The Recall Desk

State

Minnesota product recalls

73,833 recalls have nationwide distribution and so reach Minnesota. 7,851 additional recalls listed Minnesota specifically in their distribution scope.

What the numbers show

Total
7,851
Critical
1,137
Severe
2,246
Most recent
2026-11-04

Of the 7,851 recalls naming Minnesota we have scored against our rubric, 1,137 (14%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2,246 (29%) are Severe.

These figures cover only the 7,851 recalls whose distribution notice names Minnesota specifically. The 73,833 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

56326–56350 of 81684

  • HighFDA (Devices)·Z-0463-2016·2015-12-30

    Boston Scientific Recalls IntellaTip MiFi XP Catheters Due to Tip Bends

    Boston Scientific is recalling IntellaTip MiFi XP catheters because the rate of distal tip bends and kinks exceeded anticipated levels during routine monitoring.

    Product
    IntellaTip MiFi XP Asymmetric (N4) Curve 7/110/2.5/8-10 OUS; Material Number: M004EPM4790N40; Catalog number: EPM4790N4; Cardiac: The IntellaTip MiFi XP Catheter is indicated for interruption of accessory atrioventricular conduction pathways associated with tachycardia…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0460-2016·2015-12-30

    Siemens Reagent Management System Recalled for Electrical Hazard

    Siemens is recalling Reagent Management System (RMS) modules for the Dimension RxL Max clinical chemistry system because some units have missing safety covers on electrical termination blocks.

    Product
    Reagent Management System (RMS) Dimension RxL Max w/o HM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0552-2016·2015-12-30

    Airways Development Recalls WaterPAP Positive Airway Pressure Devices

    Airways Development LLC is recalling 1,464 WaterPAP Positive Airway Pressure Devices due to a reported canister leak.

    Product
    Airways Development LLC WaterPAP Positive Airway Pressure Device Reference #WP-7700.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0461-2016·2015-12-30

    Siemens Reagent Management System Recalled for Electrical Safety Hazard

    Siemens is recalling 151 Reagent Management System units because missing safety covers expose electrical termination blocks, creating a shock hazard.

    Product
    Reagent Management System (RMS) Dimension RxL Max HM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0471-2016·2015-12-30

    Biomet Black Mamba Suture Passer Recalled for Weld Defect

    Biomet is recalling 50 Black Mamba Suture Passers because a weld defect may cause the instrument's jaw to fail to actuate properly.

    Product
    Black Mamba Suture Passer
    Category
    Medical Device
    Distribution
    20 states
  • HighFDA (Devices)·Z-0423-2016·2015-12-30

    Aesculap Flexible Bone Awl Recalled for Potential Shaft Breakage

    Aesculap, Inc. is recalling 41 Flexible Bone Awls (SJ607R) because the flexible shaft may break during surgery, posing a risk of fragments remaining in patients.

    Product
    Flexible Bone Awl; SJ607R; NON STERILE; B Braun Aesculap; Aesculap AG: These instruments are intended for use during the implantations of the A-Space SIBD and Arcadius XP L systems.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0456-2016·2015-12-30

    BD Vacutainer K2 EDTA Blood Collection Tubes Recalled for Stopper Defect

    Becton Dickinson is recalling 4.8 million blood collection tubes because stoppers may not be fully inserted, posing a risk of blood exposure.

    Product
    BD Vacutainer K2 EDTA (K2E) 5.4mg Plus Blood Collection Tubes. Product Usage: BD Vacutainer¿ tubes are used to collect, transport and process blood in a closed system. These tubes are used for testing serum, plasma or whole blood in the clinical laboratory. The BD…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0458-2016·2015-12-30

    Siemens Reagent Management System Recalled for Electrical Hazard

    Siemens is recalling 106 Reagent Management System units because a missing safety cover exposes an electrical hazard on the compressor.

    Product
    Reagent Management System (RMS) Dimension EXL with LM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0453-2016·2015-12-30

    Zimmer Universal Power System Handpieces Recalled for Malfunction

    Zimmer, Inc. is recalling 89 Universal Power System handpieces that may start by themselves when the power source is connected.

    Product
    Zimmer UNIVERSAL Power System Loaner and Modular Electric/Battery Single Trigger Handpiece. Rx only Made in Switzerland
    Category
    Medical Device
    Distribution
    32 states
  • HighCPSC·16068·2015-12-30

    Origin8 Recalls Folding Bicycles Due to Frame Breakage and Fall Hazard

    Origin8 is recalling F1, F3, and F7 folding bicycles because the frame welds can crack or fail, posing a fall hazard. Consumers should stop using the bikes and return them for a free replacement.

    Product
    Origin8 folding bicycles
    Category
    Consumer Product
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0459-2016·2015-12-30

    Siemens Reagent Management System Recalled for Electrical Hazard

    Siemens is recalling four Reagent Management System units because a missing safety cover exposes an electrical termination block, creating a shock hazard.

    Product
    Reagent Management System (RMS) Dimension EXL with LM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0467-2016·2015-12-30

    Siemens MAGNETOM MRI Systems Recalled for Quench Line Installation Errors

    Siemens is recalling MAGNETOM MRI systems due to potential quench line installation errors that could cause oxygen displacement or cold burns.

    Product
    MAGNETOM systems, magnetic resonance diagnostic devices (MRDD).
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0474-2016·2015-12-30

    Becton Dickinson Recalls Enterococcus Screen Agar QUAD Plates

    Becton Dickinson is recalling 770 units of Enterococcus Screen Agar QUAD Plates due to potential breakthrough growth of Enterococcus faecalis near the end of shelf life.

    Product
    Enterococcus Screen Agar QUAD Plate
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·15V886000·2015-12-30

    Micro Bird Recalls 2015 School Buses for Stop Arm Decal Issue

    Micro Bird is recalling 66 model year 2015 school buses because the stop arm decal may peel off, reducing visibility and increasing crash risk.

    Product
    MICRO BIRD — MICRO BIRD MB II SCHOOL BUS, G5 SCHOOL BUS
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0445-2016·2015-12-30

    CareFusion Recalls Alaris Pump SmartSite Infusion Sets Due to Leakage Risk

    CareFusion is recalling Alaris Pump SmartSite Infusion Sets because disconnection and leakage may occur during infusion.

    Product
    Alaris Pump SmartSite Infusion Set, Model No. 2420-0500. Used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighCPSC·16718·2015-12-30

    KTM Recalls 2015-2016 Freeride 250R Motorcycles Due to Fuel Leak Fire Risk

    KTM North America is recalling about 550 model year 2015 and 2016 Freeride 250R off-road motorcycles because fuel can escape from the tank breather assembly, posing a fire and injury risk.

    Product
    KTM Competition/Closed Course Off-Road Motorcycles
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0452-2016·2015-12-30

    Zimmer Universal Power System Handpieces Recalled for Malfunction

    Zimmer, Inc. is recalling 87 Universal Power System handpieces because a malfunction could cause them to start unexpectedly when connected to a power source.

    Product
    Zimmer UNIVERSAL Power System Loaner and Modular Electric/Battery Double Trigger Handpiece. Rx only Made in Switzerland
    Category
    Medical Device
    Distribution
    32 states
  • HighFDA (Devices)·Z-0450-2016·2015-12-30

    Biomet Recalls A.L.P.S. 4.0mm Cancellous Locking Screws Due to Thread Defect

    Biomet is recalling A.L.P.S. 4.0mm Cancellous Locking Screws because a single-lead thread design may cause screws to sit proud or back out, potentially requiring revision surgery.

    Product
    A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones.
    Category
    Medical Device
    Distribution
    51 states
  • ModerateFDA (Devices)·Z-0545-2016·2015-12-30

    Philips Brilliance iCT CT System Recalled for Perfusion Software Issue

    Philips is recalling 51 Brilliance iCT CT systems because the perfusion scan feature may be unavailable on specific software versions.

    Product
    Brilliance iCT Computed Tomography X-ray system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0508-2016·2015-12-30

    Ameditech Immutest 11 Pnl Drug Screen Cup Recall for Reduced Reactivity

    Ameditech is recalling Immutest 11 Pnl Drug Screen Cups because they may show reduced reactivity for two specific benzodiazepine compounds.

    Product
    Immutest 11 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-11OP
    Category
    Medical Device
    Distribution
    Distributed nationwide