The Recall Desk

State

Mississippi product recalls

72,933 recalls have nationwide distribution and so reach Mississippi. 5,349 additional recalls listed Mississippi specifically in their distribution scope.

What the numbers show

Total
5,349
Critical
923
Severe
1,126
Most recent
2026-09-09

Of the 5,349 recalls naming Mississippi we have scored against our rubric, 923 (17%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1,126 (21%) are Severe.

These figures cover only the 5,349 recalls whose distribution notice names Mississippi specifically. The 72,933 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

2701–2725 of 78282

  • HighFDA (Devices)·Z-1609-2026·2026-04-01

    Beckman Coulter MicroScan gram-negative panels may contain contaminated wells

    Beckman Coulter is recalling one lot of MicroScan Neg MIC 56 panels due to the likely presence of contamination in one or more wells.

    Product
    MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1599-2026·2026-04-01

    I.T.S. cannulated screw system heats more in MRI than stated

    I.T.S. GmbH is recalling the I.T.S. Screw System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous Screw, Cannulated; Article Numbers: 31404-16, 31404-18, 31404-20, 31404-22, 31404-24, 31404-26, 31404-28, 31404-30, 31404-32, 31404-34, 31404-36, 31404-38, 31404-40, 31404-42, 314
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·26V209000·2026-04-01

    Audi Q3 recalled over remote sunroof closing from out of sight

    Volkswagen Group of America is recalling 6,077 Audi Q3 vehicles because incorrect software may allow remote sunroof closing from a distance where the operator cannot clearly see the sunroof.

    Product
    AUDI — AUDI Q3
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1643-2026·2026-04-01

    Olympus PKS cutting forceps 920000PK jaws may break from welds

    Olympus is recalling PKS Cutting Forceps with cord, model 920000PK, because a supplier did not adequately validate the welding process and defective welds can cause the jaw to break.

    Product
    Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0601-2026·2026-04-01

    Ajinomoto Vegetable Gyoza Dumplings Recalled for Glass Contamination

    Ajinomoto Foods North America Inc. is recalling Ajinomoto Vegetable Gyoza Dumplings due to the presence of glass fragments in the product. The recalled dumplings were distributed nationwide.

    Product
    Item 6430153 Ajinomoto Vegetable Gyoza Dumplings. Net wt. 24.7oz. per bag. Retail box UPC: 071757010109. Product packaging is flexible bags in a master case or flexible bags in cartons in a master case. There are 8/24.7oz bags or 12.34lb. per case. Case UPC 10071757010106. All
    Category
    Food
    Distribution
    43 states
  • HighFDA (Devices)·Z-1598-2026·2026-04-01

    I.T.S. straight and forearm plates heat more in MRI than stated

    I.T.S. GmbH is recalling Straight Plates with Angular Stability, including forearm and compression plates, because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Pla
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1608-2026·2026-04-01

    I.T.S. pelvic reconstruction plates heat more in MRI than stated

    I.T.S. GmbH is recalling the Pelvic Reconstruction System PRS RX and PRS Phoenix because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: 21218-16, 21219-16. 2. Anterior Column Plate; Article Numbers: 21218-14, 21219-14. 3. Biplanar 2-C
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1592-2026·2026-04-01

    I.T.S. volar radius plates heat more in MRI than stated

    I.T.S. GmbH is recalling Volar Radius Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1644-2026·2026-04-01

    Olympus PKS cutting forceps jaws may break from defective welds

    Olympus is recalling PKS Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visuali
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1602-2026·2026-04-01

    I.T.S. distal humeral plates heat more in MRI than stated

    I.T.S. GmbH is recalling Distal Humeral Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N, 373
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0615-2026·2026-04-01

    Savannah Bee Company Honey BBQ Sauce Mustard Undeclared Allergens

    Savannah Bee Company is recalling Honey BBQ Sauce Mustard bottles containing an undeclared different flavor product with soy and wheat allergens not listed on the label.

    Product
    SAVANNAH BEE COMPANY HONEY BBQ SAUCE MUSTARD NET 16 FL OZ (453g), PACKED IN BOTTLES. INGREDIENTS: MUSTARD (DISTILLED VINEGAR, WATER, MUSTARD SEED, SALT, SPICES (INCLUDING PAPRIKA AND TUMERIC), NATURAL FLAVOR, GARLIC), HONEY, WATER, DISTILLED VINEGAR, SUGAR, TOMATO CONCENTRATE, HA
    Category
    Food
    Distribution
    45 states
  • HighFDA (Devices)·Z-1600-2026·2026-04-01

    I.T.S. pelvic reconstruction screws heat more in MRI than stated

    I.T.S. GmbH is recalling the Pelvic Reconstruction System fixation screw and short spike because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1597-2026·2026-04-01

    I.T.S. olecranon plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling Olecranon Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1633-2026·2026-04-01

    Straumann angled custom abutments made from wrong bar material

    Straumann is recalling Custom Abutment AS Ti for Straumann RC because a material mix-up meant titanium abutments were manufactured from TAN bar material instead of Titanium Grade 4.

    Product
    Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0406-2026·2026-04-01

    Teva octreotide acetate injection recalled over sterility assurance deficiencies

    Teva is recalling twelve lots of Octreotide Acetate for Injectable Suspension after quality system deficiencies were identified during an FDA inspection at the contract manufacturer.

    Product
    OCTREOTIDE ACETATE — OCTREOTIDE ACETATE (OCTREOTIDE ACETATE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1638-2026·2026-04-01

    Olympus Everest bipolar forceps model 3005 jaws may break

    Olympus is recalling Everest Bipolar Cutting Forceps model 3005 because a supplier did not adequately validate the welding process, and defective welds can cause the jaw to break during clinical use.

    Product
    Brand Name: Olympus Everest Bipolar Cutting Forceps, 5 mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3005 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved u
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1610-2026·2026-04-01

    Olympus endoscope suction valve no longer compatible with named reprocessors

    Olympus is recalling the MAJ-1443 Endoscope Suction Valve because the MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

    Product
    Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1595-2026·2026-04-01

    I.T.S. calcaneus repair plate heats more in MRI than stated

    I.T.S. GmbH is recalling the FR.O.H. Calcaneus Repair System anterolateral calcaneus plate because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·26V207000·2026-04-01

    Mack Recalls 2025-2027 Trailers Due to Turn Signal Software Defect

    Mack Trucks is recalling 2025-2027 Pioneer and 2027 Anthem trailers because turn signals may not function properly, increasing crash risk.

    Product
    MACK — MACK PIONEER PR (4), ANTHEM AN (4)
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1632-2026·2026-04-01

    Straumann custom abutments made from wrong bar material after mix-up

    Straumann is recalling Custom Abutment Ti for Straumann RC because a material mix-up meant titanium abutments were manufactured from TAN bar material instead of Titanium Grade 4.

    Product
    Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0596-2026·2026-04-01

    Ajinomoto Ling Ling Fried Rice Recalled for Glass Contamination

    Ajinomoto Foods North America is recalling Item 81097 Ling Ling Restaurant Style Fried Rice due to glass fragments measuring 1-3 cm long and 2-4 mm wide found in the product. Affected units with Best By dates 9/21/2026 through 3/8/2027 were distributed across 40 U.S. states.

    Product
    Item 81097 Ajinomoto Ling Ling Restaurant Style Fried Rice Savory Vegetable - A Chinese Style Fried Rice with Edamame, Carrots, Fire Roasted Corn & Red Bell Peppers Prepared with Sweet Soy Sauce Infused Rice. Net wt. 17oz. Retail box UPC 010878810915. Product packaging is flexib
    Category
    Food
    Distribution
    43 states
  • HighFDA (Devices)·Z-1590-2026·2026-04-01

    GE Imactis CT navigation may show needle trajectory that differs from actual

    GE Medical Systems is recalling the Imactis CT-Navigation System because of a potential discordance between the simulated needle trajectory and tip position and the actual trajectory and tip position.

    Product
    Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1635-2026·2026-04-01

    Medtronic MiniMed 780G software defects may cause insulin delivery inaccuracies

    Medtronic is recalling MiniMed 780G insulin pumps on software 6.60 and 6.61 because three software defects could result in hypoglycemia or hyperglycemia through inaccurate insulin delivery.

    Product
    MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1613-2026·2026-04-01

    Olympus OER-Pro reprocessor no longer compatible with two valve models

    Olympus is recalling the OER-Pro Endoscope Reprocessor because the MAJ-1443 and MAJ-1444 air/water valves are no longer reprocessing compatible with the OER-Pro and OER-Elite reprocessors.

    Product
    Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1596-2026·2026-04-01

    I.T.S. pilon and tibia plates heat more in MRI than stated

    I.T.S. GmbH is recalling Pilon Plates with Angular Stability and Anterolateral Tibia Plates because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 2109
    Category
    Medical Device
    Distribution
    Distributed nationwide