-1461-2018·2018-04-25

    Cook Vacuum Pump Recalled for Electrical Shock and Burn Risk

    Cook Inc. is recalling 2,267 Cook Vacuum Pumps (K-MAR-5200-US) worldwide due to defective crimp connectors that may cause device failure, electric shock, or burns.

    Product
    Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-1411-2018·2018-04-25

    Philips Brilliance Big Bore CT Systems Recalled for Emergency Stop Defect

    Philips is recalling 119 Brilliance Big Bore CT systems because closing the Emergency Stop may disable the horizontal motor, allowing the couch to move during patient load or unload.

    Product
    Brilliance Big Bore (Radiology) - Model no. 728244, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1434-2018·2018-04-25

    Philips Panorama 1.0T MRI Recalled for Helium Gas Release Risk

    Philips is recalling Panorama 1.0T MRI scanners due to a potential risk of helium gas release into the examination room during a magnet quench.

    Product
    Panorama 1.0T, Model 781250 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1430-2018·2018-04-25

    Philips Multiva MRI Systems Recalled for Helium Gas Risk During Quench

    Philips is recalling 8,205 Multiva MRI systems worldwide due to a potential risk of helium gas release in the examination room during a magnet quench.

    Product
    Multiva systems, 1) Multiva 1.5T 8 R5 Model 781072, 2) Multiva 1.5T 16 R5 Model 781073, 3) Multiva 1.5T 8ch R5 based Model 781074, 4) Multiva 1.5T Model 781076, & 5) Multiva 1.5T 16ch R5 based Model 781078 Product Usage: Indicated for use as a diagnostic device. It can…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1382-2018·2018-04-25

    Draeger Infinity Acute Care System Recall for Alarm Failure

    Draeger Medical Systems is recalling Infinity Acute Care System monitoring solutions because invasive pressure alarms may not activate.

    Product
    Infinity Acute Care System (IACS) Monitoring Solution, Catalog Numbers: MS18841, MS20401, MS20407, MS20724, MS22956, MS25510, MS25520, MS25643, MS26196, MS26372, MS31818; UDI Information: 4049098054454, 4049098054447, 4049098054409, 4049098009799, 4049098009751…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1379-2018·2018-04-25

    Getinge Washer-Disinfector Accessory Recall for Potential Cart Unloading Hazard

    Getinge is recalling 170 automated loading/unloading accessories for Washer-Disinfector 86 Series units due to a risk of disinfected goods falling to the floor.

    Product
    Getinge AGS, FSLC, FSUC, RC accessory to Washer- Disinfector 86 Series UDI Info: 86 Series Washer Disinfector 07340153700246, 07340153700253, 07340153700260, 07340153700277 Getinge automated loading/unloading systems are intended to be used as an automatic…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1437-2018·2018-04-25

    Philips Ingenia 1.5T CX MRI Recall for Helium Gas Risk

    Philips is recalling Ingenia 1.5T CX MRI scanners due to a potential risk of helium gas release into the examination room during a magnet quench.

    Product
    Ingenia 1.5T CX, Models 781261 & 781262 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1436-2018·2018-04-25

    Philips MRI Software Recall Due to Helium Gas Risk During Quench

    Philips is recalling the SmarthPath to dStream conversion software for 1.5T MRI systems due to a potential risk of helium gas release during a magnet quench.

    Product
    Conversion SmarthPath to dStream for 1.5T, Model 781260 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1415-2018·2018-04-25

    Philips Brilliance 6 Slice CT System Recalled for Emergency Stop Defect

    Philips is recalling Brilliance 6 Slice CT systems because closing the Emergency Stop may disable the horizontal motor, allowing the couch to move during patient loading.

    Product
    Brilliance 6 Slice (Air) - Model no. 728256, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1391-2018·2018-04-25

    NeuroLogica Recalls Solid State X-Ray Imagers Due to Foot Injury Risk

    NeuroLogica Corporation is recalling 16 solid state x-ray imagers because the retrofit equipment poses a potential foot injury risk during backward transport.

    Product
    Solid state x-ray imager (flat panel/digital imager) Solid state x-ray imager (flat panel/digital imager) Medical
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1441-2018·2018-04-25

    Philips Multiva MRI Systems Recalled for Helium Gas Quench Risk

    Philips is recalling 8,205 Multiva MRI systems worldwide due to a potential risk of helium gas release during a magnet quench.

    Product
    Multiva systems; 1) Magnets Multiva 16 Model 781487, 2) Magnets 8 Model 781488, 3) Magnet Achieva 1.5T, Model 781492, 4) Magnet HFO, Model 781495, 5) Magnet Multiva 16, Model 781496, 6) Magnet Multiva 8, Model 781497 & 7) Magnets Ingenia 1.5T CX, Model 781498 Product…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1432-2018·2018-04-25

    Philips Intera 1.0T MRI Systems Recalled for Helium Gas Risk

    Philips is recalling 8,205 Intera 1.0T MRI systems worldwide due to a potential risk of helium gas release during a magnet quench.

    Product
    Intera 1.0T; 1) Stellar Model 781102, 2) Pular Model 781103 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1405-2018·2018-04-25

    Philips Ingenuity CT Recall for Emergency Stop Defect

    Philips is recalling 172 Ingenuity CT systems because closing the Emergency Stop may disable the horizontal motor, allowing the couch to move during patient load or unload.

    Product
    Ingenuity CT - Model no. 728326, 728325, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These…
    Category
    Medical Device
    Distribution
    Distributed nationwide