The Recall Desk

State

Georgia product recalls

73,817 recalls have nationwide distribution and so reach Georgia. 9,635 additional recalls listed Georgia specifically in their distribution scope.

What the numbers show

Total
9,635
Critical
1,471
Severe
2,567
Most recent
2026-11-04

Of the 9,635 recalls naming Georgia we have scored against our rubric, 1,471 (15%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2,567 (27%) are Severe.

These figures cover only the 9,635 recalls whose distribution notice names Georgia specifically. The 73,817 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

47351–47375 of 83452

  • ModerateFDA (Devices)·Z-2886-2017·2017-08-16

    Valorem Surgical Recalls Pallas M/MAXIMIS 50mm MIS Rods for Quality Issues

    Valorem Surgical is recalling 2,225 Pallas M/MAXIMIS 50mm MIS Rods because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS 50mm MIS Rod. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2843-2017·2017-08-16

    FDA Recalls Piston Syringes and Hypodermic Needles Due to Sterility Documentation Issues

    International Medsurg Connection, Inc. is recalling 54,204,100 units of piston syringes and hypodermic needles because of insufficient documentation to support sterility claims.

    Product
    Piston Syringe and Hypodermic Needle, 5 CC L/L 22 G X 1.5, 100 per box, 15 boxes/case, 1,500/case The products are used to inject fluids into, or withdraw fluids from the body. They are intended for use in the diagnosis of disease or other conditions or in the cure…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2959-2017·2017-08-16

    Valorem Surgical Recalls Pallas M/MAXIMIS 400mm Straight Rods

    Valorem Surgical LLC is recalling 2,225 Pallas M/MAXIMIS 400mm Straight Rods because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS 400mm Straight Rod. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2940-2017·2017-08-16

    Valorem Surgical Recalls Pallas M 35mm Curved Rods for Quality Issues

    Valorem Surgical is recalling 2,225 Pallas M 35mm Curved Rods because they do not meet FDA Quality System Requirements.

    Product
    PALLAS M/MAXIMIS 35mm Curved Rod. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2930-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Pedicle Screws for Quality Issues

    Valorem Surgical LLC is recalling 2,225 Pallas M/MAXIMIS MIS Pedicle Screws because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS MIS Screw 6.5 x 55mm and 6.5 x 5.5 mm screw. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2989-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Extension Guide Bushes for Quality Issues

    Valorem Surgical LLC is recalling 262 Pallas M/MAXIMIS Extension Guide Bushes because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS Extension Guide Bush. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2844-2017·2017-08-16

    FDA Recalls Piston Syringes and Hypodermic Needles Due to Sterility Documentation Issues

    International Medsurg Connection, Inc. is recalling 54,204,100 piston syringes and hypodermic needles nationwide because of insufficient documentation to support product sterility claims.

    Product
    Piston Syringe and Hypodermic Needle, 10 CC L/L 20 G X 1.0, 100 per box, 10 boxes/case, 1,000/case The products are used to inject fluids into, or withdraw fluids from the body. They are intended for use in the diagnosis of disease or other conditions or in the cure…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2947-2017·2017-08-16

    Valorem Surgical Recalls Pallas M/MAXIMIS 35mm Straight Rods for Quality Issues

    Valorem Surgical LLC is recalling 2,225 Pallas M/MAXIMIS 35mm Straight Rods because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS 35mm Straight Rod. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2968-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Anti Torque Spinal Fixation Devices

    Valorem Surgical LLC is recalling 262 Pallas M/MAXIMIS Anti Torque devices due to non-compliance with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS Anti Torque. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2809-2017·2017-08-16

    FDA Recalls Hypodermic Needles Due to Insufficient Sterility Documentation

    International Medsurg Connection, Inc. is recalling hypodermic needles because the company lacks sufficient documentation to support the product's sterility claims.

    Product
    Hypodermic Needle, 19 G X 1.5 in BD YALE DISP RG, 100 per box, 50 boxes/case, 5,000/case The products are used to inject fluids into, or withdraw fluids from the body. They are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2993-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Anti Torque Devices for Quality Issues

    Valorem Surgical is recalling 262 Pallas M/MAXIMIS Anti Torque Devices because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS Anti Torque Device. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2949-2017·2017-08-16

    Valorem Surgical Recalls Pallas M/MAXIMIS 45mm Straight Rods

    Valorem Surgical is recalling 2,225 Pallas M/MAXIMIS 45mm Straight Rods because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS 45mm Straight Rod. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2937-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Pedicle Screws for Quality Issues

    Valorem Surgical LLC is recalling Pallas M/MAXIMIS Poly Screws because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS Poly Screw 6.5 x 45mm. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-3008-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Nut Inserters for Quality System Issues

    Valorem Surgical LLC is recalling 262 Pallas M/MAXIMIS Nut Inserters because the devices are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS Nut Inserter B. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2853-2017·2017-08-16

    FDA Recalls Piston Syringes Due to Insufficient Sterility Documentation

    International Medsurg Connection, Inc. is recalling 54,204,100 piston syringes because the company lacks sufficient documentation to support the product's sterility claims.

    Product
    Piston Syringe, 5 CC L/L, 100 per box, 20 boxes/case, 2,000/case The products are used to inject fluids into, or withdraw fluids from the body. They are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment or prevention of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2917-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Pedicle Screws for Quality Issues

    Valorem Surgical LLC is recalling 2,225 Pallas M/MAXIMIS MIS Pedicle Screws because they are not consistent with FDA Quality System Requirements.

    Product
    PALLAS M/MAXIMIS MIS Screw 6.5 x 55mm. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2834-2017·2017-08-16

    FDA Recalls Piston Syringes and Hypodermic Needles Due to Sterility Documentation Issues

    International Medsurg Connection, Inc. is recalling 54,204,100 units of piston syringes and hypodermic needles because the company lacks sufficient documentation to support the products' sterility claims.

    Product
    Piston Syringe and Hypodermic Needle, 3 CC L/L 22 G X 1.5, 100 per box, 24 boxes/case, 2,400/case The products are used to inject fluids into, or withdraw fluids from the body. They are intended for use in the diagnosis of disease or other conditions or in the cure…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2996-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Rod Checkers for Quality System Defects

    Valorem Surgical LLC is recalling 262 Pallas M/MAXIMIS Rod Checkers because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS Rod Checker. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2873-2017·2017-08-16

    Valorem Surgical Recalls Pallas M/MAXIMIS Instrument Taps for Quality Issues

    Valorem Surgical is recalling 2,225 Pallas M/MAXIMIS 5.5 mm Instrument Taps because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS 5.5 mm Instrument Tap. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2938-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Pedicle Screws for Quality Issues

    Valorem Surgical LLC is recalling 2,225 Pallas M/MAXIMIS Poly Screws because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS Poly Screw 6.5 x 50mm. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2972-2017·2017-08-16

    Valorem Surgical Recalls Pallas M/MAXIMIS Marking Pins for Quality Issues

    Valorem Surgical LLC is recalling 262 Pallas M/MAXIMIS Marking Pins because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS Marking Pin (Smooth). Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2833-2017·2017-08-16

    FDA Recalls Piston Syringes and Hypodermic Needles Due to Sterility Documentation Issues

    International Medsurg Connection, Inc. is recalling 54,204,100 units of piston syringes and hypodermic needles because the company lacks sufficient documentation to support the products' sterility claims.

    Product
    Piston Syringe and Hypodermic Needle, 3 CC L/L 22 G X 1.25, 100 per box, 24 boxes/case, 2,400/case The products are used to inject fluids into, or withdraw fluids from the body. They are intended for use in the diagnosis of disease or other conditions or in the cure…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2990-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Extension Sleeves for Quality System Issues

    Valorem Surgical LLC is recalling 262 Pallas M/MAXIMIS Extension Sleeves because they are not consistent with Quality System Requirements.

    Product
    PALLAS M/MAXIMIS Extension Sleeve. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2908-2017·2017-08-16

    Valorem Surgical Recalls Pallas M Pedicle Screws for Quality Issues

    Valorem Surgical LLC is recalling 2,225 Pallas M/MAXIMIS Poly Screws because they are not consistent with FDA Quality System Requirements.

    Product
    PALLAS M/MAXIMIS Poly Screw 5.5 x 30mm. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
    Category
    Medical Device
    Distribution
    Distributed nationwide