The Recall Desk

State

Nevada product recalls

73,096 recalls have nationwide distribution and so reach Nevada. 5,696 additional recalls listed Nevada specifically in their distribution scope.

What the numbers show

Total
5,696
Critical
951
Severe
1,479
Most recent
2026-09-25

Of the 5,696 recalls naming Nevada we have scored against our rubric, 951 (17%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1,479 (26%) are Severe.

These figures cover only the 5,696 recalls whose distribution notice names Nevada specifically. The 73,096 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

9876–9900 of 78792

  • HighFDA (Devices)·Z-3281-2024·2024-10-02

    Boston Scientific Recalls WallFlex Esophageal Stents Due to Catheter Tip Detachment

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems because the delivery catheter tip may detach during use.

    Product
    WALLFLEX PC ESOPH STENT 18/23MM X 103MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophag
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3224-2024·2024-10-02

    Karl Storz Nephroscope Recall Due to Unapproved Reprocessing Instructions

    Karl Storz Endoscopy is recalling 5 nephroscopes because their instructions for use list reprocessing methods not reviewed or approved by the FDA.

    Product
    Karl Storz - Endoskope , REF: 27830KA, Nephroscope for MIP M, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3202-2024·2024-10-02

    Philips MR 7700 MRI Scanner Recall for Loose Screws

    Philips is recalling 7 MR 7700 MRI scanners because loose screws may block the tabletop, potentially delaying diagnosis or causing patient anxiety.

    Product
    MR 7700 -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782120
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3269-2024·2024-10-02

    Boston Scientific Recalls WallFlex Esophageal Stent Systems Due to Catheter Tip Detachment

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems because the delivery catheter tip may detach. The recall covers 176 units distributed worldwide.

    Product
    WALLFLEX FC ESO STENT RMV 23X125MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esoph
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3288-2024·2024-10-02

    Boston Scientific Recalls Agile Esophageal Stent Systems Due to Catheter Detachment Risk

    Boston Scientific is recalling Agile Esophageal Stent Systems because the delivery catheter tip may detach. The recall affects 15 units distributed worldwide.

    Product
    AGILE ESO OTW PC 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3301-2024·2024-10-02

    Cardinal Health Recalls Salem Sump Stomach Tubes Due to Valve Breakage

    Cardinal Health is recalling Salem Sump Dual Lumen Stomach Tubes because the Anti-Reflux Valve may break due to improper use. Labeling changes were made to address the issue.

    Product
    (a) CardinalHealth Salem Sump Dual Lumen Stomach Tube, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888264929, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube, 14 Fr/Ch (4.7 mm) x 48 in. (122 cm), Model #8888264945, Sterile; (c) CardinalHealth Salem Sump Dua
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3230-2024·2024-10-02

    American Contract Systems Recalls Open Shoulder Medical Convenience Kits

    American Contract Systems is recalling 12 Open Shoulder medical convenience kits because a faulty chart recorder prevented verification of sterilization requirements.

    Product
    Open Shoulder, LWOS39L; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3294-2024·2024-10-02

    Boston Scientific Recalls Agile Esophageal Stent Systems Due to Catheter Tip Detachment

    Boston Scientific is recalling Agile Esophageal Stent Systems because the delivery catheter tip may detach. The recall affects 10 units distributed worldwide.

    Product
    AGILE ESO FC RMV 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·24V737000·2024-10-02

    2025 Chevrolet Equinox EV door strikers recalled for possible breakage

    General Motors is recalling 2025 Chevrolet Equinox EV vehicles (along with Blazer and Blazer EV models) because door strikers may not be properly heat-treated and could break, causing doors to open unexpectedly. All affected vehicles remain in dealerships.

    Product
    CHEVROLET — CHEVROLET BLAZER EV, EQUINOX EV, BLAZER
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3201-2024·2024-10-02

    Philips MR 5300 MRI System Recalled for Loose Assembly Screws

    Philips is recalling 68 MR 5300 MRI diagnostic systems worldwide because loose screws in the RF carrier assembly could block equipment movement and delay patient diagnosis.

    Product
    MR 5300 - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782110
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3243-2024·2024-10-02

    Medical device kit recalled due to inability to confirm sterilization assurance

    American Contract Systems recalled its Cysto medical convenience kit (36 units, Lot 2405033) due to a humidity monitoring failure that prevented confirmation of adequate product sterilization.

    Product
    Cysto, FTCY03L; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3221-2024·2024-10-02

    Medical endoscope recalled for unapproved sterilization instructions

    Karl Storz Hopkins Telescope 6 endoscopes are recalled because their Instructions for Use contain reprocessing methods that lack FDA review and approval.

    Product
    Karl Storz - Endoskope , REF: 27293AA, Hopkins Telescope 6, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3191-2024·2024-10-02

    Custom Procedure Kit Sheath Introducers Mislabeled by Size

    Merit Medical Systems is recalling Custom Procedure Kits (REF: K12T-11077) with 7F sheath introducers that are incorrectly labeled as 7.5F, a labeling discrepancy affecting 24 units distributed nationwide.

    Product
    Custom Procedure Kit, REF: K12T-11077
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3226-2024·2024-10-02

    Medical Pacemaker Pack Recalled Due to Sterilization Assurance Failure

    American Contract Systems is recalling the PACEMAKER PACK (Lot 2405112) due to an inoperable chart recorder that prevented confirmation of sterilization requirements. Affected units were distributed in MO, MN, MA, OH, and NE.

    Product
    PACEMAKER PACK, SLCV01J; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3247-2024·2024-10-02

    BD Phoenix diagnostic systems may misidentify E. coli in clinical samples

    The BD Phoenix 100 system may misidentify Escherichia coli in clinical samples, potentially leading to misdiagnosis and inappropriate patient treatment. Becton Dickinson is recalling 764 units distributed worldwide.

    Product
    BD Phoenix 100-Intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria. The BD Phoenix System provides rapid results for most aerobic and facultative anaerobic Grampositive bacteria. Catalog Number: 448100
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3199-2024·2024-10-02

    MRI Diagnostic Device Recalled for Loose Assembly Screws Blocking Equipment

    Philips Ingenia Elition S MRI machines (10 units) are recalled because assembly screws may come loose and interfere with the horizontal tabletop, potentially delaying diagnosis and causing patient anxiety.

    Product
    Ingenia Elition S -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782150 2) 782106
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3271-2024·2024-10-02

    WallFlex Esophageal Stent System Recalled for Potential Catheter Tip Detachment

    Boston Scientific Corporation is recalling WallFlex Esophageal Stent Systems and Agile Esophageal Over-the-Wire Stent Systems due to potential delivery catheter tip detachment during insertion procedures.

    Product
    WALLFLEX FC ESO STENT RMV LL 18X103- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esop
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3262-2024·2024-10-02

    Airvo 3 Respiratory Device Recalled for Software Issue Affecting Oxygen Delivery

    Fisher & Paykel's Airvo 3 respiratory device has a software defect that may cause it to deliver only room air instead of oxygen during certain alarm conditions, potentially causing oxygen desaturation and hypoxia in patients.

    Product
    PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and
    Category
    Medical Device
    Distribution
    29 states
  • HighFDA (Devices)·Z-3273-2024·2024-10-02

    Wallflex and Agile Esophageal Stent Systems Recalled Due to Catheter Tip Detachment Risk

    Boston Scientific is recalling the Wallflex Esophageal Stent System and Agile Esophageal Over the Wire Stent System due to potential delivery catheter tip detachment during placement. Wallflex devices totaling 115 units are affected.

    Product
    WALLFLEX FC ESO STENT RMV LL 23X125-Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esoph
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3249-2024·2024-10-02

    IntelliVue Patient Monitors recalled for electrical grounding defect

    Philips is recalling approximately 545 IntelliVue Patient Monitor units with a broken ground bolt that may cause loss of electrical grounding, potentially affecting the device's electromagnetic immunity and emission.

    Product
    IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3298-2024·2024-10-02

    Origio Sperm Wash Medium Recalled for Low Sodium Pyruvate Concentration

    CooperSurgical is recalling Origio Sperm Wash Medium due to low sodium pyruvate concentration. The deficiency could compromise gamete development in assisted reproductive procedures.

    Product
    Origio Sperm Wash Medium Pack of Five 60ml bottles - Intended for washing of spermatozoa. Part Number: 84055060D
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3237-2024·2024-10-02

    Medical Convenience Kit Recalled for Failed Sterilization Assurance Validation

    American Contract Systems, Inc. is recalling the Hand Pack AKHD97C medical convenience kit because an inoperable chart recorder caused humidity readings to fall outside specifications, preventing confirmation of sterilization assurance.

    Product
    Hand Pack, AKHD97C; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3286-2024·2024-10-02

    WallFlex and Agile esophageal stent systems recalled for delivery catheter tip detachment

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems due to potential for delivery catheter tip detachment during insertion.

    Product
    WALLFLEX PC ESOPH STENT 23/28MM X 155MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophag
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3209-2024·2024-10-02

    Ultrasonic Cleaner Solution Recalled for Bacterial Contamination and Infection Risk

    Ultrasonic cleaner solutions contaminated with pathogenic bacteria may cause serious infections if used in medical or dental procedures, and may leak or become bloated. Patients and healthcare workers handling the products are at risk.

    Product
    Good Vibrations Ultrasonic Cleaner Solution, JorVet J0382G Ultrasonic Instrument Cleaning Solution, SpecClean Ultrasonic Cleaner Solution, Models: GVUS128-1, GVUS128-1EA, GVUS550-1, GVUS24-1, GVUS24-1EA, JGVUS128-1, VGVUSNP128-1, VGVUSNP128-1EA
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3293-2024·2024-10-02

    Boston Scientific Agile Esophageal Over-the-Wire Stent System recalled for catheter tip detachment

    Boston Scientific is recalling the Agile Esophageal Over-the-Wire Stent System due to potential for delivery catheter tip detachment. Seven affected units distributed worldwide are identified by lot numbers 32889020 and 33178844.

    Product
    AGILE ESO FC RMV 23MM X 6.2CM IN 18.5 FR-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concu
    Category
    Medical Device
    Distribution
    Distributed nationwide