72,933 recalls have nationwide distribution and so reach Ohio. 11,719 additional recalls listed Ohio specifically in their distribution scope.

What the numbers show

Total
11,719
Critical
2,106
Severe
2,413
Most recent
2026-09-09

Of the 11,719 recalls naming Ohio we have scored against our rubric, 2,106 (18%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2,413 (21%) are Severe.

These figures cover only the 11,719 recalls whose distribution notice names Ohio specifically. The 72,933 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

2751–2775 of 84652

  • HighFDA (Devices)·Z-1613-2026·2026-04-01

    Olympus OER-Pro reprocessor no longer compatible with two valve models

    Olympus is recalling the OER-Pro Endoscope Reprocessor because the MAJ-1443 and MAJ-1444 air/water valves are no longer reprocessing compatible with the OER-Pro and OER-Elite reprocessors.

    Product
    Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1629-2026·2026-04-01

    Orthalign drill plate sizes 1-5 may allow excessive bone cut

    Orthalign is recalling one lot of the Trimax Pegged Drill Plate sizes 1-5 for the Lantern Knee System because plates manufactured out of specification can result in a bone cut in excess of 0.0800 inches.

    Product
    Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0404-2026·2026-04-01

    Teva octreotide acetate 10 mg recalled over sterility assurance deficiencies

    Teva is recalling two lots of Octreotide Acetate for Injectable Suspension 10 mg after quality system deficiencies were identified during an FDA inspection at the contract manufacturer.

    Product
    OCTREOTIDE ACETATE — OCTREOTIDE ACETATE (OCTREOTIDE ACETATE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1590-2026·2026-04-01

    GE Imactis CT navigation may show needle trajectory that differs from actual

    GE Medical Systems is recalling the Imactis CT-Navigation System because of a potential discordance between the simulated needle trajectory and tip position and the actual trajectory and tip position.

    Product
    Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0406-2026·2026-04-01

    Teva octreotide acetate injection recalled over sterility assurance deficiencies

    Teva is recalling twelve lots of Octreotide Acetate for Injectable Suspension after quality system deficiencies were identified during an FDA inspection at the contract manufacturer.

    Product
    OCTREOTIDE ACETATE — OCTREOTIDE ACETATE (OCTREOTIDE ACETATE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0595-2026·2026-04-01

    Ajinomoto Golden Tiger Vegetable Fried Rice recalled for glass fragments

    Ajinomoto Foods North America Inc. is recalling Ajinomoto Golden Tiger Vegetable Fried Rice due to the presence of glass fragments varying from 1–3 cm long and 2–4 mm wide in the product.

    Product
    Item 5654390 Ajinomoto Golden Tiger Vegetable Fried Rice - Long grain white rice and crispy vegetables in a savory sauce. Net wt. 3lbs per bag. Retail bag UPC 076366565439. Product packaging is flexible bags in a master case or flexible bags in cartons in a master case. There ar
    Category
    Food
    Distribution
    43 states
  • HighFDA (Devices)·Z-1641-2026·2026-04-01

    Olympus HALO PKS cutting forceps jaws may break from defective welds

    Olympus is recalling HALO PKS Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electro
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1644-2026·2026-04-01

    Olympus PKS cutting forceps jaws may break from defective welds

    Olympus is recalling PKS Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visuali
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1632-2026·2026-04-01

    Straumann custom abutments made from wrong bar material after mix-up

    Straumann is recalling Custom Abutment Ti for Straumann RC because a material mix-up meant titanium abutments were manufactured from TAN bar material instead of Titanium Grade 4.

    Product
    Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1607-2026·2026-04-01

    I.T.S. hand locking plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling the HLS Hand Locking Plates System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11, 32151-12, 32151-13, 32151-14, 32151-15, 32151-16, 32151-18, 32151-20. 2. HLS Plate, 2 T-Shape
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1595-2026·2026-04-01

    I.T.S. calcaneus repair plate heats more in MRI than stated

    I.T.S. GmbH is recalling the FR.O.H. Calcaneus Repair System anterolateral calcaneus plate because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1600-2026·2026-04-01

    I.T.S. pelvic reconstruction screws heat more in MRI than stated

    I.T.S. GmbH is recalling the Pelvic Reconstruction System fixation screw and short spike because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1611-2026·2026-04-01

    Olympus endoscope air water valves no longer compatible with named reprocessors

    Olympus is recalling the MAJ-1444 Endoscope Air/Water Valve because the MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

    Product
    Olympus Endoscope Air/Water Valve - MAJ-1444. Model Number: MAJ-1444.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1602-2026·2026-04-01

    I.T.S. distal humeral plates heat more in MRI than stated

    I.T.S. GmbH is recalling Distal Humeral Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N, 373
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0612-2026·2026-04-01

    Pepsin 1:3000 Powder recalled for potential Salmonella contamination

    American Laboratories is recalling Pepsin 1:3000 Powder (50 kg, Item #FTI-G0193-00) due to potential Salmonella contamination. The product was distributed to wholesale accounts in multiple states and Australia.

    Product
    Pepsin 1:3000 Powder, 50 kg, Item #FTI-G0193-00, American Laboratories 5036 South 33rd Street, Omaha, Nebraska 68107. Packed in double poly-lined cardboard drums or boxes.
    Category
    Food
    Distribution
    9 states
  • HighFDA (Devices)·Z-1597-2026·2026-04-01

    I.T.S. olecranon plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling Olecranon Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1592-2026·2026-04-01

    I.T.S. volar radius plates heat more in MRI than stated

    I.T.S. GmbH is recalling Volar Radius Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0592-2026·2026-04-01

    Trader Joe's Vegetable Fried Rice recalled for glass fragments

    Ajinomoto Foods North America Inc. is recalling Trader Joe's Vegetable Fried Rice (1 lb bags, Item 5650233) due to glass fragments (1–3 cm long, 2–4 mm wide) found in the product. The recall affects approximately 9.9 million pounds distributed across 40 US states.

    Product
    Item 5650233 Trader Joe's Vegetable Fried Rice, net wt. 1lb per bag. Retail bag UPC 00521482. Product packaging is flexible bags in a master case or flexible bags in cartons in a master case. There are 24/1lb. bags per case. Case UPC 10052148200001. All products are intended t
    Category
    Food
    Distribution
    43 states
  • HighFDA (Food)·H-0600-2026·2026-04-01

    Ajinomoto Vegetable Gyoza Dumplings recalled for glass fragments

    Ajinomoto Foods North America Inc. is recalling Vegetable Gyoza Dumplings (UPC 071757055483) due to the presence of glass fragments in the product, which could pose a choking or injury hazard if ingested.

    Product
    Item 6430143 Ajinomoto Vegetable Gyoza Dumplings 12ct. Net wt. 8.92oz. per package. Retail package UPC: 071757055483. Product packaging is flexible bags in a master case or flexible bags in cartons in a master case. There are 20/8.92oz. packages or 11.15 lbs. per case. Carton UP
    Category
    Food
    Distribution
    43 states
  • HighFDA (Food)·H-0610-2026·2026-04-01

    Pepsin Full Strength Powder recalled for potential Salmonella contamination

    American Laboratories is recalling Pepsin Full Strength Powder (50 kg drums/boxes, Item #5000545) due to potential Salmonella contamination. The product was distributed to wholesale accounts in multiple US states and Australia.

    Product
    Pepsin Full Strength Powder, 50 kg, Item #5000545, American Laboratories 5036 South 33rd Street, Omaha, Nebraska 68107. Packed in double poly-lined cardboard drums or boxes.
    Category
    Food
    Distribution
    9 states
  • HighNHTSA·26V209000·2026-04-01

    Audi Q3 recalled over remote sunroof closing from out of sight

    Volkswagen Group of America is recalling 6,077 Audi Q3 vehicles because incorrect software may allow remote sunroof closing from a distance where the operator cannot clearly see the sunroof.

    Product
    AUDI — AUDI Q3
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1634-2026·2026-04-01

    TECNIS Eyhance intraocular lens haptic may stick to the optic

    AMO Puerto Rico Manufacturing is recalling TECNIS Eyhance IOLs with the TECNIS Simplicity Delivery System because a limited quantity may have a haptic that sticks to the optic, preventing consistent unfolding.

    Product
    Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1639-2026·2026-04-01

    Olympus Everest bipolar cutting forceps jaws may break from defective welds

    Olympus is recalling Everest Bipolar Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achi
    Category
    Medical Device
    Distribution
    Distributed nationwide