  • SevereFDA (Devices)·Z-1946-2023·2023-07-05

    Preformed AGT oral endotracheal tube connector disconnection recalled

    Teleflex LLC recalls 14,780 preformed AGT oral endotracheal tubes due to reports of 15mm connector disconnection. Units were distributed nationwide.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780065
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1954-2023·2023-07-05

    Preformed AGT Endotracheal Tubes Recalled for Connector Disconnection Risk

    TELEFLEX is recalling approximately 43,571 units of Preformed AGT Oral Endotracheal Tubes due to reports of 15mm connector disconnection from the tube, affecting nationwide distribution including Puerto Rico.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111781060
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1859-2023·2023-07-05

    Flexi-Set Uncuffed Endotracheal Tube Sets recalled for connector disconnection

    Teleflex is recalling approximately 10,600 Flexi-Set Uncuffed Endotracheal Tube and Stylet Sets due to reports that the 15mm connector may disconnect from the tube during use.

    Product
    Flexi-Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 506550
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1878-2023·2023-07-05

    Endotracheal Tube Connector Disconnection Recall — Teleflex Preformed AGT Units

    Teleflex is recalling 347,422 Preformed AGT Oral Endotracheal Tubes due to reports of 15mm connector disconnection. The connector may separate from the tube during use.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111781040
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1902-2023·2023-07-05

    Slick Set Endotracheal Tubes Recalled for Connector Disconnection

    Teleflex is recalling 870 units of Slick Set Uncuffed Endotracheal Tubes nationwide due to reports of disconnection of the 15mm connector from the tube.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 150030
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1879-2023·2023-07-05

    Teleflex Preformed AGT Endotracheal Tubes Connector Disconnection Recall

    Teleflex is recalling approximately 446,163 Preformed AGT Oral Endotracheal Tubes due to reports of 15mm connector disconnection. The connector failure could affect tube function during patient airway management.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111781045
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1945-2023·2023-07-05

    Endotracheal Tube Connector Disconnection Recall: TELEFLEX Preformed AGT

    TELEFLEX is recalling 16,570 units of Preformed AGT Oral Endotracheal Tubes due to reports of 15mm connector disconnection, which could compromise airway management during use.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780060
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1925-2023·2023-07-05

    Endotracheal Tube Connector Disconnection Reported by Teleflex

    Teleflex is recalling endotracheal tubes due to reports of disconnection between the tube and its 15mm connector. Affected devices were distributed nationwide including Puerto Rico.

    Product
    Endotracheal Tube oral/nasal uncuffed/plain - Magill, REF 100380025
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1892-2023·2023-07-05

    Endotracheal tube connector may disconnect during patient use

    Teleflex is recalling endotracheal tubes due to reports of 15mm connector disconnection. The FDA Class I recall affects approximately 30,423 units distributed nationwide.

    Product
    Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082090
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1928-2023·2023-07-05

    Endotracheal Tubes with Risk of Connector Disconnection Recalled

    Teleflex is recalling 3,370 endotracheal tubes nationwide after reports of disconnection of the 15mm connector. Disconnection during use could interrupt airway access.

    Product
    Endotracheal Tube oral/nasal uncuffed/plain - Magill, REF 100380040
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1843-2023·2023-07-05

    Slick Set Endotracheal Tubes Recalled for Connector Disconnection

    Teleflex LLC is recalling Slick Set Endotracheal Tube Sets for reports of 15mm connector disconnection. The connector may separate from the tube, affecting respiratory support in patients.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170150
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1871-2023·2023-07-05

    Preformed AGT Endotracheal Tubes Recalled for Connector Disconnection Risk

    Teleflex is recalling 12,852 units of Preformed AGT Oral Endotracheal Tubes due to reports of 15mm connector disconnection from the tube. Affected units were distributed nationwide including Puerto Rico.

    Product
    Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100181040
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1842-2023·2023-07-05

    Slick Set Uncuffed Endotracheal Tubes recalled for connector disconnection risk

    Teleflex is recalling Slick Set Uncuffed Endotracheal Tubes nationwide due to reported disconnection of the 15mm connector from the tube. This disconnection could result in loss of airway during patient care.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170145
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1860-2023·2023-07-05

    Endotracheal Tube Connector Disconnection Recall

    TELEFLEX LLC is recalling Slick Set Cuffed Endotracheal Tubes (REF 1700100) due to reports of the 15mm connector disconnecting from the device during use.

    Product
    Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 1700100
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1953-2023·2023-07-05

    Endotracheal Tubes Recalled for Potential 15mm Connector Disconnection

    TELEFLEX LLC is recalling Preformed AGT Oral Endotracheal Tubes because the 15mm connector may disconnect from the tube, creating a potential airway access risk.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111781055
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1939-2023·2023-07-05

    Endotracheal Tubes Recalled for Connector Disconnection by Teleflex

    TELEFLEX LLC is recalling 2,140 endotracheal tubes nationwide due to reports of the 15mm connector disconnecting from the tube. Affected units were distributed throughout the United States, including Puerto Rico.

    Product
    Endotracheal Tube oral/nasal uncuffed/plain - Murphy, REF 100382065
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1862-2023·2023-07-05

    Preformed AGT Endotracheal Tubes Recalled for Connector Disconnection Risk

    Teleflex is recalling 4,769 endotracheal tubes due to reports of 15mm connector disconnection. This defect could interrupt breathing support during airway management procedures.

    Product
    Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180035
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1996-2023·2023-07-05

    MEGADYNE MEGA SOFT Patient Return Electrode Recalled for Reported Patient Burns

    Megadyne Medical Products is recalling the MEGADYNE MEGA SOFT Dual Reusable Patient Return Electrode used in electrosurgery due to reports of patient burns during surgical procedures.

    Product
    MEGADYNE MEGA SOFT Dual Reusable Patient Return Electrode. used during electrosurgery
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1857-2023·2023-07-05

    Flexi-Set Uncuffed Endotracheal Tubes recalled for connector disconnection risk

    TELEFLEX LLC recalls Flexi-Set Uncuffed Endotracheal Tube and Stylet Sets due to reports of the 15mm connector disconnecting from the tube. Approximately 16,400 units distributed nationwide are affected.

    Product
    Flexi-Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 506540
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1870-2023·2023-07-05

    Preformed AGT Oral Endotracheal Tubes recalled for connector disconnection risk

    TELEFLEX is recalling 2,874 Preformed AGT Oral Endotracheal Tubes nationwide due to reports of the 15mm connector disconnecting from the tube. Affected users should immediately discontinue use.

    Product
    Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100181030
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1952-2023·2023-07-05

    FDA Recalls Teleflex Endotracheal Tubes for Connector Disconnection

    The FDA is recalling approximately 302,652 Teleflex preformed oral endotracheal tubes nationwide due to reports of disconnection between the 15mm connector and the tube. Healthcare providers should immediately discontinue use of affected units.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111781050
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1948-2023·2023-07-05

    Endotracheal Tube Recalled for 15mm Connector Disconnection Risk

    TELEFLEX LLC is recalling 9,070 Preformed AGT Oral Endotracheal Tubes nationwide due to reports of 15mm connector disconnection. This medical device is used for airway management in hospital settings.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780080
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1903-2023·2023-07-05

    Endotracheal Tube Connector Failure Recall: Teleflex Slick Set Nationwide

    Teleflex is recalling 510 units of Slick Set Endotracheal Tubes with reported connector disconnection issues. The defect may prevent proper airway management during critical medical use.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 150035
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1962-2023·2023-07-05

    FDA recalls Teleflex endotracheal tubes for connector disconnection

    Teleflex endotracheal tubes with high-volume, low-pressure cuffs are recalled due to reports of 15mm connector disconnection. The recall affects 13,810 units distributed nationwide.

    Product
    Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080075
    Category
    Medical Device
    Distribution
    Distributed nationwide