Most US product recalls have nationwide distribution — the manufacturer ships into every state and the recall sweeps the same way. We include those in this listing because they affect North Dakota consumers even though the source notice doesn’t name the state. State-specific recalls are usually for products with regional distribution (a regional dairy, a local store brand) or where the agency was able to narrow the affected shipments to specific destinations.

If you’re checking whether a recall affects your purchase, the recall detail page lists the exact production lot codes, UPCs, and best-by dates. Match those against your package — distribution scope is a starting filter, not the final answer.

12726–12750 of 20305

  • SevereFDA (Devices)·Z-1735-2023·2023-06-28

    Implantable defibrillators risk energy output failure during high voltage therapy

    Certain Medtronic cardiac defibrillators may produce reduced or no energy output during high voltage therapy. The defect involves a specific feedthrough component used in the manufacturing process.

    Product
    CRT-D DTBA2D1 VIVA XT IS1/DF1 INTL, Model Number DTBA2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1758-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver Therapeutic Shock

    Medtronic implantable cardioverter defibrillators with a specific feedthrough may produce reduced or no energy output during high-voltage therapy, potentially failing to deliver life-saving shocks.

    Product
    CRTD DTMA2D4 CLARIA MRI OUS DF4, Model Number DTMA2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1715-2023·2023-06-28

    Medtronic Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic is recalling approximately 10,900 ICD and CRT-D devices that may fail to deliver high-voltage therapy due to a feedthrough component defect. No illnesses reported, but the rare defect poses risk of inadequate device function during critical treatment.

    Product
    ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1740-2023·2023-06-28

    Implantable Defibrillators Risk Loss of Therapy from Feedthrough Defect

    Medtronic recalls certain CRT-D VIVA QUAD XT GOLD implantable defibrillators due to a rare risk of reduced or no electrical output during therapy. Affected devices contain a specific glassed feedthrough component.

    Product
    CRT-D VIVA QUAD XT GOLD CTD, Model Number DTBA2QQG; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1760-2023·2023-06-28

    Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Certain Medtronic implantable cardioverter defibrillators (ICDs and CRT-Ds) may fail to deliver critical therapy due to a manufacturing defect in the feedthrough. This rare issue affects 23,223 devices.

    Product
    CRTD DTMA2QQ CLARIA MRI QUAD OUS DF4, Model Number DTMA2QQ; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1714-2023·2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential Reduced Energy Output

    Medtronic implantable cardioverter defibrillators (ICDs) and CRT-Ds with specific feedthrough designs may produce reduced or insufficient energy during high voltage therapy, affecting the device's ability to treat heart arrhythmias.

    Product
    ICD-DR DDMB2D4 EVERA MRI XT IS-1/DF4 INT, Model Number DDMB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1767-2023·2023-06-28

    Implantable Defibrillators Risk Therapy Failure Due to Feedthrough Defect

    Certain Medtronic implantable cardioverter defibrillators may fail to deliver high-voltage therapy due to a specific feedthrough defect. The 4,767 affected units were distributed nationwide and worldwide.

    Product
    CRTD DTMB2Q1 AMPLIA MRI QUAD OUS DF1, Model Number DTMB2Q1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1727-2023·2023-06-28

    Implantable cardioverter defibrillators may output reduced or no energy during therapy

    Medtronic ICDs and CRT-Ds with specific feedthroughs may fail to deliver adequate high-voltage therapy. This FDA Class I recall affects 487 units distributed nationwide and worldwide.

    Product
    ICD CROME DR MRI IS1 DF1, Model Number DDPC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1739-2023·2023-06-28

    Implantable Cardioverter Defibrillators Risk Energy Output Failure

    Medtronic Inc. is recalling certain implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) due to a rare potential for reduced or no-energy output during high voltage therapy. Affected devices contain a specific glassed feedthrough.

    Product
    CRT-D DTBA2QQ VIVA QUAD XT IS4/DF4 INTL, Model Number DTBA2QQ; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1705-2023·2023-06-28

    Implantable Cardioverter Defibrillators Risk Reduced Energy Output During Therapy

    Medtronic is recalling 522 implantable cardioverter defibrillators that may deliver reduced or no energy during high-voltage therapy due to a feedthrough defect.

    Product
    ICD-DR DDBB1D4 EVERA XT IS1/DF4 US, Model Number DDBB1D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1783-2023·2023-06-28

    Implantable Defibrillator Models May Fail to Deliver Shock Therapy

    Medtronic implantable cardiac defibrillators may fail to deliver electrical therapy during treatment due to a defective feedthrough component. Approximately 1,878 devices are affected.

    Product
    CRTD COBALT HF QUAD MRI IS4 DF1, Model Number DTPB2Q1 ; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1722-2023·2023-06-28

    Implantable cardioverter defibrillators recalled for potential therapy delivery failure

    Medtronic implantable cardioverter defibrillators (ICDs) with a specific glassed feedthrough are recalled due to a rare potential for reduced or no energy output during high voltage therapy.

    Product
    ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1799-2023·2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential Energy Output Failure

    Medtronic's implantable cardioverter defibrillators may fail to deliver high-voltage therapy due to a feedthrough defect. Approximately 6,821 devices are affected worldwide.

    Product
    ICD-VR DVBC3D1 EVERA S IS1/DF1 GLOB, Model Number DVBC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1804-2023·2023-06-28

    Implantable Cardioverter Defibrillators Risk Failure to Deliver Therapy

    Medtronic implantable cardioverter defibrillators with a specific feedthrough defect may rarely fail to deliver high-voltage therapy. Approximately 34,350 devices distributed worldwide are affected.

    Product
    ICD-VR DVFB1D4 VISIA MRI AF US DF4, Model Number DVFB1D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1729-2023·2023-06-28

    Medtronic CRT-D Implants Recalled for Rare Energy Output Failure Risk

    Medtronic is recalling certain CRT-D (cardiac resynchronization therapy defibrillator) models due to a rare risk of reduced or no energy output during high-voltage therapy. The FDA has classified this as a Class I recall affecting 1,473 units.

    Product
    CRT-D DTBA1D1 VIVA XT IS1/DF1 US, Model Number DTBA1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1753-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver High-Voltage Therapy Energy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy due to a feedthrough defect. Approximately 11,794 units worldwide are affected.

    Product
    CRTD DTMA1D1 CLARIA MRI US DF1, Model Number DTMA1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1775-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillator Model DTPA2D1G Therapy Output Failure

    Medtronic Inc. is recalling five units of the CRTD DTPA2D1G COBALT XT HF QUAD OUS implantable cardioverter defibrillator due to a rare potential for reduced or no energy output during high-voltage defibrillation therapy.

    Product
    CRTD DTPA2D1G COBALT XT HF QUAD OUS, Model Number DTPA2D1G; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1803-2023·2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential High Voltage Therapy Failure

    Medtronic implantable cardioverter defibrillators (ICDs) with a specific feedthrough component may have reduced or absent energy output during high voltage therapy. The FDA Class I recall affects 12,099 units distributed nationwide and worldwide.

    Product
    ICD-VR DVFB1D1 VISIA AF MRI US DF1, Model Number DVFB1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1810-2023·2023-06-28

    Implantable cardioverter defibrillators may fail to deliver therapy

    Medtronic is recalling certain implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) with specific glassed feedthroughs that may deliver reduced or no energy output during high voltage therapy.

    Product
    ICD-VR DVMB2D1 EVERA MRI XT OUS DF1, Model Number DVMB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1723-2023·2023-06-28

    ICD COBALT XT Defibrillators Recalled for Potential Reduced Energy Output

    Medtronic recalled 22,114 ICD COBALT XT defibrillators due to a rare potential for reduced or no energy output during high-voltage therapy delivery caused by a specific feedthrough defect.

    Product
    ICD COBALT XT DR MRI IS1 DF4, Model Number DDPA2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1779-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver Critical Heart Therapy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy delivery due to a defective feedthrough component. No injuries have been reported, but patients may not receive proper treatment for life-threatening heart rhythms.

    Product
    CRTD COBALT XT HF QUAD MRI IS4 DF4, Model Number DTPA2QQ; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1814-2023·2023-06-28

    Medtronic ICD-VR PRIMO MRI Defibrillators: Potential Therapy Delivery Failure

    Certain Medtronic implantable cardioverter defibrillators and CRT-Ds may fail to deliver adequate energy during high-voltage therapy due to a feedthrough component defect. This affects 5,215 units distributed worldwide.

    Product
    ICD-VR DVMD3D1 PRIMO MRI, Model Number DVMD3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1774-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail During High-Voltage Therapy

    Some Medtronic implantable defibrillators contain a specific feedthrough that may rarely cause reduced or no energy output during high-voltage therapy. The recall affects approximately 4,151 units distributed nationwide.

    Product
    CRTD COBALT XT HF MRI IS1 DF1, Model Number DTPA2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1795-2023·2023-06-28

    Implantable Defibrillators May Deliver Reduced or No Energy During Therapy

    Certain Medtronic implantable cardioverter defibrillators with a specific feedthrough may produce reduced or no energy output during high voltage therapy. The rare defect affects 16 units distributed nationwide.

    Product
    ICD-VR DVBB1D1 EVERA XT IS1/DF1 US, Model Number DVBB1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide