  • ModerateFDA (Devices)·Z-0674-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, SCHERTEL FORCEPS, FENESTRATED, P/N 0250080312. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0639-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, PKG, HOOK SCISSORS, P/N 0250080260. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0625-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 laparoscopic manual instruments because the provided instructions for use do not support specific sterilization methods.

    Product
    PKG, AXIAL HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080235. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0700-2015·2014-12-24

    Stryker Endoscopy Recalls 10mm Clip Applying Forceps Due to Sterilization Documentation

    Stryker Endoscopy is recalling 180,573 units of 10mm Clip Applying Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 10MM CLIP APPLYING FORCEPS, P/N 0250080558. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0778-2015·2014-12-24

    Stryker Endoscopy Recalls Fundus Grasper Instruments Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 Fundus Grasper instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, FUNDUS GRASPER, 29CM, P/N 0250282050 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0749-2015·2014-12-24

    Stryker Laparoscopic Dissector Recalled for Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 laparoscopic dissectors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM X 33CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW, P/N 0250080757 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0779-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, FUNDUS GRASPER, 29CPKG, 3MM INSERT, ALLIGATOR GRASPER, 29CM, P/N 0250282051 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0687-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Grasper Forceps Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic grasper forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, GRASPER FORCEP, BOWEL, P/N 0250080324. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0721-2015·2014-12-24

    Stryker Endoscopy Recalls 10mm PEEK Monopolar Handle Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of 10mm PEEK Monopolar Handles because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 10MM PEEK MONOPOLAR HANDLE, P/N 0250080618 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0635-2015·2014-12-24

    Stryker Endoscopy Recalls Serrated Operating Scissors Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of Serrated Operating Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, OPERATING SCISSORS, SERRATED, P/N 0250080247. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0691-2015·2014-12-24

    Stryker Laparoscopic Needle Holder Recall Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic needle holders because the instructions for use do not support specific sterilization methods.

    Product
    PKG, NEEDLE HOLDER, CURVED RIGHT, P/N 0250080342. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0724-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled Due to Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 33CM, P/N 0250080694 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0769-2015·2014-12-24

    Stryker Endoscopy Recalls Maryland Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Maryland Forceps instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, MARYLAND, FORCEPS CURVED, 20CM, P/N 0250282011 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0753-2015·2014-12-24

    Stryker Johann Forceps Recalled for Sterilization Instructions Error

    Stryker Endoscopy is recalling Johann Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM X 33CM INSERT, JOHANN FORCEPS, P/N 0Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0716-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, ALLIS SINGLE ROW TEETH, P/N 0250080591 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0775-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Scissors Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of Metzenbaum scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, METZENBAUM, CURVED, LONG SCISSORS, 29CM, P/N 0250282047 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0788-2015·2014-12-24

    Stryker Laparoscopic Needleholder Recall Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic needleholders because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, STRAIGHT NEEDLEHOLDER, 29CM, P/N 0250282061 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0688-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the provided instructions for use do not support specific sterilization methods.

    Product
    PKG, MICRO GRASPER, P/N 0250080325. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0645-2015·2014-12-24

    Stryker Mini Metzenbaum Scissors Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Mini Metzenbaum Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MINI METZENBAUM SCISSORS, CURVED LONG, P/N 0250080266. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0644-2015·2014-12-24

    Stryker Metzenbaum Scissors Recalled Due to Sterilization Instructions Error

    Stryker Endoscopy is recalling Metzenbaum Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, METZENBAUM SCISSORS, CURVED LONG, P/N 0250080265. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0729-2015·2014-12-24

    Stryker Endoscopy Recalls Endoloop Applicator Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Endoloop Applicator units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, ENDOLOOP APPLICATOR, P/N 0250080701 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0636-2015·2014-12-24

    Stryker Maryland Forceps Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling Maryland Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MARYLAND FORCEPS, CURVED, P/N 0250080248. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-0299-2015·2014-12-24

    Zydus Recalls Benzonatate Capsules Due to Wet or Leaking Capsules

    Zydus Pharmaceuticals is recalling 58,920 bottles of Benzonatate Capsules, 200 mg, because some capsules are wet or leaking.

    Product
    Benzonatate Capsules, USP, 200 mg, 100-count bottles, Rx only, Manufactured by: Cadila Healthcare Ltd, Ahmedabad, India; Distributed by: Zydus Pharmaceuticals USA, In, Pennington, NJ USA 08534; NDC 68382-248-01, UPC 3 68382 24801 5.
    Category
    Drug
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0619-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, LUERLOCK, FEMALE, 6MM TUBING, P/N 0250080208. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0762-2015·2014-12-24

    Stryker Endoscopy Recalls 180,573 Alligator Graspers Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of 3mm Alligator Graspers because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, ALLIGATOR GRASPER, 20CM, P/N 0250282004 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide