What the numbers show

Total
5,723
Critical
1,045
Severe
1,228
Most recent
2026-09-02

Of the 5,723 recalls naming Kansas we have scored against our rubric, 1,045 (18%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1,228 (21%) are Severe.

These figures cover only the 5,723 recalls whose distribution notice names Kansas specifically. The 72,936 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

2551–2575 of 78659

  • HighFDA (Devices)·Z-1748-2026·2026-04-08

    Paradigm insulin pump delivery varies with pump height changes

    Medtronic is recalling Paradigm insulin pumps models MMT-712 and MMT-715 because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm Insulin Pump (MMT-712, MMT-715)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1699-2026·2026-04-08

    da Vinci grasping retractor instruments recalled over broken grip cables

    Intuitive Surgical is recalling da Vinci S and Si Grasping Retractor Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Grasping Retractor Instrument Product Name: da Vinci S, Si Grasping Retractor Instrument Model/Catalog Number: 420278 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp en
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1746-2026·2026-04-08

    MiniMed 740G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 740G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1671-2026·2026-04-08

    Philips Allura Xper FD20/15 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD20/15 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/15; System Code: 722058;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0419-2026·2026-04-08

    Prazosin hydrochloride 100-count bottles recalled over nitrosamine impurity

    A further quantity of Prazosin Hydrochloride tablets is being recalled after detection of the nitrosamine impurity N-nitroso-prazosin impurity C above acceptable limits.

    Product
    PRAZOSIN HYDROCHLORIDE — PRAZOSIN HYDROCHLORIDE (PRAZOSIN HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1657-2026·2026-04-08

    Atellica hemoglobin A1c assay may return depressed patient results

    Siemens is recalling Atellica CH Enzymatic Hemoglobin A1c because results may be depressed when the RCRP reagent is processed on the same analyzer, with bias increasing over time.

    Product
    Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1700-2026·2026-04-08

    da Vinci tenaculum forceps recalled over frayed or broken pitch cables

    Intuitive Surgical is recalling da Vinci S and Si Tenaculum Forceps Instruments following increased complaints regarding frayed or broken pitch cables on reusable surgical instruments.

    Product
    Brand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci S, Si Tenaculum Forceps Instrument Model/Catalog Number: 420207 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp end
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1757-2026·2026-04-08

    Burlington Medical wrap aprons may lose radiation attenuation over time

    Burlington Medical is recalling wrap aprons made with Xenolite 800 attenuation material because of the potential for attenuation degradation over time.

    Product
    Burlington Medical, Wrap Aprons.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1686-2026·2026-04-08

    Artoura breast tissue expander infusion set needles may be blunt

    Mentor Texas is recalling Artoura Breast Tissue Expanders because the infusion sets may contain a dull or blunt needle tip that may be difficult to advance or may break.

    Product
    Artoura Breast Tissue Expanders Reference Numbers: SDC-100H SDC-110H SDC-120H SDC-130H SDC-135H SDC-140H SDC-150H SDC-155H Smooth High Profile Tissue Expander, 225cc, 300cc, 375cc, 475cc, 500cc, 600cc, 750cc, 850cc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1655-2026·2026-04-08

    Paradise renal denervation catheter distributed despite being marked for scrapping

    Recor Medical is recalling one lot of the Paradise Ultrasound Renal Denervation System Catheter because a nonconforming product that should have been scrapped was distributed instead.

    Product
    Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0414-2026·2026-04-08

    GNP redness and dry eye relief drops recalled over sterility assurance

    GNP Eye Drops Redness and Dry Eye Relief, containing glycerin and naphazoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    LEADER EYE DROPS REDNESS RELIEF — LEADER EYE DROPS REDNESS RELIEF (GLYCERIN, NAPHAZOLINE HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1744-2026·2026-04-08

    MiniMed 700G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 700G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1687-2026·2026-04-08

    Artoura ultra high profile expander infusion needles may be blunt

    Mentor is recalling Artoura Breast Tissue Expanders in the smooth ultra high profile range because the infusion sets may contain a dull or blunt needle tip.

    Product
    Artoura Breast Tissue Expanders Reference Numbers: SDC-100UH SDC-110UH SDC-120UH SDC-130UH SDC-135UH SDC-140UH Smooth Ultra High Profile Tissue Expander, 350cc, 455cc, 535cc, 650cc, 700cc, 850cc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1680-2026·2026-04-08

    Philips Azurion 5 M20 X-ray may not start from foot switch

    Philips is recalling the Azurion 5 M20 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1749-2026·2026-04-08

    Paradigm REAL-Time insulin pump delivery varies with pump height

    Medtronic is recalling Paradigm REAL-Time insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1704-2026·2026-04-08

    Finesse BTK angioplasty balloon catheter may not meet burst specifications

    The Finesse BTK Multicath injectable angioplasty balloon catheter is being recalled because of the potential for the balloon not to meet burst specifications.

    Product
    Brand Name: Finesse BTK Multicath Product Name: Injectable Angioplasty Balloon Catheter Model/Catalog Number: FS252251502 Product Description: 5 French sheath, 0.014" guidewire compatible over-the-wire injectable angioplasty balloon catheter, balloon diameter 2.5mm, balloon le
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1679-2026·2026-04-08

    Philips Azurion 5 M12 X-ray may not start from foot switch

    Philips is recalling the Azurion 5 M12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 5 M12; System Code: (1)722227, (2)722231;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0415-2026·2026-04-08

    Rugby and Walgreens lubricant eye drops recalled over sterility assurance

    K.C. Pharmaceuticals is recalling Sterile Eye Drops Soothing Tears, sold as Rugby Lubricating Eye Drops and Walgreens Lubricant Eye Drops, for lack of assurance of sterility.

    Product
    STERILE EYE DROPS SOOTHING TEARS (polyethylene glycol 400 0.4% and propylene glycol 0.3%), 0.5 OZ 0.5 fl oz (15 mL) bottles; a) Rugby LUBRICATING EYE DROPS, Distributed by: Rugby Laboratories, Livonia, MI 48152, NDC 0536-1219-94; b) Walgreens, Lubricant Eye Drops, DISTRIBUTED BY
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1659-2026·2026-04-08

    Philips Allura Xper FD10C X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD10C because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD10C; System Code: 722001;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1676-2026·2026-04-08

    Philips Azurion 7 B20 X-ray may not start from foot switch

    Philips is recalling the Azurion 7 B20 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1697-2026·2026-04-08

    da Vinci double fenestrated grasper recalled over broken grip cables

    Intuitive Surgical is recalling da Vinci S and Si Double Fenestrated Grasper Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci S, Si Double Fenestrated Grasper Instrument Model/Catalog Number: 420189 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including bl
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1667-2026·2026-04-08

    Philips Allura Xper FD10 OR Table X-ray may not start reliably

    Philips is recalling the Allura Xper FD10 OR Table because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1677-2026·2026-04-08

    Philips Azurion 7 M12 X-ray may not start from foot switch

    Philips is recalling the Azurion 7 M12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1673-2026·2026-04-08

    Philips Azurion 3 M12 X-ray may not start from foot switch

    Philips is recalling the Azurion 3 M12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1675-2026·2026-04-08

    Philips Azurion 7 B12 X-ray may not start from foot switch

    Philips is recalling the Azurion 7 B12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;
    Category
    Medical Device
    Distribution
    Distributed nationwide