  • HighFDA (Devices)·Z-1665-2026·2026-04-08

    Philips Allura Xper FD20 OR Table X-ray may not start

    Philips is recalling the Allura Xper FD20 OR Table because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD20 OR Table; System Code: (1) 722015, (2)722023, (3)722035;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1682-2026·2026-04-08

    CPX 4 low height tissue expander infusion needles may be blunt

    Mentor is recalling the CPX 4 US Breast Tissue Expander with Suture Tabs, smooth low height 650 cc, because the infusion sets may contain a dull or blunt needle tip.

    Product
    CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-157LH Smooth Low Height Tissue Expander, 650cc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1657-2026·2026-04-08

    Atellica hemoglobin A1c assay may return depressed patient results

    Siemens is recalling Atellica CH Enzymatic Hemoglobin A1c because results may be depressed when the RCRP reagent is processed on the same analyzer, with bias increasing over time.

    Product
    Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1760-2026·2026-04-08

    Burlington Medical caps may lose radiation attenuation over time

    Burlington Medical is recalling protective caps made with Xenolite 800 attenuation material because of the potential for attenuation degradation over time.

    Product
    Burlington Medical, Caps.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1757-2026·2026-04-08

    Burlington Medical wrap aprons may lose radiation attenuation over time

    Burlington Medical is recalling wrap aprons made with Xenolite 800 attenuation material because of the potential for attenuation degradation over time.

    Product
    Burlington Medical, Wrap Aprons.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1744-2026·2026-04-08

    MiniMed 700G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 700G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1678-2026·2026-04-08

    Philips Azurion 7 M20 X-ray may not start from foot switch

    Philips has identified instances where X-ray imaging on the Azurion 7 M20 may not initiate, or may run intermittently, when the wired foot switch is used. 5,167 units are affected.

    Product
    Azurion 7 M20; System Code: (1)722079, (2)722224, (3)722234, (4)722282 (OUS only);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1755-2026·2026-04-08

    Burlington Medical kilts may lose radiation attenuation over time

    Burlington Medical is recalling kilts made with Xenolite 800 attenuation material because of the potential for attenuation degradation over time.

    Product
    Burlington Medical, Kilt
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1743-2026·2026-04-08

    MiniMed 640G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 640G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1750-2026·2026-04-08

    Paradigm REAL-Time Revel pump delivery varies with pump height changes

    Medtronic is recalling Paradigm REAL-Time Revel insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1756-2026·2026-04-08

    Burlington Medical frontal aprons may lose radiation attenuation over time

    Burlington Medical is recalling frontal aprons made with Xenolite 800 attenuation material because of the potential for attenuation degradation over time.

    Product
    Burlington Medical, Frontal Aprons
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1696-2026·2026-04-08

    da Vinci monopolar curved scissors recalled over frayed grip cables

    Intuitive Surgical is recalling da Vinci S and Si Monopolar Curved Scissors Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1691-2026·2026-04-08

    Medline ENFit G-tube connector kits may leak from dimensional defect

    Medline is recalling kits containing ENFit G-tube connectors because the connectors were not made to required dimensional specifications and may not form a proper seal.

    Product
    Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1747-2026·2026-04-08

    MiniMed 620G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 620G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 620G Insulin Pump (MMT-1750)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1708-2026·2026-04-08

    Medline endoscopy kits contain recalled Olympus biopsy valves shedding fragments

    Medline is recalling BAL and ENDO kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0416-2026·2026-04-08

    Leader artificial tears 15 mL recalled over sterility assurance failure

    Leader Artificial Tears 15 mL, containing polyvinyl alcohol and povidone, are being recalled for lack of assurance of sterility.

    Product
    LEADER ARTIFICIAL TEARS 15 ML — LEADER ARTIFICIAL TEARS 15 ML (POLYVINYL ALCOHOL, POVIDONE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1752-2026·2026-04-08

    Paradigm REAL-Time Veo pump delivery varies with pump height changes

    Medtronic is recalling Paradigm REAL-Time Veo insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1706-2026·2026-04-08

    Medline bronchoscopy packs contain recalled Olympus biopsy valves shedding fragments

    Medline is recalling three bronchoscopy convenience kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1768-2026·2026-04-08

    Monoject tuberculin syringe packages actually contain U-100 insulin syringes

    Cardinal Health is recalling two lots of Monoject Tuberculin Syringes because the packaging identifies them as 1 mL Luer Lock tuberculin syringes but the contents are U-100 insulin syringes.

    Product
    Cardinal Health Monoject Tuberculin Syringe, 1 mL Luer Lock, Catalog Number 1180100777
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1661-2026·2026-04-08

    Philips Allura Xper FD10 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD10 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10; System Code: (1) 722003, (2)722010, (3)722026;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1697-2026·2026-04-08

    da Vinci double fenestrated grasper recalled over broken grip cables

    Intuitive Surgical is recalling da Vinci S and Si Double Fenestrated Grasper Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci S, Si Double Fenestrated Grasper Instrument Model/Catalog Number: 420189 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including bl
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1671-2026·2026-04-08

    Philips Allura Xper FD20/15 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD20/15 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/15; System Code: 722058;
    Category
    Medical Device
    Distribution
    Distributed nationwide