The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

8726–8750 of 40095

  • HighFDA (Devices)·Z-0322-2024·2023-11-29

    Medtronic Mo.Ma Ultra Cerebral Protection Device manifold stickers mislabeled

    Medtronic has recalled the Mo.Ma Ultra Proximal Cerebral Protection Device due to mislabeled manifold stickers. Both sides of affected units were labeled identically, potentially causing confusion about which ports control the proximal versus distal balloons.

    Product
    Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130068X5
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0332-2024·2023-11-29

    Rhinolaryngoscope Models Recalled for Unvalidated Air Drying Process

    Aizu Olympus rhinolaryngoscopes (models ENF-T3, ENF-VT2, ENF-VT3) are recalled. The air-drying process for the instrument channel was not validated, and water remaining in some devices after repair poses contamination and infection risk.

    Product
    Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-VT3.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0327-2024·2023-11-29

    Bronchoscopes Recalled Due to Inadequate Air Drying Process

    Aizu Olympus recalls bronchoscopes due to unvalidated air drying process. Some returned units after repair contain water in channels, creating infection risk.

    Product
    Bronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190, BF-MP190F, BF-MP60, BF-P150, BF-P190, BF-P60, BF-Q170, BF-Q190, BF-XP190, BF-XP60, BF-XT160, BF-XT
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0358-2024·2023-11-29

    Blood Gas Analyzer Cartridges Recalled for Potential Oxygen Measurement Errors

    The FDA is recalling GEM Premier 3000 PAK cartridges manufactured by Instrumentation Laboratory because they may incorrectly report high oxygen levels in patients with low blood oxygen, potentially delaying critical treatment.

    Product
    GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-0399-2024·2023-11-29

    ORION Robotic Arm Recalled for Sudden Acceleration and Collision Risk

    The FDA recalls ORION robotic positioning arms used in radiation and radiology due to sudden acceleration or unintended movement that could cause patient collision with treatment room equipment. No injuries reported.

    Product
    ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation Therapy, Radiology, or Radiology image-guided treatment environment. Part Number: 1SOPT00009
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0352-2024·2023-11-29

    Roche Tina-quant D-Dimer Test System Recalled for Inaccurate Results

    Roche is recalling Tina-quant D-Dimer Test System kits because they produce elevated results when used with lithium-heparin plasma samples, potentially affecting diagnostic accuracy.

    Product
    Tina-quant D-Dimer Test System, Fibrinogen/Fibrin Degradation Products Assay, Material Numbers 03001245322 - (cobas c 311, 501 and 502; INTEGRA 400 plus), 08105618190 - (cobas c 303 and 503)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0373-2024·2023-11-29

    Biomet Compress Device Segmental Anchor Plugs Recalled for Metal Burrs

    Biomet has recalled certain Compress Device Segmental Anchor Plugs (16 MM) due to metal burrs in the transverse holes. These burrs may prevent the drill or pins from passing through, potentially extending surgical time.

    Product
    Compress Device Segmental Anchor Plug, 16 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178406
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0383-2024·2023-11-29

    Biomet Compress Device Anchor Plug Recalled for Metal Burrs

    Biomet is recalling Compress Device Short Anchor Plugs (20 MM) because metal burrs in transverse holes prevent surgical instruments from passing through, potentially extending surgery time.

    Product
    Compress Device Short Anchor Plug, 20 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178562
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0345-2024·2023-11-29

    Ultrasonic Gastrovideoscope Model TGF-UC180J Recalled for Inadequate Channel Drying

    Aizu Olympus is recalling ultrasonic gastrovideoscopes due to an unvalidated channel drying process that left some scopes with water in the channels after repair. Water remaining in the device poses a risk of contamination and infection.

    Product
    Ultrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding organs Model Number: TGF-UC180J.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0361-2024·2023-11-29

    Ventlab AirFlow Manual Resuscitators Recalled for CO2 Rebreathing Risk

    SunMed Holdings is recalling Ventlab AirFlow manual resuscitators nationwide due to a backwards leak in the integrated manometer that allows CO2 rebreathing. Users should immediately discontinue use and contact the manufacturer.

    Product
    Ventlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult Resuscitator REF #s: AF1140MB-M5, AF1140MB-D5, AF5142OB, AF5142MB, AF5142MB-D235, AF1142MBP, AF5142MBP, AF5140MB, AF5140MB-2, AF5140MB-D6
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0320-2024·2023-11-29

    Medtronic Mo.Ma Ultra cerebral protection device labeling error recalled

    Medtronic is recalling 2,767 units of the Mo.Ma Ultra Proximal Cerebral Protection Device due to labeling errors on the manifold that result in incorrect identification of inflation and deflation ports.

    Product
    Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0388-2024·2023-11-29

    Biomet Compress Device Short Anchor Plug recalled due to metal burrs

    Biomet is recalling Compress Device Short Anchor Plugs (28 MM, Item 178570) due to potential metal burrs in transverse holes that may prevent surgical instruments from passing through, extending surgery duration.

    Product
    Compress Device Short Anchor Plug, 28 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178570
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0382-2024·2023-11-29

    Biomet Compress Device Anchor Plug Recall - Metal Burrs May Extend Surgery

    Biomet is recalling Compress Device Short Anchor Plugs due to metal burrs that may prevent surgical instruments from passing through, potentially extending operative time.

    Product
    Compress Device Short Anchor Plug, 18 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178560
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0389-2024·2023-11-29

    Philips Achieva 1.5T MRI systems recalled for potential gradient coil fire risk

    Philips is recalling Achieva 1.5T MRI systems due to potential gradient coil failure that could produce smoke and/or fire. Approximately 520 units have been distributed nationwide and internationally.

    Product
    Achieva 1.5T
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0325-2024·2023-11-29

    IRON assay kits recalled for inaccurate diagnostic results

    Sentinel CH SpA is recalling IRON assay kits (Reference Numbers 6K95-41 and 6K95-30) that may produce falsely elevated results. Affected lots could cause inaccurate iron measurements and delays in patient test reporting.

    Product
    IRON assay, Reference Numbers 6K95-41 and 6K95-30
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0350-2024·2023-11-29

    BD Gravity Infusion Sets Labeled DEHP-Free But Contain DEHP

    BD gravity blood infusion sets labeled as free of DEHP (di(2-ethylhexyl) phthalate) actually contain the substance. The recall affects products distributed worldwide through July 31, 2026.

    Product
    BD gravity sets (Product Name, Catalog #) 32-IN GRAV SET W/0.2 MICRON FILTER BAG ACCESS PORT SS VALVE TEXIUMTM LOW ABSORPTION, Catalog # 4030B-07T and BD GRAVITY BLOOD SET 15 DROP 180 MICRON FILTER SmartSiteTM VALVE NONVENTED, Catalog # 42081E.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0397-2024·2023-11-29

    EXALT Model D Single-Use Duodenoscope Recalled for Lens Fluid Ingress

    Boston Scientific is recalling EXALT Model D Single-Use Duodenoscopes due to fluid ingress in the lens, which causes poor image quality during procedures.

    Product
    EXALT Model D Single-Use Duodenoscope, 2 pack: Outer box UPN# M00542421, Inner box UPN # M00542420 & Outer box UPN# M0054242CE1, Inner box UPN # M0054242CE0.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0390-2024·2023-11-29

    Philips Achieva 1.5T MRI Device Recalled for Gradient Coil Fire Risk

    Philips is recalling 68 Achieva 1.5T MRI machines due to potential gradient coil component failure that could produce smoke and fire. No injuries have been reported.

    Product
    Achieva 1.5T Conversion
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0371-2024·2023-11-29

    Compress Device Anchor Plugs recalled due to metal burrs

    Biomet is recalling Compress Device Segmental Anchor Plugs due to metal burrs in transverse holes that may prevent drill or pins from passing through, potentially extending surgical time.

    Product
    Compress Device Segmental Anchor Plug, 12 MM -For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178402
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0351-2024·2023-11-29

    BD Extension Sets Mislabeled as DEHP-Free Despite Containing Phthalates

    Certain BD Extension Sets sold as free of DEHP (di(2-ethylhexyl) phthalate) actually contain this restricted phthalate. The products have been distributed worldwide.

    Product
    BD Extension Sets (Product Name, Catalog #) BD SmartSite Filter Extension Set PE lined Standard bore 1 SmartSite bonded 0.2um Large PES Filter 1 PC RLL. L: 16 in PV: 4mL, Catalog # 10013902 BD EXTENSION SET TRI-PORT SMALLBORE TUBING CHECK VALVE 3 SmartSite" VALVES, Catalog
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0329-2024·2023-11-29

    Colonoscopes Recalled for Unvalidated Channel Drying Process and Infection Risk

    Aizu Olympus is recalling certain colonoscope models because the channel air drying process was not validated. A small percentage of repaired units had water remaining in the channel, which could lead to infection risk.

    Product
    Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I, CF-2T160L, CF-H170I, CF-H170L, CF-H180AI, CF-H180AL, CF-H180DI, CF-H180DL, CF-H190I, CF-H190L, CF-
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0393-2024·2023-11-29

    Philips MRI systems recalled for potential gradient coil fire hazard

    Philips is recalling 102 Intera 1.5T MRI systems worldwide due to a potential gradient coil failure that could produce smoke and fire. The affected devices are identified by product numbers 781195 and 781295.

    Product
    Intera 1.5T
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0336-2024·2023-11-29

    Hysterovideoscope Models Recalled Due to Improper Channel Drying Risk

    Aizu Olympus is recalling Hysterovideoscope Model HYF-V units (27 devices) because the channel air drying process was not validated. Water remaining in device channels could pose an infection risk.

    Product
    Hysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0384-2024·2023-11-29

    Compress Device Short Anchor Plugs Recalled for Metal Burrs

    Biomet is recalling Compress Device Short Anchor Plugs (22 MM) used in orthopedic surgery because they may contain metal burrs in the transverse holes that could prevent drill or pin passage and extend surgical procedures.

    Product
    Compress Device Short Anchor Plug, 22 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178564
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0328-2024·2023-11-29

    Ultrasonic Bronchoscopes Recalled Due to Inadequate Drying Validation

    Aizu Olympus is recalling 823 ultrasonic bronchoscopes (models BF-UC180F, BF-UC190F) because the channel air drying process was not validated, and some repaired units retained water that could cause contamination and infection.

    Product
    Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.
    Category
    Medical Device
    Distribution
    Distributed nationwide