The Recall Desk

State

California product recalls

73,187 recalls have nationwide distribution and so reach California. 13,255 additional recalls listed California specifically in their distribution scope.

What the numbers show

Total
13,255
Critical
1,605
Severe
2,847
Most recent
2026-10-02

Of the 13,255 recalls naming California we have scored against our rubric, 1,605 (12%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2,847 (21%) are Severe.

These figures cover only the 13,255 recalls whose distribution notice names California specifically. The 73,187 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

13226–13250 of 86442

  • SevereFDA (Devices)·Z-1485-2024·2024-04-24

    Tablo Hemodialysis System recalled for PCB contamination in tubing

    Outset Medical is recalling Tablo Hemodialysis Systems because their silicone tubing may contain unremoved polychlorinated biphenyl acids (PCBAs), which could cause skin conditions, liver damage, reproductive harm, or cancer.

    Product
    Tablo Hemodialysis System, REF: PN-0003000, PN-0006000 and PN-0007001
    Category
    Medical Device
    Distribution
    42 states
  • SevereFDA (Devices)·Z-1501-2024·2024-04-24

    LifeVentEVO2 Ventilator Recalled for Potential Loss of Ventilation

    Philips Respironics is recalling the LifeVentEVO2 ventilator due to a software malfunction that can trigger a false Battery Depleted or Loss of Power alarm when power is sufficient, potentially causing sudden loss of ventilation.

    Product
    LifeVentEVO2, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1500-2024·2024-04-24

    Philips Garbin EVO Ventilator False Battery Alarm Causes Ventilation Loss

    Philips Respironics recalls 1,233 Garbin EVO ventilators. A software defect in the battery and power alarm system may cause sudden loss of ventilation despite adequate power supply.

    Product
    Garbin EVO, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1521-2024·2024-04-24

    FDA Recalls SonarMed Airway Sensor for Obstruction Detection Failure

    SonarMed Inc is recalling the AW-S035 NEO airway monitoring sensor due to failure to detect partial obstructions. The defect may prevent the device from alerting clinicians to airway obstruction.

    Product
    AW-S035 NEO SONARMED SENSOR 3.5MM. Used in conjunction with SonarMed Monitor for airway monitoring.
    Category
    Medical Device
    Distribution
    19 states
  • SevereNHTSA·24V288000·2024-04-24

    BMW Recalls 2014-2015 Models for Head Air Bag Inflator Defect

    BMW is recalling 2014-2015 2, 3, and 4 Series vehicles because head air bag inflators may fail, posing an injury risk.

    Product
    BMW — BMW 328D, 335I, 228I, 328I
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1535-2024·2024-04-24

    Medical Device Airway Monitor Software Defect Prevents Obstruction Detection

    A software anomaly in SonarMed's AW-M0001 airway monitor prevents detection of partial airway obstructions in 2.5mm and larger sensors. The device may fail to alert clinicians to blockages, affecting approximately 145 units across 19 states.

    Product
    AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring.
    Category
    Medical Device
    Distribution
    17 states
  • SevereFDA (Drugs)·D-0444-2024·2024-04-24

    Atovaquone Oral Suspension Recalled Due to Potential Bacillus Contamination

    AvKARE recalls Atovaquone Oral Suspension due to potential Bacillus cereus contamination of a non-sterile product. The recalled lot (AW0221A, Exp. 08/30/2025) was distributed nationwide; 153 cartons affected.

    Product
    Atovaquone Oral Suspension USP, 750 mg per 5 mL sachets, 20 Units x 5 mL cartons, Rx Only, Manufactured for: AvKARE, Pulaski, TN 38478, NDC 50268-086-12.
    Category
    Drug
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1502-2024·2024-04-24

    Philips Trilogy EV300 Ventilator False Power Alarm Defect Recall

    Philips is recalling 56,012 Trilogy EV300 ventilators due to a software defect that can trigger a false Battery Depleted or Loss of Power alarm despite adequate power, risking sudden loss of ventilation.

    Product
    Trilogy EV300, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Drugs)·D-0442-2024·2024-04-24

    Vancomycin Hydrochloride Oral Solution Recalled Nationwide for Superpotency Due to Overfilling

    Amneal Pharmaceuticals is recalling Vancomycin Hydrochloride Oral Solution nationwide due to manufacturing overfilling causing superpotency with higher-than-labeled concentration. The FDA classified this as Class I.

    Product
    VANCOMYCIN HYDROCHLORIDE — VANCOMYCIN HYDROCHLORIDE (VANCOMYCIN HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1520-2024·2024-04-24

    SonarMed Airway Monitoring Sensor Fails to Detect Partial Obstruction

    SonarMed Inc is recalling 690 units of the AW-S030 NEO SONARMED SENSOR because the sensor fails to detect partial airway obstruction. The devices were distributed across 20 US states.

    Product
    AW-S030 NEO SONARMED SENSOR 3.0MM. Used in conjunction with SonarMed Monitor for airway monitoring.
    Category
    Medical Device
    Distribution
    19 states
  • SevereFDA (Devices)·Z-1519-2024·2024-04-24

    SonarMed 2.5mm Airway Sensor Recalled for Obstruction Detection Failure

    SonarMed Inc is recalling the AW-S025 NEO SONARMED SENSOR 2.5MM airway monitoring device due to a malfunction that causes it to fail detecting partial obstruction, potentially preventing clinicians from identifying compromised airways.

    Product
    AW-S025 NEO SONARMED SENSOR 2.5MM. Used in conjunction with SonarMed Monitor for airway monitoring.
    Category
    Medical Device
    Distribution
    19 states
  • SevereFDA (Devices)·Z-1504-2024·2024-04-24

    Trilogy Evo Ventilator Recall Due to Potential Loss of Ventilation

    Philips Respironics is recalling the Trilogy Evo ventilator (76,713 units) because it may suddenly stop ventilating when a false Battery Depleted or Loss of Power alarm triggers despite sufficient power available.

    Product
    Trilogy Evo, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1499-2024·2024-04-24

    Philips Respironics Aeris EVO Ventilators Recalled for Loss of Ventilation Risk

    Philips Respironics is recalling Aeris EVO ventilators due to a software defect that may cause false battery or power alarms while the device retains adequate power, risking sudden loss of ventilation in pediatric and adult patients. Approximately 775 units are affected.

    Product
    Aeris EVO, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereUSDA FSIS·014-2024·2024-04-24

    ConSup Recalls Ready-to-Eat Sliced Prosciutto Produced Without Inspection

    ConSup North America Inc. is recalling approximately 85,984 pounds of ready-to-eat sliced prosciutto produced in Germany without the benefit of equivalent inspection.

    Product
    ConSup North America Inc. Recalls Ready-To-Eat Sliced Prosciutto Product Produced Without Benefit of Inspection
    Category
    Food
    Distribution
    8 states
  • HighFDA (Devices)·Z-1581-2024·2024-04-24

    Philips Intera 1.5T MRI System Patient Support Table Installation Defect

    Philips is recalling certain Intera 1.5T MRI systems because the patient support table floor plate may be incorrectly installed, which could affect patient safety.

    Product
    Intera 1.5T Omni/Stellar, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1594-2024·2024-04-24

    Microalbumin Calibrator Series Recalled for Negative Bias and Patient Result Misclassification

    Randox Laboratories is recalling its Microalbumin Calibrator Series due to a negative bias in its calibration. This could misclassify patient results and affect Quality Control procedures by up to 12%.

    Product
    Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1584-2024·2024-04-24

    Philips MRI System Patient Support Table Floor Plate Installation Issue

    Philips is recalling 9 Intera 3.0T Quasar Dual MRI systems (Model 781150) distributed worldwide. The patient support table floor plate may be incorrectly installed on these units.

    Product
    Intera 3.0T Quasar Dual, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1580-2024·2024-04-24

    Philips MRI patient support table floor plate installation defect recall

    Philips is recalling 107 Intera 1.5T MRI systems worldwide due to potential incorrect installation of the patient support table floor plate, which may create safety risks during patient use.

    Product
    Intera 1.5T Master/Nova, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1566-2024·2024-04-24

    Ingenia 3.0T magnetic resonance system patient support floor plate installation defect

    Philips is recalling Ingenia 3.0T MRI systems due to incorrect patient support table floor plate installation. Affected healthcare facilities should verify installation status and contact Philips for corrective action.

    Product
    Ingenia 3.0T, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1537-2024·2024-04-24

    Masimo Rad-G Pulse Oximeter May Randomly Power Off, Causing Loss of Monitoring

    Approximately 21,723 Masimo Rad-G pulse oximeters may randomly power off and on, causing temporary loss of patient monitoring. The FDA classified this as a Class II recall affecting devices distributed worldwide.

    Product
    Masimo Rad-G, Pulse Oximeter (W/Sensor), REF:9847, Rx Only,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1530-2024·2024-04-24

    Cook Incorporated Dilators Recalled for Low Packaging Seal Strength

    Cook Incorporated is recalling 5,998 dilators worldwide because packaging may not meet seal strength specifications, potentially compromising device sterility.

    Product
    Dilator - Intended to be used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the venous, arterial, biliary and renal systems. 1) JCD10.0-38-20 2) JCD20.0-38-20 3) JCD22.0-38-20 4)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1573-2024·2024-04-24

    MRI System Patient Support Table Floor Plate Installation Defect

    Philips Intera 1.0T MRI systems may have an incorrectly installed patient support table floor plate. This defect could affect patient safety during magnetic resonance imaging procedures.

    Product
    Intera 1.0T Omni/Stellar, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1565-2024·2024-04-24

    Philips Ingenia MRI System Patient Support Table Floor Plate Installation Defect

    Philips is recalling 135 Ingenia 1.5T S MRI systems worldwide because the patient support table floor plate may be incorrectly installed, creating a potential safety risk to patients and staff.

    Product
    Ingenia 1.5T S, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1536-2024·2024-04-24

    Masimo Rad-G Pulse Oximeter Recall: Potential Automatic Power Loss

    Masimo Corporation is recalling 443 Rad-G pulse oximeters that may automatically power off and on, interrupting patient monitoring. Affected units are distributed worldwide.

    Product
    Masimo Rad-G, Pulse Oximeter with temperature (W/Sensor), REF:9210, Rx Only,
    Category
    Medical Device
    Distribution
    Distributed nationwide