72,919 recalls have nationwide distribution and so reach Vermont. 1,983 additional recalls listed Vermont specifically in their distribution scope.

What the numbers show

Total
1,983
Critical
199
Severe
409
Most recent
2026-09-02

Of the 1,983 recalls naming Vermont we have scored against our rubric, 199 (10%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 409 (21%) are Severe.

These figures cover only the 1,983 recalls whose distribution notice names Vermont specifically. The 72,919 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

2501–2525 of 74902

  • HighFDA (Devices)·Z-1679-2026·2026-04-08

    Philips Azurion 5 M12 X-ray may not start from foot switch

    Philips is recalling the Azurion 5 M12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 5 M12; System Code: (1)722227, (2)722231;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1659-2026·2026-04-08

    Philips Allura Xper FD10C X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD10C because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD10C; System Code: 722001;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1676-2026·2026-04-08

    Philips Azurion 7 B20 X-ray may not start from foot switch

    Philips is recalling the Azurion 7 B20 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1741-2026·2026-04-08

    MiniMed 780G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 780G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1667-2026·2026-04-08

    Philips Allura Xper FD10 OR Table X-ray may not start reliably

    Philips is recalling the Allura Xper FD10 OR Table because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1668-2026·2026-04-08

    Philips Allura Xper FD20/10 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD20/10 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/10; System Code: 722029;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1666-2026·2026-04-08

    Philips Allura Xper FD20 Biplane OR Table X-ray may not start

    Philips is recalling the Allura Xper FD20 Biplane OR Table because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020, (2)722025;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1675-2026·2026-04-08

    Philips Azurion 7 B12 X-ray may not start from foot switch

    Philips is recalling the Azurion 7 B12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1685-2026·2026-04-08

    Artoura textured tissue expander infusion set needles may be blunt

    Mentor is recalling textured Artoura Breast Tissue Expanders with integral injection dome because the infusion sets may contain a dull or blunt needle tip.

    Product
    Artoura Breast Tissue Expanders Reference Numbers: TEXP110RH TEXP120RH TEXP120RUH TEXP130RH TEXP135RH TEXP140RH TEXP150RH Textured, High & Ultra High Profile, Integral Injection Dome, 300cc, 375cc, 535cc, 475cc, 500cc, 600cc, 750cc,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1681-2026·2026-04-08

    CPX 4 SILTEX low height expander infusion needles may be blunt

    Mentor is recalling CPX 4 US Breast Tissue Expanders with Suture Tabs in the SILTEX low height range because the infusion sets may contain a dull or blunt needle tip.

    Product
    CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: 354-9113, 3549313, 354-9314 SILTEX Low Height Tissue Expander Style (9100, 450cc), (9300, 450cc), (9300, 550cc)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1671-2026·2026-04-08

    Philips Allura Xper FD20/15 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD20/15 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/15; System Code: 722058;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1661-2026·2026-04-08

    Philips Allura Xper FD10 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD10 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10; System Code: (1) 722003, (2)722010, (3)722026;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0414-2026·2026-04-08

    GNP redness and dry eye relief drops recalled over sterility assurance

    GNP Eye Drops Redness and Dry Eye Relief, containing glycerin and naphazoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    LEADER EYE DROPS REDNESS RELIEF — LEADER EYE DROPS REDNESS RELIEF (GLYCERIN, NAPHAZOLINE HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1744-2026·2026-04-08

    MiniMed 700G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 700G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1690-2026·2026-04-08

    Medline ENFit G-tube connectors may leak from dimensional specification failure

    Medline is recalling ENFit G-tube connectors because they were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices.

    Product
    ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1701-2026·2026-04-08

    da Vinci permanent cautery hook recalled over frayed pitch cables

    Intuitive Surgical is recalling da Vinci S and Si Permanent Cautery Hook Instruments following increased complaints regarding frayed or broken pitch cables on reusable surgical instruments.

    Product
    Brand Name: da Vinci S, Si Permanent Cautery Hook Instrument Product Name: da Vinci S, Si Permanent Cautery Hook Instrument Model/Catalog Number: 420183 Software Version: NA Product Description: EndoWrist Instruments or da Vinci S/Si Reusable instruments, including blunt and
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1694-2026·2026-04-08

    Olympus SOLTIVE Premium laser power supply may fail producing smoke

    Olympus is recalling SOLTIVE Premium SuperPulsed Laser Systems because a potential defect in the 24V power supply module may render the system inoperable and produce smoke or a burning smell.

    Product
    Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1655-2026·2026-04-08

    Paradise renal denervation catheter distributed despite being marked for scrapping

    Recor Medical is recalling one lot of the Paradise Ultrasound Renal Denervation System Catheter because a nonconforming product that should have been scrapped was distributed instead.

    Product
    Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1710-2026·2026-04-08

    Medline bronchoscopy kits contain Olympus biopsy valves that shed fragments

    Medline is recalling six bronchoscopy and thoracic kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRON
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1766-2026·2026-04-08

    Oxoid Microbact Listeria kit may identify the incorrect organism

    Oxoid, part of Thermo Fisher Scientific, is recalling one lot of the Microbact 12L Listeria identification kit because it may identify the incorrect organism.

    Product
    OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1755-2026·2026-04-08

    Burlington Medical kilts may lose radiation attenuation over time

    Burlington Medical is recalling kilts made with Xenolite 800 attenuation material because of the potential for attenuation degradation over time.

    Product
    Burlington Medical, Kilt
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1762-2026·2026-04-08

    Burlington Medical half aprons may lose radiation attenuation over time

    Burlington Medical is recalling half aprons because of the potential for attenuation degradation over time, decreasing the lifespan of the product.

    Product
    Burlington Medical, Half Aprons.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0413-2026·2026-04-08

    GoodSense original formula eye drops recalled over sterility assurance

    GoodSense Eye Drops Original Formula, containing tetrahydrozoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    GOODSENSE EYE DROPS ORIGINAL FORMULA — GOODSENSE EYE DROPS ORIGINAL FORMULA (TETRAHYDROZOLINE HCL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0410-2026·2026-04-08

    Quality Choice moisturizing relief eye drops recalled over sterility assurance

    Quality Choice Moisturizing Relief Eye Drops, containing dextran 70, polyethylene glycol 400, povidone and tetrahydrozoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    QUALITY CHOICE MOISTURIZING RELIEF EYE DROPS — QUALITY CHOICE MOISTURIZING RELIEF EYE DROPS (DEXTRAN 70, POLYETHYLENE GLYCOL 400, POVIDONE, TETRAHYDROZOLINE HCL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1687-2026·2026-04-08

    Artoura ultra high profile expander infusion needles may be blunt

    Mentor is recalling Artoura Breast Tissue Expanders in the smooth ultra high profile range because the infusion sets may contain a dull or blunt needle tip.

    Product
    Artoura Breast Tissue Expanders Reference Numbers: SDC-100UH SDC-110UH SDC-120UH SDC-130UH SDC-135UH SDC-140UH Smooth Ultra High Profile Tissue Expander, 350cc, 455cc, 535cc, 650cc, 700cc, 850cc
    Category
    Medical Device
    Distribution
    Distributed nationwide