I.T.S. pilon and tibia plates heat more in MRI than stated
I.T.S. GmbH is recalling Pilon Plates with Angular Stability and Anterolateral Tibia Plates because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.
- Product
- Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 2109
- Category
- Medical Device
- Distribution
- Distributed nationwide