  • HighFDA (Devices)·Z-0751-2025·2025-01-01

    Daytona Small Stature Locking Caps recalled for manufacturing defects

    SeaSpine Orthopedics is recalling 92 Daytona Small Stature CoCr Locking Caps due to manufacturing defects that prevent proper component mating. The devices were distributed in seven U.S. states.

    Product
    Brand Name: Daytona Small Stature Product Name: CoCr LockingCap 4.5mm, Small Stature, T25, BT Model/Catalog Number: 79-0002 Software Version: N/A Product Description: Locking Cap Component: No
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0165-2025·2025-01-01

    Duloxetine delayed-release capsules recalled due to manufacturing impurity

    Duloxetine delayed-release 60mg capsules are being recalled nationwide due to N-nitroso-duloxetine impurity above recommended interim limits. Two lot numbers are affected: J0786744-061724 (blister cards) and B3002625-060524 (bottles).

    Product
    Duloxetine Delayed-Release Capsules, 60 mg, a) 30 count blister cards (NDC 70518-0937-04), b) 30 count bottles (NDC 70518-0937-03), Rx only, Source NDC 57237-0019-99, MFG: Rising Pharma, Inc., Allendale, NJ, Repackaged by: RemedyRepack Inc., Indiana, PA
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0159-2025·2025-01-01

    Prescription chlorpromazine tablets recalled due to manufacturing quality defect

    Glenmark Pharmaceuticals is recalling chlorpromazine hydrochloride 10mg tablets nationwide due to manufacturing deviations that allowed excess N-nitroso impurity.

    Product
    CHLORPROMAZINE HYDROCHLORIDE — CHLORPROMAZINE HYDROCHLORIDE (CHLORPROMAZINE HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0723-2025·2025-01-01

    Medline surgical patties and strips recalled for endotoxin contamination

    Medline surgical patties and strips used in certain procedure convenience kits were found to contain higher-than-expected endotoxin levels that may exceed safety specifications. The affected kits were distributed nationwide and in Canada.

    Product
    Medline brand, medical procedure convenience kits, labeled as: 1) CRANIOTOMY, SKU CDS982719X; 2) MVFF/FIBU #54-RF, SKU DYNJ21927Q; 3) MVFF/FIBU #54-RF, SKU DYNJ21927S; 3) NEURO PEDIATRIC, SKU DYNJ61607A; 4) CRANIOTOMY PACK, SKU DYNJ66419C; 5) LYMPHATIC BYPASS PACK, SKU DYNJ69907;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0725-2025·2025-01-01

    Surgical Patties and Strips with Endotoxin Contamination in Medline Convenience Kits

    Medline Industries is recalling surgical patties and strips in specific lots of KIT,SPINE LH Convenience Kits due to higher-than-expected endotoxin levels in raw materials. The contamination may result in out-of-specification endotoxin in finished sterile products.

    Product
    KIT,SPINE LH Convenience kit, SKU DYNJ906448A.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0158-2025·2025-01-01

    Duloxetine Delayed-Release Capsules Recalled Nationwide for Excess N-Nitroso Impurity

    Amerisource Health Services LLC is recalling Duloxetine Delayed-Release Capsules due to N-nitroso-duloxetine impurity above recommended limits. The affected product was distributed nationwide in 1,815 bottles with two specific lot numbers and expiration dates.

    Product
    Duloxetine Delayed-Release Capsules, USP, 60 mg, Rx only, 1,000 count bottles, Manufactured by: Aurobindo Pharma Limited, Hyderabad-500 090, India, For BluePoint Laboratories NDC 68001-415-08
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0160-2025·2025-01-01

    Chlorpromazine Hydrochloride tablets recalled for manufacturing impurity excess

    Glenmark Pharmaceuticals is recalling Chlorpromazine Hydrochloride tablets due to manufacturing deviations that resulted in an impurity exceeding FDA limits. The affected lot (17230133, expiring 12/31/2024) was distributed nationwide.

    Product
    CHLORPROMAZINE HYDROCHLORIDE — CHLORPROMAZINE HYDROCHLORIDE (CHLORPROMAZINE HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0767-2025·2025-01-01

    Siemens magnesium reagent cartridges recalled for false measurement results

    Siemens is recalling two lots of Magnesium Flex reagent cartridges due to potential falsely low magnesium measurements. The intermittent error may not be detected by standard quality control procedures.

    Product
    Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0719-2025·2025-01-01

    Surgical Patties and Strips Recalled for Endotoxin Contamination

    Medline Industries recalls DIEP FLAP Convenience kits containing surgical patties and strips due to higher-than-expected endotoxin contamination that may not meet sterile product specifications.

    Product
    DIEP FLAP Convenience kit, SKU DYNJ910120. Convenience kits used for various procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0742-2025·2025-01-01

    Galaxy System bronchoscope power supply defect may injure lung tissue

    Noah Medical's Galaxy System bronchoscope has defective power supply semiconductors that may fail during procedures, potentially causing pneumothorax or lung tissue injury when the scope is manually removed.

    Product
    Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
    Category
    Medical Device
    Distribution
    15 states
  • HighFDA (Devices)·Z-0740-2025·2025-01-01

    Siemens Blood Analysis Device Recalled Over Patient Name Mix-up Software Defect

    Siemens is recalling the epoc NXS blood analysis system due to a software flaw that can assign incorrect patient names to test results, risking misdiagnosis or mistreatment if the operator doesn't verify the patient name.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413583(CN, KR) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0741-2025·2025-01-01

    Siemens epoc blood analyzer software assigns incorrect patient names

    Siemens epoc NXS blood analysis system has a software defect that can incorrectly assign patient names to blood test results, potentially causing misdiagnosis or mistreatment.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413879(IN) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capi
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0735-2025·2025-01-01

    Blood analyzer system recalled for patient name assignment software error

    Siemens recalled its epoc NXS Host blood analyzer due to a software malfunction that can assign incorrect patient names to test results, potentially leading to misdiagnosis or mistreatment.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413497 (EU); Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capi
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0755-2025·2025-01-01

    Medline Surgical Scissors Recalled for Sterility Seal Defect

    Medline O.R. scissors may have weak seals that compromise sterility, increasing risk of contamination and infection. Approximately 21,440 units from lot 2023052490 have been recalled nationwide.

    Product
    O.R. SCISSORS, S/B, 5.5 STERILE, Medline Item No. DYND04000
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0732-2025·2025-01-01

    Abbott Vascular Inflation Device May Leak During Cardiovascular Procedures

    Abbott Vascular is recalling the 20/30 Priority Pack Accessory Kit w/Copilot inflation device used in cardiovascular procedures because it may leak at connection points, potentially causing air embolism.

    Product
    20/30 Priority Pack Accessory Kit w/Copilot, REF: 1003327, used for cardiovascular procedures.
    Category
    Medical Device
    Distribution
    40 states
  • HighFDA (Devices)·Z-0762-2025·2025-01-01

    MISHA Knee System Implants recalled due to fracture risk

    Moximed is recalling 115 MISHA Knee System Implants in multiple US states due to a supplier component defect that may cause the implant to fracture, potentially leading to pain, swelling, and stiffness.

    Product
    MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0156-2025·2025-01-01

    Duloxetine Delayed-Release Capsules Recalled Due to N-nitroso Impurity Contamination

    Amerisource Health Services is recalling certain lots of Duloxetine Delayed-Release Capsules due to N-nitroso impurity levels exceeding recommended limits. Consumers with the medication should contact their healthcare provider.

    Product
    Duloxetine Delayed-Release Capsules, USP, 30 mg, Rx only, 90 count bottles, Manufactured by: Aurobindo Pharma Limited, Hyderabad-500 090, India, For BluePoint Laboratories NDC 68001-414-05
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0164-2025·2025-01-01

    Diagnostic drug kit recalled due to sterility concerns and vial integrity issues

    Curium US is recalling Technetium TC 99m Sestamibi diagnostic kits (Lot 092-24006) due to improper vial crimps that could compromise sterility assurance. The recall affects approximately 5,160 vials distributed nationwide.

    Product
    TECHNETIUM TC 99M SESTAMIBI — TECHNETIUM TC 99M SESTAMIBI (TECHNETIUM TC 99M SESTAMIBI)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0731-2025·2025-01-01

    Abbott Vascular inflation device may leak during cardiovascular procedures

    Abbott Vascular is recalling 20/30 Priority Pack Accessory Kit inflation devices due to potential leaks that could cause air embolism during cardiovascular procedures.

    Product
    20/30 Priority Pack Accessory Kit/.096 RHV, REF: 1000186, or use during vascular procedures in conjunction with interventional and / or diagnostic devices (e.g., balloon dilatation catheters, atherectomy devices, stent delivery systems, intravascular ultrasound devices).
    Category
    Medical Device
    Distribution
    40 states
  • HighFDA (Devices)·Z-0736-2025·2025-01-01

    Siemens epoc Blood Analysis System Recalled for Patient Identification Software Error

    Siemens is recalling epoc NXS Host blood analysis systems due to a software malfunction that can assign incorrect patient names to test results, potentially leading to misdiagnosis and mistreatment if not verified by the user.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capi
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0157-2025·2025-01-01

    Duloxetine Delayed-Release 60 mg Capsules Recalled for N-Nitroso Impurity

    Amerisource Health Services is recalling Duloxetine Delayed-Release Capsules 60 mg due to N-nitroso impurity levels exceeding FDA recommended limits. The recall affects 20,734 bottles distributed nationwide.

    Product
    Duloxetine Delayed-Release Capsules USP, 60 mg, Rx only, 30 count bottles, Manufactured by: Aurobindo Pharma Limited, Hyderabad-500 090, India, For BluePoint Laboratories NDC 68001-415-04
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0724-2025·2025-01-01

    Medline Surgical Patties and Strips Recalled for Out-of-Spec Endotoxin Contamination

    Medline is recalling surgical patties and strips in custom surgical kits due to higher-than-expected endotoxin levels in raw materials. These may result in out-of-specification endotoxin levels in finished sterile products.

    Product
    Medline brand, medical procedure convenience kits, labeled as: 1) NEURO CRANIOTOMY CDS #36-RF, SKU CDS780119T; 2) NEURO KIT, SKU CDS981327Q; 3) NEURO POSTERIOR SPINE PACK-LF, SKU DYNJ0311501I; 4) ACOUSTIC NEUROMA PACK-LF, SKU DYNJ37741D; 5) SCRIPPS SW NEURO PACK-LF, SKU D
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0739-2025·2025-01-01

    Blood Analysis System Software Error Can Assign Wrong Patient Name

    Siemens epoc NXS Host blood analysis devices have a software malfunction that may assign the wrong patient name to blood test results, potentially leading to misdiagnosis or mistreatment.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413518 (ROW) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or ca
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0746-2025·2025-01-01

    BD FACSLyric Flow Cytometer Power Supply May Fail During Use

    BD FACSLyric flow cytometers are recalled for power supply failures that may prevent the device from powering on, risking specimen loss and potential respiratory effects from fume inhalation.

    Product
    Device Description: FACSLyric 2L6C Instrument US, Catalog Number 662876, UDI 00382906628763 The K170974 clearance document describes the device as follows: The BD FACSLyric" flow cytometer (3-1, 4-2, 4-2-2, and 4-3-3 optical configurations) systems consist of a flow cy
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0738-2025·2025-01-01

    Siemens Blood Analysis System Software Malfunction May Cause Patient Identification Error

    Siemens epoc NXS Host blood analysis systems are recalled due to a software malfunction that can assign incorrect patient names to blood tests, potentially leading to misdiagnosis or mistreatment.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413517(MX) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capi
    Category
    Medical Device
    Distribution
    Distributed nationwide