Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

33851–33875 of 40489

  • ModerateFDA (Devices)·Z-0910-2015·2015-01-14

    McKesson Cardiology ECG Management Software Recalled for Incorrect Time Stamps

    McKesson Israel Ltd. is recalling McKesson Cardiology ECG Management software versions 13.1 and 13.1.1 because a software error causes incorrect ECG procedure time stamps.

    Product
    McKesson Cardiology ECG Management It is a software application designed to import, display, store, analyze, distribute and manage information related to ECG procedures of adult and pediatric patients from external ECG devices.
    Category
    Medical Device
    Distribution
    6 states
  • ModerateFDA (Devices)·Z-0943-2015·2015-01-14

    Centurion Nerve Block Tray Recalled for Human Hair Contamination

    Centurion Medical Products is recalling 480 Nerve Block Trays due to a confirmed report of human hair embedded in a stopper.

    Product
    Nerve Block Tray (Nerve Block convenience kit)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0930-2015·2015-01-14

    Steris Recalls Harmony vLED Surgical Lighting Systems Due to Inconsistent Illumination

    Steris is recalling 140 Harmony vLED Surgical Lighting Systems because LED modules may illuminate inconsistently. The units were distributed worldwide.

    Product
    Harmony¿ vLED Surgical Lighting System, one unit packaged per box The Harmony¿ vLED Surgical Lighting System is a variable pattern, variable intensity surgical lighting fixture designed to provide visible illumination of the surgical field or the patient for the operating room…
    Category
    Medical Device
    Distribution
    25 states
  • ModerateFDA (Devices)·Z-0941-2015·2015-01-14

    CSA Medical truFreeze System Recalled for Timer Sound Loss

    CSA Medical is recalling 46 truFreeze cryogenic surgical devices because the timer may lose its audible sound, potentially affecting surgical procedures.

    Product
    CSA Medical truFreeze System; Model: CC3-01, a cryogenic surgical device ***The involved units are those with the newest version of a panel PC referred to as POC-127. These panels are identified by serial number. *** The truFreeze is intended for cryogenic destruction of…
    Category
    Medical Device
    Distribution
    25 states
  • ModerateFDA (Devices)·Z-0946-2015·2015-01-14

    Siemens Recalls Dimension Vista Myoglobin Calibrator Due to Positive Result Shift

    Siemens Healthcare Diagnostics is recalling Dimension Vista Myoglobin Calibrators because they may cause test results to shift up to 12% higher than expected.

    Product
    Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic product for the calibration of Myoglobin (MYO) method on the Dimension Vista System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0938-2015·2015-01-14

    Stryker Orthobiologics Recalls Imbibe Bone Marrow Aspiration Needles

    Stryker Orthobiologics is recalling Imbibe Bone Marrow Aspiration Needles due to a potential breach in the inner or outer packaging pouches.

    Product
    Stryker Orthobiologics Imbibe Bone Marrow Aspiration Needle, Model Numbers: 2090-0027 Needle (bullet-tip), 11 gauge x 4 inch 2090-0028 Needle (bullet-tip), 11 gauge x 6 inch 2090-0029 Needle (bullet-tip), 8 gauge x 6 inch 2090-0047 Needle (bullet-tip), 8 gauge x 8 inch…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0937-2015·2015-01-14

    Stryker Orthobiologics Imbibe Aliquot Needle Bone Cement Needle Recall

    Stryker Orthobiologics is recalling Imbibe Aliquot Needle Bone Cement Needles due to potential breaches in the inner or outer packaging pouches.

    Product
    Stryker Orthobiologics Imbibe Aliquot Needle Bone Cement Needle- Model Numbers: 2110-0505 Beveled Needle, 11 gauge x 4 inch 2110-0524 Beveled Needle, 11 gauge x 6 inch 2110-0506 Diamond Needle, 11 gauge x 4 inch 2110-0529 Diamond Needle, 11 gauge x 6 inch Intended Use: Can…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0907-2015·2015-01-14

    Spinal Elements Recalls Ti-Bond Marketing Literature Due to Misleading Claims

    Spinal Elements is recalling Ti-Bond marketing literature because the content regarding the Ti-Bond coating is unsubstantiated and misleading for spine applications.

    Product
    Ti-Bond Marketing Literature/Labeling, Catalog No. LT10000, MP-14000 This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months non-operative treatment prior to being treated with this device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0932-2015·2015-01-14

    Baxter Amia Automated PD Systems Recalled for Battery Charging Error

    Baxter Healthcare is recalling 55 Amia Automated PD systems due to a battery charging error that prevents the device from functioning.

    Product
    Baxter Amia Automated PD systems are used in the treatment of adult renal failure patients to perform automated peritoneal dialysis.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0889-2015·2015-01-07

    Roche Recalls ACCU-CHEK Tender II Infusion Sets Due to Tubing Detachment

    Roche Diagnostics is recalling ACCU-CHEK Tender II infusion sets because the tubing may detach, interrupting insulin delivery without an alarm.

    Product
    ACCU-CHEK Tender II 13/60 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0896-2015·2015-01-07

    Hansen Medical Recalls Sensei X Robotic Catheter Systems Due to Loose Clamp

    Hansen Medical is recalling Sensei X Robotic Catheter Systems because a mechanical clamp may become loose, causing the device to fall and potentially injure patients or users.

    Product
    Sensei X Robotic Catheter System, model numbers 02057 and 08257. - Set up joint (SUJ)mechanical clamp that secures the SUJ to the catheter laboratory table. Product Usage: The Sensei X Robotic Catheter System is an electromechanical software controlled system designed to…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0893-2015·2015-01-07

    Roche Recalls ACCU-CHEK Tender II Infusion Sets Due to Tubing Detachment

    Roche is recalling ACCU-CHEK Tender II infusion sets because tubing may detach, interrupting insulin delivery without an alarm.

    Product
    ACCU-CHEK Tender II 17/80 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0899-2015·2015-01-07

    Smiths Medical Recalls Hypodermic Needles Due to Safety Sheath Failure

    Smiths Medical ASD, Inc. is recalling 10,000 units of Needle Pro Edge Safety Hypodermic Needles because the needle may not be captured in the safety sheath.

    Product
    Hypodermic Needle Pro¿ Edge" Safety Device with Syringe 20G x1 5mL Luer lock syringe; Reorder Number: 452010 Product Usage: The Needle Pro¿ Edge Safety Hypodermic Needle is intended for injection or aspiration of fluids utilizing a luer lock or luer slip syringe.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0886-2015·2015-01-07

    Roche Recalls ACCU-CHEK Tender I Infusion Sets Due to Tubing Detachment Risk

    Roche Diagnostics is recalling 13,041 boxes of ACCU-CHEK Tender I infusion sets because the tubing may detach, interrupting insulin delivery without an alarm.

    Product
    ACCU-CHEK Tender I 17/60 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0901-2015·2015-01-07

    Nonin Medical Recalls Regional Oximetry Sensors Due to Signal Loss

    Nonin Medical is recalling specific regional oximetry sensors because they may lose signal when used on diaphoretic patients. No injuries or illnesses have been reported.

    Product
    SenSmart, Regional Oximetry EQUANOX Classic Plus Adult/Pediatric Sensor > 40 kg., Model 8003CA (REF 9824-001, 3-Wavelength), 8004CA (REF 9705-001, 4-Wavelength), with associated materials (INT-100 cable, operator's manual; and Quick Start Guide). Sensors are single-packed…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0892-2015·2015-01-07

    Roche Recalls ACCU-CHEK Tender II Infusion Sets Due to Tubing Detachment

    Roche is recalling ACCU-CHEK Tender II infusion sets because tubing may detach, interrupting insulin delivery without an alarm.

    Product
    ACCU-CHEK Tender II 17/60 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0885-2015·2015-01-07

    Roche Recalls ACCU-CHEK Tender I Infusion Sets Due to Tubing Detachment Risk

    Roche Diabetes Care is recalling ACCU-CHEK Tender I infusion sets because tubing may detach, interrupting insulin delivery without an alarm.

    Product
    ACCU-CHEK Tender I 13/110 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0884-2015·2015-01-07

    Roche Recalls ACCU-CHEK Tender I Infusion Sets Due to Tubing Detachment Risk

    Roche Diagnostics is recalling ACCU-CHEK Tender I infusion sets because the tubing may detach, interrupting insulin delivery without an alarm.

    Product
    ACCU-CHEK Tender I 13/80 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0891-2015·2015-01-07

    ACCU-CHEK Tender II Infusion Sets Recalled for Tubing Detachment Risk

    Roche Diagnostics is recalling ACCU-CHEK Tender II infusion sets because tubing may detach, interrupting insulin delivery without an alarm.

    Product
    ACCU-CHEK Tender II 13/110 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0890-2015·2015-01-07

    ACCU-CHEK Tender II Infusion Sets Recalled for Tubing Detachment Risk

    Roche Diagnostics is recalling ACCU-CHEK Tender II infusion sets because tubing may detach, interrupting insulin delivery without an alarm.

    Product
    ACCU-CHEK Tender II 13/80 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0887-2015·2015-01-07

    Roche Recalls ACCU-CHEK Tender I Infusion Sets Due to Tubing Detachment Risk

    Roche Diagnostics is recalling ACCU-CHEK Tender I infusion sets because the tubing may detach, interrupting insulin delivery without an alarm.

    Product
    ACCU-CHEK Tender I 17/80 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0883-2015·2015-01-07

    Roche Recalls ACCU-CHEK Tender I Infusion Sets Due to Tubing Detachment Risk

    Roche Diagnostics is recalling ACCU-CHEK Tender I infusion sets because the tubing may detach, interrupting insulin delivery without an alarm.

    Product
    ACCU-CHEK¿ Tender I 13/60 10 pieces Product Usage: Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0876-2015·2015-01-07

    Bard Caliber Inflation Device Recalled for Pressure Limitations

    Perouse Medical is recalling Bard Caliber inflation devices that may fail to reach required pressure, potentially delaying angioplasty procedures.

    Product
    Bard Caliber (CL3030) Inflation Device Product Usage: Inflation Device intended for use during cardiovascular to create and maintain, monitor pressure in the balloon catheter - balloon inflation device for angioplasty procedures
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0898-2015·2015-01-07

    Smiths Medical Recalls Hypodermic Needles Due to Safety Sheath Failure

    Smiths Medical ASD, Inc. is recalling specific lots of Needle Pro Edge safety hypodermic needles because the needle is not captured in the safety sheath.

    Product
    Hypodermic Needle Pro¿ Edge" Safety Device with Syringe 20G x1 10mL Luer lock syringe; Reorder Number: 4102010 Product Usage: The Needle Pro¿ Edge Safety Hypodermic Needle is intended for injection or aspiration of fluids utilizing a luer lock or luer slip syringe.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0895-2015·2015-01-07

    Atricure CryoICE Cryogenic Probe Recalled for Potential Sterility Compromise

    Atricure Inc is recalling 19,190 CryoICE cryogenic probes due to packaging defects that may have compromised sterility.

    Product
    Cryogenic probe for cardiac ablation surgery, packaged individually in Tyvek pouches Product Usage: Cryogenic probe for cardiac ablation surgery. The CryoICE cryoablation probe is a sterile, single use device intended to use in the cryosurgical treatment of cardiac…
    Category
    Medical Device
    Distribution
    Distributed nationwide

Looking for a different category? Browse all recalls.