  • HighFDA (Devices)·Z-0340-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 53 mm Item Number: 00-8065-550-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0423-2021·2020-11-11

    3M Red Dot Monitoring Electrode Recall Due to Corrosion Risk

    3M is recalling Red Dot Monitoring Electrodes because corrosion could cause performance failures in the ECG monitoring product.

    Product
    3M Red Dot Monitoring Electrode with Foam Tape - Product Usage: intended to be used by healthcare professionals for ECG monitoring. 50 electrodes per pouch; 20 pouches per shipper box (1000 electrodes total).
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0361-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Cup Hip Prosthesis Components

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 47 mm Number: 00-8065-644-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0357-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling specific acetabular cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 43 mm Number: 00-8065-640-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0349-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 59 mm Number: 00-8065-556-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0314-2021·2020-11-11

    GE Healthcare Recalls 9100c NXT Systems Due to Flow Sensor Defects

    GE Healthcare is recalling 37 9100c NXT systems because flow sensors may have damaged tubes with small punctures or cuts.

    Product
    9100c NXT system - Product Usage: Device is licensed and distributed outside of the US only.
    Category
    Medical Device
    Distribution
    36 states
  • HighFDA (Devices)·Z-0354-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 61 mm Number: 00-8065-558-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0380-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Cup Hip Prosthesis Components

    Zimmer Biomet is recalling 412 acetabular cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 57 mm Number: 00-8065-654-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0324-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Component Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling specific ZCA All Poly Acetabular Cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 45mm Item Number: 00-8065-542-22 Hip prosthesis component
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0391-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 49 mm Number: 00-8065-946-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0374-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Prosthesis Cups Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 acetabular hip prosthesis cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup with Spacers, 53 mm Number: 00-8065-650-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0313-2021·2020-11-11

    GE Datex-Ohmeda Aisys CS2 Anesthesia System Recalled for Damaged Flow Sensors

    GE Healthcare is recalling 101 Aisys CS2 Anesthesia Systems because flow sensors may have damaged tubes, posing a risk to patients.

    Product
    GE Datex-Ohmeda Aisys CS2 Anesthesia System - Product Usage: is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric adult). The device is intended for volume or pressure control ventilation.
    Category
    Medical Device
    Distribution
    36 states
  • HighFDA (Devices)·Z-0327-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Prosthesis Cups Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 acetabular hip prosthesis cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 47 mm Item Number: 00-8065-544-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0350-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 59 mm Number: 00-8065-556-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0362-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 47 mm Number: 00-8065-644-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0386-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup with Spacers, 59 mm Number: 00-8065-656-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0363-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cup Components Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 49 mm Number: 00-8065-646-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0337-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cup components because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 53 mm Item Number: 00-8065-550-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0395-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 57 mm Number: 00-8065-954-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0418-2021·2020-11-11

    Angiodynamics 5F Micro-Introducer Kits Recalled for Incorrect Device Packaging

    Angiodynamics, Inc. is recalling 32 kits of 5F Standard Micro-Introducer Kits because they may contain a 4F Sheath/Dilator component instead of the correct device.

    Product
    Angiodynamics 5F Standard Micro-Introducer Kit - indicated for percutaneous introduction of a guidewire or catheter into the vascular system, Catalog Number: 06597035 UPN: H787065970355
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0311-2021·2020-11-11

    GE Healthcare Recalls Aestiva/5 MRI Anesthesia Systems Due to Flow Sensor Defect

    GE Healthcare is recalling two Aestiva/5 MRI Anesthesia Systems because flow sensors may have damaged tubes with small punctures or cuts.

    Product
    Aestiva/5 MRI Anesthesia System - Product Usage: The Aestiva/5 MRI Anesthesia System provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment. Among those standard Aestiva/5 features is the…
    Category
    Medical Device
    Distribution
    36 states
  • ModerateFDA (Devices)·Z-0402-2021·2020-11-11

    Microbiologics Recalls Klebsiella Pneumoniae Strain Due to Incorrect VIM-1 Status

    Microbiologics Inc is recalling 46 units of a Klebsiella pneumoniae strain because it is VIM-1 negative instead of the expected VIM-1 positive status.

    Product
    Catalog 02145P, Klebsiella pneumoniae derived from NCTC 13439 - Product Usage: intended to be used as controls to verify the performance of assays, reagents or media that are intended to be used in microbial testing for the detection and identification of a cultured…
    Category
    Medical Device
    Distribution
    Distributed nationwide