9526–9550 of 75776

  • HighFDA (Devices)·Z-3235-2024·2024-10-02

    IR Procedure Pack Recalled Due to Sterilization Assurance Failure

    American Contract Systems recalls IR Procedure Pack (Lot 2405102, expiration 05/10/2024) because it cannot confirm sterilization requirements were met. An inoperable chart recorder caused humidity levels to be out-of-specification.

    Product
    IR Procedure Pack, CEAT19W; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3282-2024·2024-10-02

    Boston Scientific recalls WallFlex and Agile esophageal stents for delivery catheter detachment

    Boston Scientific has recalled 308 units of WallFlex and Agile esophageal stent systems due to the potential for delivery catheter tip detachment during stent placement. The affected devices are distributed worldwide.

    Product
    WALLFLEX PC ESOPH STENT 18/23MM X 123MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophag
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3280-2024·2024-10-02

    FDA Recalls WallFlex Esophageal Stent Systems Due to Catheter Tip Detachment Risk

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems (516 units) due to potential for delivery catheter tip detachment during stent placement procedures.

    Product
    WALLFLEX ESOPHAGEAL FC 23/28MMX15CM-Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal f
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3299-2024·2024-10-02

    Arrow Arterial Kits Recalled Due to Open Packaging Seals

    Arrow International recalls over 14,000 arterial kits due to reports of open packaging seals. The recall affects multiple models distributed nationwide.

    Product
    Arrow Arterial Kits and Trays as follows: (1) Arrow Arterial Access Tray with 0.025" Spring-Wire Guide (Catheter Not Included), REF ASK-04001-BW1; (2) Arrow Arterial Access Tray with 0.021" Spring-Wire Guide (Catheter Not Included), REF ASK-04001-CK; (3) Arrow Arterial
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3265-2024·2024-10-02

    Esophageal stent system recalled due to potential delivery catheter detachment

    Boston Scientific is recalling Wallflex Esophageal Stent Systems due to potential delivery catheter tip detachment that could affect proper stent deployment during medical procedures.

    Product
    WALLFLEX FC ESO STENT RMV 18X103MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esoph
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3272-2024·2024-10-02

    Esophageal stent systems recalled due to delivery catheter tip detachment risk

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems due to potential delivery catheter tip detachment during stent placement procedures.

    Product
    WALLFLEX FC ESO STENT RMV LL 23X105- Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN): M00516300
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3250-2024·2024-10-02

    Cook Biotech Biodesign Fistula Plug implants recalled for premature expiration

    Cook Biotech is recalling Biodesign Fistula Plug C-FPB-0.4 implants because products expire before their printed expiration dates.

    Product
    Biodesign Fistula Plug C-FPB-0.4 - For implantation to reinforce soft tissue for repair of recto-vaginal or anorectal fistulas. Order Number (GPN): G24667
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3242-2024·2024-10-02

    Nasal Sinus Pack Recalled Due to Sterilization Verification Failure

    American Contract Systems is recalling Nasal Sinus Pack units due to manufacturing control failures that prevent confirmation of sterilization. Affected units were distributed to multiple states.

    Product
    Nasal Sinus Pack, MTSN26B; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0664-2024·2024-10-02

    FDA Recalls Refresh P.M. Eye Ointment Due to Potential Sterility Breach

    AbbVie is recalling Refresh P.M. lubricant eye ointment due to potential breaches in tube seals that may compromise product sterility. Approximately 2.4 million tubes were distributed nationwide and internationally.

    Product
    Refresh P.M., (Mineral Oil 42.5%, White Petrolatum 57.3%) Lubricant Eye Ointment, Net wt, 0.12 oz (3.5g) Tube, Distributed by: Allergen, an AbbVie company, Madison, NJ 07940, UPC code: 300230667043,
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3252-2024·2024-10-02

    Biodesign Fistula Plug recalled for expiring before printed expiration date

    Cook Biotech is recalling 35 units of Biodesign Fistula Plug surgical implants that expire before their printed expiration dates, potentially compromising device function.

    Product
    Biodesign Fistula Plug C-FPS-0.2 - For implantation to reinforce soft tissue for repair of recto-vaginal or anorectal fistulas. Order Number (GPN): G54612
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3287-2024·2024-10-02

    Esophageal Stent Systems Recalled for Potential Delivery Catheter Detachment

    Boston Scientific is recalling esophageal stent systems due to potential delivery catheter tip detachment during placement. Patients and providers with affected devices should contact the manufacturer for guidance.

    Product
    AGILE ESO OTW PC 23MM X 6.2CM IN 18.5 FR- Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of con
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3291-2024·2024-10-02

    Boston Scientific Esophageal Stent Systems Recalled for Catheter Tip Detachment Risk

    Boston Scientific is recalling WallFlex and Agile Esophageal Stent Systems for potential delivery catheter tip detachment during insertion. Eighteen units were distributed worldwide including the United States.

    Product
    AGILE ESO OTW FC 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3193-2024·2024-10-02

    MRI Diagnostic Device Recalls Due to Loose Screw Assembly Risk

    Philips recalls Evolution Upgrade 3.0T MRI devices due to loose screws that may block the tabletop and delay diagnosis. Devices with the affected serial numbers should be taken out of service.

    Product
    Evolution Upgrade 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: (1) 782117 (2) 782143
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3285-2024·2024-10-02

    Esophageal Stent System Recalled Due to Potential Catheter Tip Detachment

    Boston Scientific recalls WallFlex and Agile esophageal stent systems (305 units) due to potential delivery catheter tip detachment. Affected products were distributed worldwide.

    Product
    WALLFLEX PC ESOPH STENT 23/28MM X 125MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophag
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3227-2024·2024-10-02

    Endovascular AAA Packs Recalled Due to Sterilization Process Defect

    American Contract Systems is recalling 93 units of Endovascular AAA Packs due to inability to confirm sterilization assurance requirements were met following an environmental monitoring failure.

    Product
    Endovascular AAA Pack, ESED52B; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3231-2024·2024-10-02

    Day Surgery General Pack medical kit recalled for sterilization concern

    American Contract Systems recalled 280 units of its Day Surgery General Pack due to humidity control failures that prevent confirmation of adequate sterilization.

    Product
    Day Surgery General Pack, FHGE18T; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3205-2024·2024-10-02

    MRI diagnostic device screws may loosen and block tabletop

    Philips is recalling the Ingenia Elition X MRI device due to loose screws in the RF carrier assembly that could block the tabletop and delay diagnosis or cause patient anxiety.

    Product
    SmartPath to Ingenia Elition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782118
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3295-2024·2024-10-02

    Boston Scientific Agile Esophageal Stent Recall: Catheter Tip Detachment Risk

    Boston Scientific is recalling Agile Esophageal Stent Systems due to potential delivery catheter tip detachment. The Class II recall affects 6 units distributed worldwide.

    Product
    AGILE ESO FC RMV 23MM X 15.0CM IN 18.5 F- Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of conc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3297-2024·2024-10-02

    CooperSurgical Recalls Origio Sperm Wash Medium Due to Low Sodium Pyruvate

    CooperSurgical is recalling Origio Sperm Wash Medium because low sodium pyruvate levels could compromise gamete development.

    Product
    Origio Sperm Wash Medium Pack of Five 60ml bottles - Intended for washing of spermatozoa. Part Number: 84055060A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3190-2024·2024-10-02

    Prelude Sheath Introducer 7.5F devices recalled due to labeling error

    Merit Medical Systems is recalling 1,892 Prelude Sheath Introducer devices due to a labeling error: devices labeled as 7.5F are actually sized as 7F. The dimensional mislabeling could result in incorrect device selection during medical procedures.

    Product
    Prelude Sheath Introducer 7.5F, REF: PSI-7F-11-038 Version F (PSI-7F-11-038/F)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3219-2024·2024-10-02

    Karl Storz Hopkins Telescope Recalled for Unapproved Reprocessing Instructions

    Karl Storz Hopkins Telescope 6 endoscopes are being recalled because the instructions for use contain reprocessing procedures not reviewed or approved by the FDA. Affected units should not be used until corrected instructions are provided.

    Product
    Karl Storz - Endoskope , REF: 27093AA Hopkins Telescope 6, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0665-2024·2024-10-02

    Refresh LACRI-LUBE Eye Ointment Recalled for Potential Tube Seal Breach

    Allergen is recalling specific lots of Refresh LACRI-LUBE eye ointment (763,426 tubes) due to potential breaches in the tube seal that may compromise sterility. Consumers should verify their product lot number against the recall list.

    Product
    Refresh LACRI-LUBE, (42.5% Mineral Oil, 57.3% White Petrolatum) Lubricant Eye Ointment, Net wt, 0.12 oz (3.5g) Tube, Distributed by: Allergen, an AbbVie company, Madison, NJ 07940, UPC code: 300230312042
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3222-2024·2024-10-02

    FDA Recalls Karl Storz Endoscopes Due to Unapproved Reprocessing Instructions

    Karl Storz is recalling 78 endoscopes because their instructions for use contain reprocessing modalities that have not been reviewed or approved by the FDA.

    Product
    Karl Storz - Endoskope , REF: 27294AA, Hopkins Telescope 6, 18 Fr, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0660-2024·2024-10-02

    Mupirocin Antibiotic Ointment Recalled Nationwide Due to Subpotency

    Glenmark Pharmaceuticals recalls Mupirocin Ointment nationwide due to subpotency, meaning some lots contain less active ingredient than specified. Affected customers should contact their pharmacy or prescriber.

    Product
    MUPIROCIN — MUPIROCIN (MUPIROCIN)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3243-2024·2024-10-02

    Medical device kit recalled due to inability to confirm sterilization assurance

    American Contract Systems recalled its Cysto medical convenience kit (36 units, Lot 2405033) due to a humidity monitoring failure that prevented confirmation of adequate product sterilization.

    Product
    Cysto, FTCY03L; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide