Nevada product recalls

73,096 recalls have nationwide distribution and so reach Nevada. 5,696 additional recalls listed Nevada specifically in their distribution scope.

What the numbers show

Total
5,696
Critical
951
Severe
1,479
Most recent
2026-09-25

Of the 5,696 recalls naming Nevada we have scored against our rubric, 951 (17%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1,479 (26%) are Severe.

These figures cover only the 5,696 recalls whose distribution notice names Nevada specifically. The 73,096 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

9901–9925 of 78792

  • HighFDA (Devices)·Z-3256-2024·2024-10-02

    Biodesign Fistula Plug Implant Recalled for Premature Expiration

    Cook Biotech is recalling the Biodesign Fistula Plug implant because devices expire before their labeled expiration dates. Thirty-nine units distributed worldwide are affected.

    Product
    Biodesign Fistula Plug C-FPS-0.4-2 - For implantation to reinforce soft tissue for repair of recto-vaginal or anorectal fistulas. Order Number (GPN): G46372
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3294-2024·2024-10-02

    Boston Scientific Recalls Agile Esophageal Stent Systems Due to Catheter Tip Detachment

    Boston Scientific is recalling Agile Esophageal Stent Systems because the delivery catheter tip may detach. The recall affects 10 units distributed worldwide.

    Product
    AGILE ESO FC RMV 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3224-2024·2024-10-02

    Karl Storz Nephroscope Recall Due to Unapproved Reprocessing Instructions

    Karl Storz Endoscopy is recalling 5 nephroscopes because their instructions for use list reprocessing methods not reviewed or approved by the FDA.

    Product
    Karl Storz - Endoskope , REF: 27830KA, Nephroscope for MIP M, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3269-2024·2024-10-02

    Boston Scientific Recalls WallFlex Esophageal Stent Systems Due to Catheter Tip Detachment

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems because the delivery catheter tip may detach. The recall covers 176 units distributed worldwide.

    Product
    WALLFLEX FC ESO STENT RMV 23X125MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esoph
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3202-2024·2024-10-02

    Philips MR 7700 MRI Scanner Recall for Loose Screws

    Philips is recalling 7 MR 7700 MRI scanners because loose screws may block the tabletop, potentially delaying diagnosis or causing patient anxiety.

    Product
    MR 7700 -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782120
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3288-2024·2024-10-02

    Boston Scientific Recalls Agile Esophageal Stent Systems Due to Catheter Detachment Risk

    Boston Scientific is recalling Agile Esophageal Stent Systems because the delivery catheter tip may detach. The recall affects 15 units distributed worldwide.

    Product
    AGILE ESO OTW PC 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3301-2024·2024-10-02

    Cardinal Health Recalls Salem Sump Stomach Tubes Due to Valve Breakage

    Cardinal Health is recalling Salem Sump Dual Lumen Stomach Tubes because the Anti-Reflux Valve may break due to improper use. Labeling changes were made to address the issue.

    Product
    (a) CardinalHealth Salem Sump Dual Lumen Stomach Tube, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888264929, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube, 14 Fr/Ch (4.7 mm) x 48 in. (122 cm), Model #8888264945, Sterile; (c) CardinalHealth Salem Sump Dua
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3230-2024·2024-10-02

    American Contract Systems Recalls Open Shoulder Medical Convenience Kits

    American Contract Systems is recalling 12 Open Shoulder medical convenience kits because a faulty chart recorder prevented verification of sterilization requirements.

    Product
    Open Shoulder, LWOS39L; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3281-2024·2024-10-02

    Boston Scientific Recalls WallFlex Esophageal Stents Due to Catheter Tip Detachment

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems because the delivery catheter tip may detach during use.

    Product
    WALLFLEX PC ESOPH STENT 18/23MM X 103MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophag
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3221-2024·2024-10-02

    Medical endoscope recalled for unapproved sterilization instructions

    Karl Storz Hopkins Telescope 6 endoscopes are recalled because their Instructions for Use contain reprocessing methods that lack FDA review and approval.

    Product
    Karl Storz - Endoskope , REF: 27293AA, Hopkins Telescope 6, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3204-2024·2024-10-02

    MRI Diagnostic Device Loose Screws May Block Patient Tabletop

    Philips SmartPath dStream MRI devices have loose screws in the RF carrier assembly that may interfere with the moving tabletop, potentially delaying diagnosis. Affected units should be inspected and repaired immediately.

    Product
    SmartPath to dStream for XR and 3.0T- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782113; 2) 781270; 3) 782129
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·24V737000·2024-10-02

    2025 Chevrolet Equinox EV door strikers recalled for possible breakage

    General Motors is recalling 2025 Chevrolet Equinox EV vehicles (along with Blazer and Blazer EV models) because door strikers may not be properly heat-treated and could break, causing doors to open unexpectedly. All affected vehicles remain in dealerships.

    Product
    CHEVROLET — CHEVROLET BLAZER EV, EQUINOX EV, BLAZER
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3192-2024·2024-10-02

    Biopsy instrument packaging holes compromise sterile barrier integrity

    Argon Medical Devices recalls BioPince Ultra and Tru-Core II biopsy instruments due to holes in packaging trays compromising the sterile barrier, which may allow microorganism introduction and infection risk.

    Product
    BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02, 360-2080-03, 370-1080-01, 370-1080-02, 370-1080-03, 370-1580-01, 370-1580-02, 370-1580-03; Tru-Core II Biopsy Instrument,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3187-2024·2024-10-02

    Roche Creatine Kinase test reagents recalled for calibration errors

    Roche Diagnostics is recalling Creatine Kinase test reagents that may exhibit abnormal calibrations, producing inaccurate results. Lot 755471 distributed nationwide.

    Product
    Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative determination of creatine kinase (CK) in human serum and plasma. Catalog Number: 07190794190
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3244-2024·2024-10-02

    Pacemaker COPM11B Medical Kits Recalled for Sterilization Verification Issue

    American Contract Systems is recalling 3 units of the COPM11B pacemaker medical kits due to inability to confirm sterilization requirements were met. An inoperable chart recorder caused humidity readings to exceed specifications during manufacturing.

    Product
    Pacemaker, COPM11B; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3249-2024·2024-10-02

    IntelliVue Patient Monitors recalled for electrical grounding defect

    Philips is recalling approximately 545 IntelliVue Patient Monitor units with a broken ground bolt that may cause loss of electrical grounding, potentially affecting the device's electromagnetic immunity and emission.

    Product
    IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3198-2024·2024-10-02

    Philips Ingenia Ambition X MRI Scanner Recalled Due to Loose Screws

    Philips is recalling 63 units of the Ingenia Ambition X MRI scanner due to loose screws in the IRF carrier assembly that may block the imaging tabletop and potentially delay diagnosis.

    Product
    Ingenia Ambition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782109
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3267-2024·2024-10-02

    Boston Scientific esophageal stents recalled due to catheter tip detachment risk

    Boston Scientific is recalling Wallflex and Agile esophageal stent systems due to potential detachment of the delivery catheter tip during stent placement. These devices are used to maintain patency in esophageal strictures caused by malignant tumors or esophageal fistulas.

    Product
    WALLFLEX FC ESO STENT RMV 18X153MM-Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esopha
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3213-2024·2024-10-02

    Cervical disc implant recalled due to foreign contamination during assembly

    Centinel Spine is recalling the Prodisc C SK cervical disc implant due to blue discoloration identified as foreign contamination on the inlay during assembly. The contamination was not fully evaluated in manufacturing validations.

    Product
    prodisc C SK, Total cervical disc replacement, Model/Catalog Number: PDSM5
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-1820-2024·2024-10-02

    Montreal Fudge Chocolate Fudge with Nuts Recalled for Mold Contamination

    Authentik Fudge is recalling Montreal Fudge brand Chocolate Fudge with Nuts due to potential mold contamination. The affected products were distributed through TJ Maxx and Marshalls locations.

    Product
    Montreal Fudge brand Chocolate Fudge with Nuts; 5.6oz
    Category
    Food
    Distribution
    40 states
  • HighFDA (Devices)·Z-3270-2024·2024-10-02

    Wallflex and Agile Esophageal Stents Recalled for Catheter Tip Detachment Risk

    Boston Scientific recalls Wallflex FC and Agile Esophageal Stent Systems due to the potential for delivery catheter tip detachment during use. Approximately 145 units are affected worldwide.

    Product
    WALLFLEX FC ESO STENT RMV 23X155MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esoph
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3293-2024·2024-10-02

    Boston Scientific Agile Esophageal Over-the-Wire Stent System recalled for catheter tip detachment

    Boston Scientific is recalling the Agile Esophageal Over-the-Wire Stent System due to potential for delivery catheter tip detachment. Seven affected units distributed worldwide are identified by lot numbers 32889020 and 33178844.

    Product
    AGILE ESO FC RMV 23MM X 6.2CM IN 18.5 FR-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3290-2024·2024-10-02

    Esophageal stent systems recalled due to delivery catheter defect

    Boston Scientific has recalled AGILE and WallFlex esophageal stent systems due to potential delivery catheter tip detachment. This defect could occur during stent deployment and affect proper stent placement.

    Product
    AGILE ESO OTW PC 23MM X 15.0CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3226-2024·2024-10-02

    Medical Pacemaker Pack Recalled Due to Sterilization Assurance Failure

    American Contract Systems is recalling the PACEMAKER PACK (Lot 2405112) due to an inoperable chart recorder that prevented confirmation of sterilization requirements. Affected units were distributed in MO, MN, MA, OH, and NE.

    Product
    PACEMAKER PACK, SLCV01J; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3228-2024·2024-10-02

    Endovascular AAA Pack recalled for sterilization assurance failure

    American Contract Systems recalls 360 units of Endovascular AAA Pack due to inability to confirm sterilization requirements were met. An inoperable chart recorder prevented verification that environmental conditions during sterilization met required specifications.

    Product
    Endovascular AAA Pack, ESED52B; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide