The Recall Desk

State

New York product recalls

73,833 recalls have nationwide distribution and so reach New York. 12,956 additional recalls listed New York specifically in their distribution scope.

What the numbers show

Total
12,956
Critical
1,508
Severe
3,702
Most recent
2026-11-04

Of the 12,956 recalls naming New York we have scored against our rubric, 1,508 (12%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 3,702 (29%) are Severe.

These figures cover only the 12,956 recalls whose distribution notice names New York specifically. The 73,833 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

54026–54050 of 86789

  • HighFDA (Devices)·Z-0650-2017·2016-11-30

    DePuy ACTIS Flex Reamers Recalled Due to Breakage Risk

    DePuy Orthopaedics is voluntarily recalling ACTIS Flex Reamers because the instruments may break and leave pieces in patients.

    Product
    ACTIS Flex Reamers SZ 0/1; 2/3; 4/5; 6/7; 8/9; 10/11; 12, Product is sold non-sterile, these instruments are packed non-sterile, within a rigid tube with foam protectors, with label and IFU
    Category
    Medical Device
    Distribution
    11 states
  • HighFDA (Devices)·Z-0622-2017·2016-11-30

    Toshiba Aquilion Prime CT System Recalled for Cooling Oil Leak Risk

    Toshiba American Medical Systems is recalling Aquilion Prime CT systems because cooling oil in the x-ray tube heat exchanger may leak during operation.

    Product
    Aquilion Prime TSX-303A Aquilion PRIME is a multislice helical CT system with an 40-row detector. High-speed rotation allows rapid data acquisition and shortens scan times, while the fast reconstruction unit further improves throughput, reducing the time required for diagnosis.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·16V861000·2016-11-30

    Cummins Recalls 2015-2016 Generator Sets Due to Door Hinge Failure Risk

    Cummins is recalling 164 model year 2015-2016 generator sets because door hinges may fail, causing doors to detach and create road hazards.

    Product
    CUMMINS POWER GENERATION — CUMMINS POWER GENERATION GENERATOR SET
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0375-2017·2016-11-30

    Philips CT Scanner Recall Due to Tabletop Movement Failure

    Philips is recalling 1,363 CT scanners because the patient support tabletop may fail to move horizontally during scans due to a mechanical linkage failure.

    Product
    Philips Healthcare System X-ray Tomography Computed :
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·16V863000·2016-11-30

    LGS Recalls 2017 Pace, Cargo Express, and Look Cargo Trailers

    LGS Industries is recalling 2017 Pace, Cargo Express, and Look cargo trailers because a defective weld could cause the ramp door bracket to detach, creating a crushing hazard.

    Product
    PACE — PACE, CARGO EXPRESS, LOOK CARGO TRAILER
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0633-2017·2016-11-30

    Merit Medical Procedure Kits Recalled for Compromised Sterile Barrier

    Merit Medical Systems is voluntarily recalling specific Procedure Kits due to a compromised sterile barrier, which may affect the sterility of the contents.

    Product
    Waste Management Kit. Catalog Number K10-04381AP
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0652-2017·2016-11-30

    Siemens Dimension Vista Chemistry 1 Calibrator Recall for Low Calcium Recovery

    Siemens is recalling Dimension Vista Chemistry 1 Calibrators (lots 5GM081 and 5GM082) because they may cause discrepant low calcium recovery in quality control and patient samples.

    Product
    Dimension Vista Chemistry 1 Calibrator The Chemistry 1 Calibrator (KC110B) is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA/CRE2), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0651-2017·2016-11-30

    Bausch + Lomb Recalls PeroxiClear Peroxide Solution for Excess Residuals

    Bausch + Lomb is recalling PeroxiClear Peroxide Solution because the product was found to contain excess residual levels of hydrogen peroxide.

    Product
    PeroxiClear Peroxide Solution, (3oz. and 12 oz., 12 oz. Twin Pack, 2x12 oz. Twin Pack for Target stores, 12 oz. for Walmart Vision Center and 3 oz. Starter Kit US), Manufactured by Bausch + Lomb, Rochester, NY 14609. PeroxiClear 3% Hydrogen Peroxide Cleaning & Disinfecting…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0645-2017·2016-11-30

    DiaSorin Recalls LIAISON Estradiol II Gen Assay Due to Cross-Reactivity

    DiaSorin is recalling the LIAISON Estradiol II Gen assay because it may produce inaccurate results for patients treated with fulvestrant.

    Product
    LIAISON Estradiol II Gen, 310680, 310680-CN The DiaSorin LIAISON Estradiol II Gen assay is a chemiluminescent immunoassay (CLIA) intended for the quantitative determination of estradiol in human serum. Assay results should be used in conjunction with other clinical and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0376-2017·2016-11-30

    Philips CT Scanner Recall Due to Potential Tabletop Movement Failure

    Philips is recalling 1,363 CT scanners because the patient support tabletop may fail to move horizontally during scans due to a mechanical linkage failure.

    Product
    Philips Healthcare System X-ray Tomography Computed :
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0623-2017·2016-11-30

    Greatbatch Stiffer Coaxial Micro-Introducer Recalled for Detaching Luer

    Greatbatch Medical is recalling Stiffer Coaxial Micro-Introducers because a rotating luer detached from the dilator hub during use.

    Product
    Greatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002 Stiffer Coaxial Micro-Introducer 5FR. The radiopaque Stiffer Coaxial Micro-Introducer (Figure 1) consists of a dilator and sheath, available in 4FR and 5FR.…
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0614-2017·2016-11-30

    Merge Healthcare Recalls RadSuite Software Due to Image Display Malfunction

    Merge Healthcare is recalling RadSuite software versions 8.30.7.8 through 8.31 because images may not appear readable, potentially delaying care or causing patient re-exposure to radiation.

    Product
    Merge RadSuite,f/k/a: Emageon Advanced Visualization , RadSuite versions 8.30.7.8, 8.30.7.9, 8.30.8, and 8.31. Advanced Visualization (Image Viewing) includes: Full featured 2D imaging, 3D surface and volume rendering, Real-time Multi-Planar Reformatting (MPR), Real-time…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Food)·F-0499-2017·2016-11-30

    Infiniti Creations Recalls Amino Complex Formula Due to Misbranding

    Infiniti Creations, LLC is voluntarily recalling Amino Complex Formula HCG Drops due to misbranding and potential undeclared HCG.

    Product
    Amino Complex Formula. 2oz HCG Drops. Blue Glass.
    Category
    Food
    Distribution
    Distributed nationwide
  • ModerateFDA (Food)·F-0493-2017·2016-11-30

    Infiniti Creations Recalls HCG Complex Drops for Misbranding

    Infiniti Creations, LLC is voluntarily recalling HCG Complex drops due to misbranding and potential undeclared HCG content.

    Product
    HCG Complex. 2oz HCG Drops. Blue Plastic.
    Category
    Food
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0648-2017·2016-11-30

    Beckman Coulter Recalls AQUIOS Lyse Reagent Kit for Increased Flags

    Beckman Coulter is recalling the AQUIOS Lyse Reagent Kit because it may cause increased notifications and flags on AQUIOS CL systems, potentially delaying test results.

    Product
    AQUIOS Lyse Reagent Kit, Catalog No. B23538 Product Usage: AQUIOS Lysing Reagent Kit is used as part of the AQUIOS flow cytometer system. The kit consists of two reagents used by AQUIOS flow cytometers to prepare whole blood samples for analysis of white blood cells.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-0152-2017·2016-11-30

    Akorn Recalls Desoximetasone Ointment for Stability Failures

    Akorn Inc. is recalling Desoximetasone Ointment USP, 0.25%, due to failed stability specifications and phase separation.

    Product
    Desoximetasone Ointment USP, 0.25%, a) 15 g tubes (NDC 61748-206-15 and b) 60 g tubes (NDC 61748-206-60, Marketed by Versapharm Inc., An Akorn Company, Lake Forest, IL, Manufactured by Hi-Tech, Pharmacal Co, Inc., Amityville, NY
    Category
    Drug
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0644-2017·2016-11-30

    Synthes 2.5 mm Reaming Rods Recalled for Delaminating Peel Pouches

    Synthes (USA) Products LLC is recalling 2,157 sterile 2.5 mm reaming rods because their peel pouches are delaminating.

    Product
    2.5 mm Reaming Rod W/Ball Tip 650 mm, Sterile
    Category
    Medical Device
    Distribution
    Distributed nationwide