  • SevereFDA (Devices)·Z-1578-2026·2026-04-01

    DuraMax hemodialysis catheter sheath introducer may not split as intended

    Merit Medical is recalling DuraMax Chronic Hemodialysis Catheters because a design defect in the 16F dual-valved splittable sheath introducer may prevent it splitting as intended.

    Product
    DuraMax Chronic Hemodialysis Catheter, REF: H787103028015/A, H787103028025/A, H787103028031/A, H787103028035/A, H787103028041/A, H787103028045/A, H787103028051/A, H787103028055/A, H787103028061/A, H787103028075/A, H787103028085/A, H787103028095/A, H787103028165/A, H787103028181/A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1615-2026·2026-04-01

    Trilogy Evo O2 ventilator recalled as Class I over tidal volume

    Philips Respironics is recalling the Trilogy Evo O2 ventilator because using non-pneumatic nebulizers may cause a discrepancy between the set tidal volume and the volume the patient receives.

    Product
    Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1614-2026·2026-04-01

    Trilogy Evo ventilator recalled as Class I over tidal volume

    Philips Respironics is recalling the Trilogy Evo ventilator because using non-pneumatic nebulizers may cause a discrepancy between the set tidal volume and the volume the patient receives.

    Product
    Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1616-2026·2026-04-01

    Trilogy EV300 ventilator recalled as Class I over tidal volume

    Philips Respironics is recalling the Trilogy EV300 ventilator because using non-pneumatic nebulizers may cause a discrepancy between the set tidal volume and the volume the patient receives.

    Product
    Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1566-2026·2026-04-01

    Erbe flexible cryoprobe 20402-401 recalled as Class I over rupture

    Erbe USA is recalling the Flexible Cryoprobe REF 20402-401 because the probes may rupture or burst during activation. The FDA has classified this recall as Class I.

    Product
    Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1602-2026·2026-04-01

    I.T.S. distal humeral plates heat more in MRI than stated

    I.T.S. GmbH is recalling Distal Humeral Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N, 373
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1595-2026·2026-04-01

    I.T.S. calcaneus repair plate heats more in MRI than stated

    I.T.S. GmbH is recalling the FR.O.H. Calcaneus Repair System anterolateral calcaneus plate because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1606-2026·2026-04-01

    I.T.S. headless compression screws heat more in MRI than stated

    I.T.S. GmbH is recalling the HCS Headless Compression Screw System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1598-2026·2026-04-01

    I.T.S. straight and forearm plates heat more in MRI than stated

    I.T.S. GmbH is recalling Straight Plates with Angular Stability, including forearm and compression plates, because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Pla
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1593-2026·2026-04-01

    I.T.S. clavicle plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling Clavicula Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm; Article
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0598-2026·2026-04-01

    Kroger Chinese Inspiration Vegetable Fried Rice recalled for glass

    Ajinomoto Foods North America Inc. is recalling Kroger Chinese Inspiration Vegetable Fried Rice due to the presence of glass fragments varying from 1–3 cm long and 2–4 mm wide.

    Product
    Item 81153 Kroger Chinese Inspiration Vegetable Fried Rice - Blend of long grain rice assorted vegetables & egg. Net wt. 22oz. Retail box UPC 011110041623. Product packaging is flexible bags in a master case or flexible bags in cartons in a master case. There are 6/22oz. boxes o
    Category
    Food
    Distribution
    43 states
  • HighFDA (Devices)·Z-1633-2026·2026-04-01

    Straumann angled custom abutments made from wrong bar material

    Straumann is recalling Custom Abutment AS Ti for Straumann RC because a material mix-up meant titanium abutments were manufactured from TAN bar material instead of Titanium Grade 4.

    Product
    Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1639-2026·2026-04-01

    Olympus Everest bipolar cutting forceps jaws may break from defective welds

    Olympus is recalling Everest Bipolar Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achi
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1613-2026·2026-04-01

    Olympus OER-Pro reprocessor no longer compatible with two valve models

    Olympus is recalling the OER-Pro Endoscope Reprocessor because the MAJ-1443 and MAJ-1444 air/water valves are no longer reprocessing compatible with the OER-Pro and OER-Elite reprocessors.

    Product
    Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0405-2026·2026-04-01

    Teva octreotide acetate 20 mg recalled over sterility assurance deficiencies

    Teva is recalling nine lots of Octreotide Acetate for Injectable Suspension 20 mg after quality system deficiencies were identified during an FDA inspection at the contract manufacturer.

    Product
    OCTREOTIDE ACETATE — OCTREOTIDE ACETATE (OCTREOTIDE ACETATE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0615-2026·2026-04-01

    Savannah Bee Company Honey BBQ Sauce Mustard Undeclared Allergens

    Savannah Bee Company is recalling Honey BBQ Sauce Mustard bottles containing an undeclared different flavor product with soy and wheat allergens not listed on the label.

    Product
    SAVANNAH BEE COMPANY HONEY BBQ SAUCE MUSTARD NET 16 FL OZ (453g), PACKED IN BOTTLES. INGREDIENTS: MUSTARD (DISTILLED VINEGAR, WATER, MUSTARD SEED, SALT, SPICES (INCLUDING PAPRIKA AND TUMERIC), NATURAL FLAVOR, GARLIC), HONEY, WATER, DISTILLED VINEGAR, SUGAR, TOMATO CONCENTRATE, HA
    Category
    Food
    Distribution
    45 states
  • HighFDA (Devices)·Z-1612-2026·2026-04-01

    Olympus OER-Elite reprocessor no longer compatible with two valve models

    Olympus is recalling the OER-Elite Endoscope Reprocessor because the MAJ-1443 and MAJ-1444 air/water valves are no longer reprocessing compatible with the OER-Pro and OER-Elite reprocessors.

    Product
    Olympus Endoscope Reprocessor OER-Elite. Model Number: N5789340.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1632-2026·2026-04-01

    Straumann custom abutments made from wrong bar material after mix-up

    Straumann is recalling Custom Abutment Ti for Straumann RC because a material mix-up meant titanium abutments were manufactured from TAN bar material instead of Titanium Grade 4.

    Product
    Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1605-2026·2026-04-01

    I.T.S. foot locking plates heat more in MRI than stated

    I.T.S. GmbH is recalling the FLS Foot Locking Plates System and HOL Hallux Osteotomy Locking Plate because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18, 373
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0600-2026·2026-04-01

    Ajinomoto Vegetable Gyoza Dumplings recalled for glass fragments

    Ajinomoto Foods North America Inc. is recalling Vegetable Gyoza Dumplings (UPC 071757055483) due to the presence of glass fragments in the product, which could pose a choking or injury hazard if ingested.

    Product
    Item 6430143 Ajinomoto Vegetable Gyoza Dumplings 12ct. Net wt. 8.92oz. per package. Retail package UPC: 071757055483. Product packaging is flexible bags in a master case or flexible bags in cartons in a master case. There are 20/8.92oz. packages or 11.15 lbs. per case. Carton UP
    Category
    Food
    Distribution
    43 states
  • HighFDA (Food)·H-0595-2026·2026-04-01

    Ajinomoto Golden Tiger Vegetable Fried Rice recalled for glass fragments

    Ajinomoto Foods North America Inc. is recalling Ajinomoto Golden Tiger Vegetable Fried Rice due to the presence of glass fragments varying from 1–3 cm long and 2–4 mm wide in the product.

    Product
    Item 5654390 Ajinomoto Golden Tiger Vegetable Fried Rice - Long grain white rice and crispy vegetables in a savory sauce. Net wt. 3lbs per bag. Retail bag UPC 076366565439. Product packaging is flexible bags in a master case or flexible bags in cartons in a master case. There ar
    Category
    Food
    Distribution
    43 states
  • HighFDA (Devices)·Z-1644-2026·2026-04-01

    Olympus PKS cutting forceps jaws may break from defective welds

    Olympus is recalling PKS Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visuali
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1640-2026·2026-04-01

    Olympus PK cutting forceps jaws may break from defective welds

    Olympus is recalling PK Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Number: PK-CF0533 Product Description: An assembly of devices intended to generate and apply radio-frequency (RF) alternating current to soft tissues for cutting and coagulation dur
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1610-2026·2026-04-01

    Olympus endoscope suction valve no longer compatible with named reprocessors

    Olympus is recalling the MAJ-1443 Endoscope Suction Valve because the MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

    Product
    Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1609-2026·2026-04-01

    Beckman Coulter MicroScan gram-negative panels may contain contaminated wells

    Beckman Coulter is recalling one lot of MicroScan Neg MIC 56 panels due to the likely presence of contamination in one or more wells.

    Product
    MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
    Category
    Medical Device
    Distribution
    Distributed nationwide