Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

34126–34150 of 40489

  • ModerateFDA (Devices)·Z-0599-2015·2014-12-24

    Stryker Laparoscopic Knot Pushers Recalled for Sterilization Labeling Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic knot pushers because the instructions for use do not support specific sterilization methods.

    Product
    PKG, KNOT PUSHER 5MM OPEN, P/N 0250080132 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0735-2015·2014-12-24

    Stryker Endoscopy Recalls 5mm Right Angle Forceps Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of 5mm Right Angle Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 45CM P/N 0250080729 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0675-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BLUNT FORCEPS, BLUNT, SERRATED, SPOON, P/N 0250080313. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0652-2015·2014-12-24

    Stryker Maryland Forceps Recalled Due to Sterilization Parameter Errors

    Stryker Endoscopy is recalling Maryland Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MARYLAND FORCEPS, CURVED LEFT, P/N 0250080282. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0730-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Issues

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, ASPIRATION NEEDLE, 33CM, P/N 0250080702 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0737-2015·2014-12-24

    Stryker Endoscopy Recalls Fenestrated Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 fenestrated forceps units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM, 45CM, INSERT, FENESTRATED FORCEPS, P/N 0250080732 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0792-2015·2014-12-24

    Stryker Endoscopy Recalls 180,573 Laparoscopic Graspers Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 laparoscopic atraumatic graspers because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, ATRAUMATIC GRASPER, DOUBLE ACTION, 20CM, P/N 0250282100 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0776-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Scissors Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, PERITONEAL SCISSORS, 29CM, P/N 0250282048 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0611-2015·2014-12-24

    Stryker Endoscopy Recalls Fixable Cone for 11mm Hasson Due to Sterilization Labeling

    Stryker Endoscopy is recalling 180,573 units of the Fixable Cone for 11mm Hasson because the instructions for use do not support specific sterilization methods.

    Product
    PKG, FIXABLE CONE FOR 11MM HASSON, P/N 0250080180. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0631-2015·2014-12-24

    Stryker Pennington Forceps Recalled Due to Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 Pennington Forceps units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, PENNINGTON FORCEPS, P/N 0250080242. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0798-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Instruments Due to Sterilization Labeling Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, CANNULA 8MM AUTOMATIC VALVE STOP COCK, W/O TROCAR, P/N 0250080172 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0633-2015·2014-12-24

    Stryker Laparoscopic Claw Forceps Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic claw forceps units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, CLAW FORCEPS, DOUBLE ACTION, P/N 0250080245. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0773-2015·2014-12-24

    Stryker Endoscopy Recalls 3mm PEEK Monopolar Handle Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of 3mm PEEK Monopolar Handles because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM PEEK MONOPOLAR HANDLE, P/N 0250282045 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0616-2015·2014-12-24

    Stryker Laparoscopic Reducers Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic reducers because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 12.5/11MM - 5.5 REDUCER, P/N 0250080204. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0747-2015·2014-12-24

    Stryker Endoscopy Recalls Allis Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of Allis Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM X 45CM INSERT, ALLIS FORCEPS, DOUBLE ROW TEETH, P/N 0250080754 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0605-2015·2014-12-24

    Stryker Laparoscopic Cannula Recall Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic cannulas because the instructions for use do not support specific sterilization methods.

    Product
    PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O TROCAR, P/N 0250080171. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0823-2015·2014-12-24

    Biomet Sirius Cemented Hip Stem Recall Due to Packaging Mix-up

    Biomet is recalling eight Sirius Polish Cemented Hip Stems because parts were mixed up prior to packaging and shipment.

    Product
    Sirius Polish Cemented Stem 38C Catalog Number: 51-199335 Lot Number Identification: 141680, 180920 The Biomet's Sirius Femoral Hip Stem is a double-tapered, highly polished cemented hip stem, designed for either primary or revision surgery. A distal centralizer is provided…
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-0742-2015·2014-12-24

    Stryker Endoscopy Recalls Park Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling specific Park Forceps instruments because the instructions for use do not support certain sterilization methods.

    Product
    PKG, 5MM INSERT, 45CM, PARK FORCEPS, P/N 0250080745 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0831-2015·2014-12-24

    Integra Licox Brain Tissue Oxygen Monitoring Catheters Recalled for Inaccurate Values

    Integra LifeSciences is recalling 2,506 Licox brain tissue oxygen monitoring catheters because they contain a smart card programmed with an inaccurate lkN2 value.

    Product
    Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX¿ Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX¿ Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX¿ Complete Brain Tunnelling Probe Kit (comprised of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0820-2015·2014-12-24

    Carestream DirectView CR Software Recalled for Reduced Mammographic Image Quality

    Carestream Health is recalling 59 units of DirectView CR Software due to reduced image quality when printing true size multi-format images.

    Product
    CARESTREAM DIRECTVIEW CR Software Generates digital mammographic images that can be used for screening and diagnosis of breast cancer.
    Category
    Medical Device
    Distribution
    3 states
  • ModerateFDA (Devices)·Z-0606-2015·2014-12-24

    Stryker Maryland Forceps Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling Maryland Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MARYLAND FORCEPS, CURVED, P/N 0250080009. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0796-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling

    Stryker Endoscopy is recalling 180,573 laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM X 29CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW, P/N 0250282105 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0654-2015·2014-12-24

    Stryker Endoscopy Recalls Tissue Grasper Forceps Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of tissue grasper forceps because the instructions for use do not support two specific sterilization methods.

    Product
    PKG, GRASPER FORCEPS, TISSUE, P/N 0250080284. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0600-2015·2014-12-24

    Stryker Myoma Drill Recall Due to Sterilization Instructions

    Stryker Endoscopy is recalling Myoma Drill instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MYOMA DRILL 5MM LENGTH 320MM, P/N 0250080133 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0740-2015·2014-12-24

    Stryker Dorsey Grasper Forceps Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling Dorsey Grasper Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM, 45CM, INSERT, DORSEY GRASPER FORCEPS, P/N 0250080736 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide

Looking for a different category? Browse all recalls.